Qianqian Chen1, Li Yang1, Keke Guo1, Jialin Yang1, Ji-Min Han1. 1. Key Laboratory of Explosion Science and Technology of China, Beijing Institute of Technology, 5 Zhongguancun South Street, Haidian District, Beijing 100081, China.
Abstract
We report a practical fluorescent sensor device for the trace amount detection of hydrogen peroxide vapor. In this paper, we have significantly improved the performance of fluorescence analysis for the detection of peroxides by solving the problems of packaging and storage of active materials and transferring the chemical experiment phenomenon to the actual project output. The fluorescent sensor molecule, test substrates, mixing methods, and the way to improve the life time are carefully studied. Combined with the design of circuit and programming, a field-test prototype was designed for peroxide explosives and its performance and algorithm were screened and optimized. In the detection of traces of H2O2 generated by ultraviolet separation or leaked as inherent impurities, the high-efficiency and rapid detection of peroxide-based explosives is achieved. The detection limit of H2O2 is expected to reach 2 ppb, and the response time can reach <0.5 s.
We report a practical fluorescent sensor device for the trace amount detection of hydrogen peroxide vapor. In this paper, we have significantly improved the performance of fluorescence analysis for the detection of peroxides by solving the problems of packaging and storage of active materials and transferring the chemical experiment phenomenon to the actual project output. The fluorescent sensor molecule, test substrates, mixing methods, and the way to improve the life time are carefully studied. Combined with the design of circuit and programming, a field-test prototype was designed for peroxide explosives and its performance and algorithm were screened and optimized. In the detection of traces of H2O2 generated by ultraviolet separation or leaked as inherent impurities, the high-efficiency and rapid detection of peroxide-based explosives is achieved. The detection limit of H2O2 is expected to reach 2 ppb, and the response time can reach <0.5 s.
With
the emergence of new explosives and the development of improvised
explosive devices, the accidental explosions and terrorism attacks
have seriously threatened the national security as well as the safety
and property of the people.[1] In the last
decades, the management of energetic materials and effective detection
of bombs thus have attracted great attention from many countries,
especially for those with long borders, dense population, and many
transportation hubs and ports. Compared to the conventional explosive
detection methods, the vapor detection has proven to be a promising
way suitable for nondestructive remote explosive monitoring. For example,
the vapor detection of hydrogen peroxide (H2O2) implies many applications in industrial and biorelated monitoring
and moreover will provide a new way for detecting the peroxide-based
improvised explosives such as triacetone triperoxide (TATP), diacetone
diperoxide (DADP), and hexamethylene triperoxide diamine, from which
H2O2 is considered as a signature compound.[2]As a representative of peroxide explosives,
TATP was first synthesized
by the German scholar Richard Wolffenstein in 1895. However, this
explosive has not been used due to its poor stability. TATP is soluble
in a series of organic solvents such as toluene, acetone, and ethanol,
but insoluble in water.[3−7] During the synthesis of organic peroxide closed-loop trimer TATP,
its dimer DADP is produced as a byproduct of the reaction.[8] The byproduct DADP is less stable than the main
product TATP.[9,10] This increases the instability
of the reaction product to a certain extent. Because peroxide explosives
generally contain more than one peroxide group (−O–O−),
this type of group is prone to photolysis under the irradiation of
ultraviolet and visible light to generate H2O2, and so, steam detection of peroxide explosives is realized.However, vapor detection of H2O2, particularly
at the trace level (ppb), remains challenging for conventional sensor
techniques.[11−14] The challenge lies in the combined difficulty of molecular design
and materials engineering to produce a sensor system that not only
enables strong binding with H2O2 (for efficient
vapor sampling) but also expedient, selective reaction with H2O2 to transduce the readable signal.[15] Although few papers reported on chemical sensors
such as colorimetric sensors[16−21] that can be employed for vapor detection of H2O2, the reported sensors either suffer from the long response time
(>10 min) or complicated instrument alignments (e.g., involving
laser
and cooled charge-coupled device) or other deficiencies.In
recent years, the application of fluorescence turn-on (or enhanced)
molecular sensors in explosive detection has received increasing attention.[22] Therefore, the researchers have employed fluorescence
analysis to detect H2O2 vapor efficiently and
quickly. Xu[23] et al. used a fluorescent
probe molecule to react with H2O2 vapor generated
by the decomposition of peroxide explosives and realized the detection
of H2O2 vapor through the change of the fluorescence
spectra. Under H2O2 vapor concentration of 1
