Yao-Hsuan Lai1, Jin-Chun Lim1, Ya-Chu Lee1, Jian-Jang Huang1,2. 1. Graduate Institute of Photonics and Optoelectronics, National Taiwan University, No. 1, Sec. 4, Roosevelt Road, Taipei 106, Taiwan. 2. Department of Electrical Engineering, National Taiwan University, No. 1, Sec. 4, Roosevelt Road, Taipei 10617, Taiwan.
Abstract
Traditional methods of monitoring biochemical reactions measure certain detectable reagents or products while assuming that the undetectable species follow the stoichiometry of the reactions. Here, based upon the metal-oxide thin-film transistor (TFT) biosensor, we develop a real-time molecular diffusion model to benchmark the concentration of the reagents and products. Using the nicotinamide adenine dinucleotide (NADH)-oxaloacetic acid with the enzyme of malate dehydrogenase as an example, mixtures of different reagent concentrations were characterized to extract the ratio of remaining concentrations between NAD+ and NADH. We can thus obtain the apparent equilibrium constant of the reaction, (8.06 ± 0.61) × 104. Because the whole analysis was conducted using a TFT sensor fabricated using a semiconductor process, our approach has the advantages of exploring biochemical reaction kinetics in a massively parallel manner.
Traditional methods of monitoring biochemical reactions measure certain detectable reagents or products while assuming that the undetectable species follow the stoichiometry of the reactions. Here, based upon the metal-oxide thin-film transistor (TFT) biosensor, we develop a real-time molecular diffusion model to benchmark the concentration of the reagents and products. Using the nicotinamide adenine dinucleotide (NADH)-oxaloacetic acid with the enzyme of malate dehydrogenase as an example, mixtures of different reagent concentrations were characterized to extract the ratio of remaining concentrations between NAD+ and NADH. We can thus obtain the apparent equilibrium constant of the reaction, (8.06 ± 0.61) × 104. Because the whole analysis was conducted using a TFT sensor fabricated using a semiconductor process, our approach has the advantages of exploring biochemical reaction kinetics in a massively parallel manner.
The analysis of biochemical
reactions is important for molecular
biology research and is critical for tissue distribution, the metabolism,
and drug discovery.[1−6] Biomaterials or biochemical reactions are usually monitored from
measurable reagents or products by methods such as gas monitoring,[7] colorimetric determination,[8,9] and
pH detection.[10] However, not all reactants
and products in the reaction are detectable due to the limitation
of the measurement method. The concentration of the undetectable material
is derived by calculating the concentration of the detectable substance
based on the stoichiometric coefficients. In addition, biochemical
reactions are often mediated by catalysts or stoichiometric reactants,
which affect the reaction rates. Slight errors in the measurement
of the concentrations of reactants and products may cause distortions
in the calculated equilibrium constants. Sophisticated measurement
and analysis methods are required to enhance the accuracy.Transistor-based
biosensors have been widely studied in detecting
proteins, ligands, nucleotides, and cells.[11−15] Because most reactants or products in the biochemical
reactions carry electrical charges, transistor-based biosensors are
very sensitive in detecting target analytes in the reactions. They
were also employed for monitoring the kinetics of biochemical reactions,
in which the complete and incomplete reactions, the equilibrium coefficient,
and the reaction rate can be obtained.[16−18] The abovementioned approaches
of extracting the concentrations of target analytes or reaction parameters
were often conducted by establishing their correlations with transistor
parameter changes, such as variations of current, threshold voltage,
and so forth. For example, we previously analyzed the lysozyme and
tri-N-acetyl-d-glucosamine (NAG3) reaction kinetics by monitoring the thin-film transistor (TFT)
drain current change with the molecular concentration.[18] Target analytes are recognized from the specific
time drain currents start to change.[15] However,
the accuracy of determining the concentration of target analytes becomes
a big challenge because the single-parameter changes are very sensitive
to the environmental interference, instability of the sensing devices,
and improper design of the experimental flow.In this paper,
a new approach is proposed to study the biochemical
reaction kinetics. We employed a molecular diffusion model to correlate
the concentration of molecules detected and the drain current response
of the TFT-microfluidic biosensor. Instead of extracting the transistor
current difference at a specific time target molecules arrived at
the sensing pad, this work monitors the current change profile when
molecules diffuse through the microfluidic channel over a period of
time. We demonstrate the applicability of the biosensor on extracting
the equilibrium constant of nicotinamide adenine dinucleotide, reduced
form (NADH)–oxaloacetic acid (OAA, the conjugate acid of oxaloacetate)
with its catalyst, malate dehydrogenase (MDH),[19,20] without labeling or immobilization. Even though the lack of a cross-linker
may lead to a higher limit of detection, the diffusion behavior of
biomolecules in the microfluidic channel will be less affected so
that more accurate results can be obtained.
Results
and Discussion
Bare Chip Current Response
Under
the condition that the microfluidic channel is prefilled with only
phosphate-buffered saline (PBS) (pH 7.4), the transient response of
the bare chip is shown in Figure . The drain current first decreases because the amorphous
layers of the device are under a constant electrical stress, which
is attributed to the device defects in the indium–gallium–zinc
oxide (IGZO) channel, gate dielectric, and thin-film interfaces,[21] and the ionic redistribution in solution.[22] Also shown in Figure is the instability of drain current caused
by the injection. A sudden change of drain current is observed at
around 50–100 s after 0.01 × PBS is injected. The time
interval between the injection and response detected using the TFT
is nearly independent of the microfluidic channel length adopted in
our design. Therefore, for the subsequent experiment, the length of
the microfluidic channel is chosen to be 7.9 mm (excluding the 6.1
mm-in-diameter inject inlet) to ensure that target molecules arrive
at the sensing pad much longer than the injection turbulence.
Figure 1
Transient drain
current profile of a bare TFT with the microfluidic
channel filled with PBS (black line). The current response profile
of the TFT when the 0.01 × PBS solution is injected to the microfluidic
channel at t = 200 s (red line) is shown. The corresponding
turbulence occurs at around 250–350 s.
Transient drain
current profile of a bare TFT with the microfluidic
channel filled with PBS (black line). The current response profile
of the TFT when the 0.01 × PBS solution is injected to the microfluidic
channel at t = 200 s (red line) is shown. The corresponding
turbulence occurs at around 250–350 s.
Real-Time Analysis of NADH and NAD+ Molecules
Diffusing toward the Sensing Pad
In the next
step, we separately measured and analyzed the drain current responses
of NADH and NAD+ solutions with the concentrations of 10,
3.33, 1, and 0.33 mM. When the target analyte arrives at the sensing
pad, the drain current changes correspondingly. The drain current
response of each solution is shown in Figure a,b. The decrease of drain current at t = 200–300 s is caused by the turbulence of the
injection (see Figure ), while the signal at t > 1000 s is induced
by
the target molecules. Since both NADH and NAD+ biomolecules
carry negative charges under pH 7.4, the mobile molecules arrive at
the Au metal plate (without cross-linkers) and deplete negative charges
in the upper side of the transistor channel, leading to the decrease
of drain current. The trend of current change is opposite to the case
in which cross-linkers are employed to capture target molecules.[18] When the positive-charged target molecules are
immobile on the sensing pad, they attract opposite charges that are
fixed in the upper channel, resulting in the decease of drain current.[18] To indicate the time interval where electric
charges carried by the analyte are sensed, the starting and ending
points are defined as the time the current slope changes beyond a
certain average. Judging from the results in Figure a,b, the slope of the point of interest is
defined to be 40% higher or lower than the average of previous at
most 20 points of the same stage. Based on the definition, the arrived
time of NADH and NAD+ is around 1000–1200 s, as
shown in Figure a,b.
Figure 2
(a) Transient
drain current responses of NADH solutions of different
concentrations. (b) Transient drain currents of NAD+ solutions
of different concentrations. Note that those marked in red are the
POI. (c) Curve fitting of NADH in the POI. (d) Curve fitting of NAD+ in the POI.