ppm, the response time could reach no more than 0.5 s. Fan[24] et al. investigated fluorenyl boronate ester
chromophore-based thin films for the detection of TATP via H2O2 from the decomposed product. An[25] et al. constructed a low-cost, portable, reusable, and
visible paper-based fluorescence sensor for sensitive detection of
TATP by steam sampling. Zhang[26] reported
two new fluorescent compounds, DB-WCZ and DB-W, which were served
as turn-off-type fluorescent film probes to H2O2 vapor. Therefore, the fluorescence analysis method reveals the advantages
of high sensitivity, good selectivity, simple operation, and easy
portability, which makes it one of the future development trends in
the field of explosive detection. However, most of the current studies
are still on the laboratory level. For example, their detection limits
and sensitivity response times are usually obtained by data fitting,
and there is no real practical exploration.Therefore, we are
working to fabricate a practical fluorescent
device for trace amount detection of H2O2 vapor,
which is expected to find broad applications in the areas of security,
environment, and biological monitoring. C6NIB is a naphthalimide-based
fluorescence turn-on sensor, which can be fabricated into a porous
matrix to enable trace vapor detection of H2O2 down to a few parts per billion. The sensor is extremely selective
to H2O2 with no response to other common reagents
even at orders of magnitude higher vapor concentrations. However,
there are still many problems to hinder it to be fabricated into a
practical, expedient, and reliable instrument suitable for trace vapor
detection of H2O2 (detection limit down to the
low ppb level). The challenges mainly focus on the lifetime of test
papers and engineering design to bring all the sophisticated testing
parts into a compact chamber.In this paper, we have significantly
improved the performance of
fluorescence analysis for the detection of peroxides by solving the
problems of packaging and storage of active materials and transferring
the chemical experiment phenomenon to the actual project output. First,
the performance the C6NIB fluorescent molecule is optimized to be
suitable toward engineering applications. Our work combines molecular
design and engineering technology and integrates detection parts and
devices. In the past, laboratory testing was an instant reaction of
the test paper without considering the storage problem and the problem
of testing efficiency changing with storage time. Due to the high
sensitivity of the test paper, it will be slowly oxidized in the air
and the simple packaging of plastic bags will shorten the life of
the test paper. In this paper, the vacuum tester is used to seal the
test paper to prevent it from contacting the air to a certain extent
and the life is improved. Also, we convert the fluorescence change
of the test strip before and after the reaction into an electrical
signal, which determines the existence of peroxides according to the
electrical signal change and sets an alarm threshold for the simple
output by careful coding. Really exploring the performance of fluorescence
analysis for the detection of peroxides from practical applications
has increased the practical value of previous fluorescence analysis
studies.
Experimental Section
Screening
of the Test Paper Matrix
First, we screened the preparation
process conditions of the test
substrates used in the project. Many test substrates that commonly
used in the laboratory were considered, such as the filter paper,
cotton, gauze, glass-based silica gel plate, aluminum foil-based silica
gel plate, and polyester-based silica gel plate. The purpose of this
step is to look for a balanced performance between the fluorescent
response and material processability. The reaction effect is shown
in Figure . The results
showed that the filter paper, cotton, and gauze did not meet the expected
experimental phenomenon. The pristine blue fluorescent lights were
immediately turned to yellow, which was the characteristic emission
wavelength of C6NIO (Figure a,b). Because the manufacturers normally used peroxide chemicals
to bleach the filter paper, cotton, and gauze, this fast transformation
from C6NIB to C6NIO was probably due to the residual peroxide bleaches
in substrates. Also, this phenomenon implied that the chemical sensor
we used in this work was extremely sensitive to trace the amount of
peroxide compounds. Then, we found that all types of the silica gel
plates could give pristine weak blue-violet emission in the absence
of H2O2 vapor and with a significant increase
in the yellow-green response after contacting H2O2 vapor. However, the glass substrate was heavy and hard to be folded
and cut, making it impossible to be fabricated into light soft materials
(Figure c). Although
polyester boards could be folded and cut and have obvious phenomena
before and after the reaction, the limited cohesive affinity between
the substrates and silica gels lead to the exfoliation of the silica
gel together with the active sensor materials. Additionally, the background
emission of the polyester substrate under UV light was also blue-violet
(Figure e), which
might interfere with the experimental phenomenon. The combined problems
hindered further engineering applications of the polyester substrate.