(a) Transient
drain current responses of NADH solutions of different
concentrations. (b) Transient drain currents of NAD+ solutions
of different concentrations. Note that those marked in red are the
POI. (c) Curve fitting of NADH in the POI. (d) Curve fitting of NAD+ in the POI.To formulate the drain
current response of NADH and NAD+ at different concentrations,
we derive the one-dimensional (1D)
diffusion model for the biomolecules. Assuming a 1D microfluidic channel,
biomolecules diffusing in the channel are expressed as[23]where C (mM) is the concentration
of the biomaterial, t (s) is the diffusion time, D (m2/s–1) is the diffusion
coefficient, and x (m) is the position
in the channel. Note that x = 0 is the location where
the solution is injected. The above equation is subject to the following
initial and boundary conditions in our experimentwhere C0 (mM)
is the initial concentration we constantly inject the biomaterial
to the channel at the inlet. Based on the boundary conditions, we
obtainwhere x0 (m) is the length of the sensing pad.
By assuming a linear
relationship between current changes and electrical charges carried
by the molecules, the drain current change after target molecules
arrive at the region of interest (ROI) becomes[23]where β is a constant and ΔI (A) is the drain current change induced by the biomolecules.The measured drain current within the time period molecules are
sensed, defined as the period of interest (POI), is fitted by eq . The results are shown
in Figure c,d. Parameters
extracted, such as constant B and τ, are shown in Table . From eq , since the length of the sensing pad and
the diffusion coefficient remain the same under different concentrations,[24] τ should be the same, which agrees with
the results extracted in Table . Because the parameters are extracted from the drain current
profile of each concentration, by correlating the constant B to the
initial concentration C0, as plotted in Figure , β can be
obtained for both NADH and NAD+. By plugging in the values
in Table , the functions
of drain current change of NADH and NAD+ are expressed
as
Table 1
Values of Parameters for Fitting Drain
Current Responses of NADH and NAD+ Molecules of Different
Concentrations
solution
concentration C0 (mM)
B of NADH (mA/M)
B of NAD+ (mA/M)
τ of NADH (s)
τ of NAD+ (s)
10
–0.288
–2.245
161.64
169.67
3.33
–0.096
–0.738
160.73
164.38
1
–0.028
–0.233
169.10
163.68
0.33
–0.009
–0.070
163.37
162.87
Figure 3
(a)
Correlation of B and the concentration of NADH; (b) Correlation
of B and the concentration of NAD+. The equation of the
correlation is shown in the figure.
(a)
Correlation of B and the concentration of NADH; (b) Correlation
of B and the concentration of NAD+. The equation of the
correlation is shown in the figure.In addition to predicting the diffusion behavior,
the concentration
of the biomolecules can be determined by fitting eqs and 6 to the measured
drain current. Under the same concentration, NAD+ has a
higher drain current decrement than NADH over the POI, which indicates
that NAD+ has higher net charges than NADH.[16]
Determination of the Apparent
Equilibrium
Constant of NADH–OAA Reaction
With the knowledge of
the correlation between the concentration of NADH and NAD+ and the transient drain current responses, the kinetics of the following
biochemical reaction is next analyzedSince the mixture of NADH and OAA involves
the dynamic change of OAA in the TFT biosensor, before characterizing
the reaction, the drain current response of OAA was first analyzed.
An example of 10 mM OAA is shown in Figure . We observe an increase of current at the
time around 1250 s, which will be considered as the arrival time of
OAA molecules to the ROI for the subsequent experiment.
Figure 4
Transient drain
current response when injecting 10 mM of OAA into
the microfluidic channel.
Transient drain
current response when injecting 10 mM of OAA into
the microfluidic channel.To extract the equilibrium constant of the reaction, we prepared
four different solutions {i.e., 10 mM NADH + 10 mM OAA + 80 units/mL
MDH [denoted as solution (1)], 3.33 mM NADH + 3.33 mM OAA + 26.6 units/mL
MDH [solution (2)], 1 mM NADH + 1 mM OAA + 8 units/mL MDH [solution
(3)], 0.33 mM NADH + 0.33 mM OAA + 2.66 units/mL MDH [solution (4)]},
with all premixed in separate microcentrifuges for 30 min before applying
to the microfluidic channel. In the solutions, the unit of MDH implies
the conversion of 1 μmole of NADH and OAA to NAD+ and malate per minute at pH of 7.5 and a temperature of 25 °C.
In our case, the amount of MDH mixed in each solution is sufficient
to ensure a complete reaction.The transient drain current responses
of the premixed solutions
are shown in Figure a. Among the curves, the first turning point occurs at the time around
1050–1100 s, which, as compared with the drain current profiles
in Figure a,b, corresponds
to the starting point of NADH and NAD+. For the mixture
with a higher concentration of reagents, such as solution (1) or (2),
a second turning point is observed around t = 1250
s, which is attributed to the detection of OAA molecules. When more
OAA molecules are detected, the drain current starts to increase.
Within the time interval where only NADH and NAD+ are detected,
by taking eqs and 6 into consideration, the drain current of the mixture
is expressed aswhere C0-NADH (mM) is the initial concentration of NADH in the solution and p is the proportion of NADH converted to NAD+. Equation was employed
to fit the transient response of drain current within the POI. The
fitting results are shown in Figure b, where the starting point is defined as t = 0. The values of p extracted at different mixtures are summarized
in Table . The ratio
of NAD+ to NADH is expressed as
Figure 5
(a) Transient drain current responses
of various mixtures and (b)
corresponding fitting curves in the POI.
Table 2
Ratio of NAD+ to NADH and
Apparent Equilibrium Constant Keq′ of Various Mixtures
solution
NAD+/NADH
Keq′
(1)
279.11
7.79 × 104
(2)
286.36
8.20 × 104
(3)
299.30
8.96 × 104
(4)
270.00
7.29 × 104
(a) Transient drain current responses
of various mixtures and (b)
corresponding fitting curves in the POI.The ratio is shown in Table . Furthermore, the apparent equilibrium constant Keq′ is
obtained by the following correlationIdeally, the stoichiometric number of both NADH and OAA is 1, which
implies that the amount of OAA converted to malate is the same as
that of NADH converted to NAD+. Therefore, the apparent
equilibrium constant can be expressed asBased on eq , the
apparent equilibrium constants of various solutions are shown in Table . For comparisons, Table summarizes Keq′ values of NADH–OAA reaction reported using various methods.
The apparent equilibrium constant we obtained is slightly lower than
the database value[25] but is within the
range from other reports.[26,27] Traditionally, the
method for detecting the NADH–OAA reaction relies on the unique
optical absorption property of NADH at the wavelength of 340 nm.[28] However, since the existence of NAD+ is optically undetectable, the optical absorption method may not
be suitable for the quantitative determination of reaction species.[28] For our approach, once the correlation between
reagent concentration and the drain current response profile is established,
the apparent equilibrium constant of different mixtures can be determined.
Table 3
Comparisons of the Apparent Equilibrium
Constant, Keq′, of NADH–OAA Reaction Reported
in the Literature and This Work
method
Keq′
eQuilibrator[25]
1.85 × 105
optical absorbance spectrophotometer[26]
9.80 × 104
optical absorbance spectrophotometer[27]
4.72 × 104
this work
(8.06 ± 0.61) × 104
The parameters explored in this work,
such as the ratio of NAD+ to NADH and the equilibrium constant,
are the key to the
investigation of biochemical reactions, especially for discovering
the metabolism mechanism. The IGZO-TFT biosensor as proposed in this
work can effectively monitor biochemical reactions. By analyzing multiple
drain current points in the time domain, the proposed diffusion model
is less sensitive to background instability that typical transistor
sensors may encounter. Even though the lack of cross-linkers limits
the minimum concentration that can be detected (limit of detection)
in the reaction, our approach avoids labeling or immobilization of
the target protein. It has the advantage of maintaining the molecular
structure of the protein so that the properties of the molecules are
not altered during the reaction. Also, the TFT and microfluidic channels
can be mass-manufactured in an array form, which is a high-throughput
method in detecting biochemical reactions.
Materials
and Methods
Device Design and Fabrication
The
TFT-microfluidic channel biosensor is shown in Figure , along with a microscopic image of the TFT
image. The TFT and microfluidic channel were fabricated on the separate
glass substrate; both are connected by Au bond wires. We employed
a dual-gate IGZO TFT (see Figure ) as the transducer to convert biological signals into
electric signals and the microfluidic channel for biomolecular diffusion.