Compared with the glass and polyester substrates, the aluminum foil-based
silica gel plate demonstrates the best performance in both fluorescent
response and material processability (Figure d). The aluminum foil was soft and light
weighted and stuck firmly with the silica gel. Moreover, the fluorescent
background was satisfyingly low, and C6NIB can maintain stable for
very long time without H2O2 vapor. Therefore,
the aluminum foil-based silica gel board is ultimately used as the
test substrate in this work.
Figure 1
Reaction effect diagram on (a) two different
filter papers, (b)
cotton and gauze, (c) glass-based silica gel plate, (d) aluminum foil-based
silica gel plate, and (e) polyester-based silica gel plate. (f) Comparison
of polyester-based and aluminum foil-based substrates.
Reaction effect diagram on (a) two different
filter papers, (b)
cotton and gauze, (c) glass-based silica gel plate, (d) aluminum foil-based
silica gel plate, and (e) polyester-based silica gel plate. (f) Comparison
of polyester-based and aluminum foil-based substrates.
Testing of Material Mixing Methods
To fabricate the optimized test substrate, we used an organic base,
tetrabutylammonium hydroxide (TBAH), to produce the basic reaction
condition. C6NIB was proven stable in the fluorescence spectra within
the experimental time, but the fluorescence turn-on reaction of C6NIB
was found to be dependent on the way of mixing TBAH. Therefore, we
studied the different mixing methods of these two materials. The reaction
effect is shown in Supporting Information Figure S7. The preliminary results show that the three drop methods
have similar results. In comparison, mixing C6NIB and TBAH together
illustrates a more uniform spot, while the other two methods increase
difficulties to generate even films because two compounds have to
be added sequentially.For further quantitative analysis, the
excitation wavelength of 458 nm is used for the test paper of different
mixing methods (C6NIB/TBAH = 2:1, C6NIB concentration is 2 mmol/L).
The solid-state fluorescence emission spectra were tested before and
after the H2O2 solution was introduced to quantitatively
investigate the effect of the mixing methods on the reaction performance
of the test substrates (Figure ).
Figure 2
Fluorescent spectra of (a) C6NIB first, (b) TBAH first, and (c)
C6NIB and TBAH together. (d) Comparison of the emission peak growth
multiples of different mixing methods.
Fluorescent spectra of (a) C6NIB first, (b) TBAH first, and (c)
C6NIB and TBAH together. (d) Comparison of the emission peak growth
multiples of different mixing methods.The increase of the emission peak is calculated at the wavelength
of 522 nm. The results show that the growth of the peaks varies with
changed mixing methods. The mixing method of adding C6NIB and then
TBAH dropwise has the largest increment in the absorption peak in
5 and 15 s. Similarly, adding TBAH and then C6NIB can also achieve
the best performance, while the direct mixing can only give half the
value. However, in the case where the two substances are added separately,
it is impossible to ensure that the positions of dropwise addition
and diffusion are completely coincident. Considering the practical
application and low-cost fabrication, we select the direct mixing
method because of its acceptable performance and much easier operation.
Research on Concentration, Ratio, and Excitation
Wavelength
The concentration, ratio, and excitation wavelength
were screened and explored. While maintaining the fixed ratio of C6NIB
and TBAH, the concentration of C6NIB was changed, and a preliminary
qualitative investigation was made on its concentration change law.
It is shown in Supporting Information Figure
S8.Qualitative analysis of concentration shows that under different
ratios, the same reaction performance is exhibited with the change
of C6NIB concentration. Before the H2O2 gas
was introduced, as the concentration decreased, the blue-violet light
gradually weakened. When the solubility was too small, even the blue-violet
light was not seen; after the H2O2 gas was introduced,
the greater the concentration, the more obviously the blue-violet
changed to yellow-green, the faster the reaction during the H2O2 gas flow. Therefore, a series of single-substance
C6NIB solutions with different concentrations were prepared, and solid-state
fluorescence spectroscopy was performed on test papers of different
concentrations using an excitation wavelength of 458 nm. The curve
of the absorption peak growth multiple at 522 nm within 5 and 15 s
with concentration was made to quantitatively investigate the effect
of the concentration of a single substance C6NIB on the reaction performance,
and the optimal concentration was selected accordingly (Figure ).