The detailed device fabrication steps for IGZO TFTs and fluidic channels
have been described elsewhere.[16,17] The microfluidic channel
is composed of one inlet and one outlet with a diameter of 6.1 mm.
The dimensions of the fluidic channel are 14, 1, and 1 mm in length
(including the 6.1 mm diameter inlet), width, and height, respectively.
Figure 6
Biosensor
employed in this work for monitoring biochemical reaction.
The sensor is composed of a TFT in the left and a microfluidic channel
in the right. ROI is located on the Au sensing pad, which is connected
to the top gate through Au bond wires.
Figure 7
Illustration
of the diffusion of target molecules in the microfluidic
channel. The electrical charges carried by the molecules above the
Au sensing pad will be detected by the TFT.
Biosensor
employed in this work for monitoring biochemical reaction.
The sensor is composed of a TFT in the left and a microfluidic channel
in the right. ROI is located on the Au sensing pad, which is connected
to the top gate through Au bond wires.Illustration
of the diffusion of target molecules in the microfluidic
channel. The electrical charges carried by the molecules above the
Au sensing pad will be detected by the TFT.
Operation Principle and Measurements
The
concentration of target molecules is determined by the amount
of electrical charges carried. The TFT detects charges and converts
them to the drain current for readout. In our device, the bottom gate
(see Figure ) provides
bias voltage and regulates the operating drain current in the saturation
region. It also acts as a reference electrode for the transistor sensor.
As schematically shown in Figure , with the diffusion of target molecules in the microfluidic
channel, the negative electrical charges carried by the molecules
will be sensed in the ROI and deplete electrons in the TFT channel
through the Au wire. The drain current will be changed correspondingly.In this work, the biomolecules, NADH, NAD+, OAA, and
MDH, were diluted in 0.01 × PBS (pH = 7.4) to obtaine the desired
concentration. The drain–source voltage, VDS, of 5 V and the bottom gate–source voltage, VGS, of 10 V, were provided to the TFT during
measurement. The sample rate is 10 s. Since the whole experimental
setup is sensitive to the static charges in the environment and the
TFT current can be perturbed during the solution injection to the
microfluidic channel, before characterizing biochemical reactions,
we conducted two types of test runs to understand the stress behavior
of the TFT device and the interference on drain current when the solution
was injected. First, with the microfluidic channel prefilled with
only PBS (without any target molecules), the transient TFT current
was recorded. Second, the drain current response of injecting only
the buffer (0.01 × PBS) (without any target molecules) to the
microfluidic channel prefilled with 14 μL of the buffer was
analyzed.To characterize biochemical reactions, we first calibrated
current
responses of pure NADH and NAD+ solutions. The concentrations
of NADH and NAD+ selected in this work are 10, 3.33, 1,
and 0.33 mM. The transient drain current profiles were fitted following
the diffusion equation to correlate the concentration of the target
analyte with the current change. Second, the reaction kinetics of
NADH and OAA was investigated by monitoring the diffusion behavior
of the mixture. We then determined the amount of NADH molecules converted
to NAD+ so that the equilibrium constant of the bioreaction
was derived.
Conclusions
In this
work, a molecular diffusion model was demonstrated to extract
parameters of the biochemical reaction without labeling and forehanded
protein immobilization. Using the reaction between NADH and OAA with
the catalyst MDH as an example, correlation between the diffusion
model and the time-domain drain current profile of biomaterials was
established so that the function for determining reactants and products
can be derived. The ratio of NAD+ to NADH and the equilibrium
constant, Keq, (8.06 ± 0.61) ×
104, are comparable to other approaches in the literature.
Our method avoids determining the molecular concentration based solely
on the absolute current variation as in the previous works. Thus,
more reliable results that mitigate background instabilities can be
obtained. Since the microfluidic channels and TFTs were fabricated
based on the semiconductor process procedure, the determination of
biochemical characteristics of various solutions using our method
can be conducted in a parallel manner. Our sensors have the potential
to be applied to the pharmaceutical industry which requires high-throughput
measurements.