Figure 3
Curves of the increase
of the emission peak with C6NIB concentration.
Curves of the increase
of the emission peak with C6NIB concentration.The test results show that the absorption peak growth multiple
increases first and then decreases with increasing concentration.
When the concentration is too large, the surface of the test (the
test paper is sealed for 6 h) has obviously turned yellow, and it
will not react with H2O2 gas. If the concentration
is too small, the reaction performance is poor. When the concentration
of C6NIB was 0.5 mmol/L, the absorption fold increase of the absorption
peak reached the maximum within 5 and 15 s, that is, the optimal concentration
of single substance C6NIB was initially determined to be 0.5 mmol/L.Under the condition that the C6NIB concentration was kept at 0.5
mmol/L, the ratio of C6NIB to TBAH was changed to investigate the
effect of the ratio on the reaction performance. It is shown in Supporting Information Figure S9.The qualitative
reaction results show that from left to right,
the blue-violet color becomes lighter and deeper before the gas is
introduced. After passing in the gas, the left half becomes yellow-green,
the phenomenon is obvious, and the right half is basically unchanged.
The closer it is to the left end, the faster is the response speed.
In the case of the same concentration and ratio, further quantitative
investigations were conducted. The test paper for the solid-state
fluorescence spectrum was tested when H2O2 gas
is introduced for 0, 5, and 15 s. The change of the absorption peak
growth factor was calculated when H2O2 gas is
introduced for 5 and 15 s, and a graph of the growth factor was plotted.
Then, the conclusions were analyzed to find the best match.The test results show that under a certain C6NIB concentration,
the absorption peak multiple fluctuates with the change in the ratio
of C6NIB to TBAH (Figure ). Compared with the test paper with different proportions
of TBAH, the response of single substance C6NIB is relatively good.
When C6NIB/TBAH = 1:2, the effect is particularly prominent, which
is used as the best ratio.
Figure 4
Curves of the increase of the emission peak
with the ratio of C6NIB
to TBAH.
Curves of the increase of the emission peak
with the ratio of C6NIB
to TBAH.With the selected concentration
and ratio, the absorption spectra
of the same test strip before and after the H2O2 gas was tested were tested using excitation wavelengths of 365,
415, and 458 nm (Figure ). The peak of the absorption peak corresponding to different excitation
wavelengths was found, and the growth multiples under the absorption
peaks before and after the H2O2 is introduced
at different excitation wavelengths were calculated (Table ).
Figure 5
Fluorescence spectra
measured at (a) 365, (b) 415, and (c) 458
nm.
Table 1
Analysis of the Influence
of Different
Excitation Wavelengths
Fluorescence spectra
measured at (a) 365, (b) 415, and (c) 458
nm.The experiment is performed under the condition that
other conditions
are the same and only the excitation wavelength is changed. The results
show that when the same test strip is tested at different excitation
wavelengths, the absorption peaks before and after the reaction occur
at different wavelengths. When the excitation wavelength changes from
365 to 458 nm, the peak wavelength gradually red-shifts. The same
test paper is tested at different excitation wavelengths, and the
absorption peak growth times before and after the H2O2 gas is passed are different. When the 458 nm excitation wave
is detected, the absorption peak growth times are the largest.
Test Paper Storage Life Tracking
Based on known conditions,
the storage life of the test paper is
tracked. First, under the laboratory conditions, qualitative observation
experiments were performed on test papers with the same concentration,
the same ratio, the same storage method, and different storage times
before and after the H2O2 was introduced. It
is shown in Supporting Information Figure
S10.The laboratory qualitatively observed that the test paper
was partially oxidized after several days of storage, and the reaction
rate became slower. The test strip 1 month ago had been completely
oxidized, and no reaction phenomenon was observed even when H2O2 was added again. However, the degree to which
the test paper is oxidized, that is, the change in the life decay
amount with storage time is unknown. Therefore, test papers with a
ratio of C6NIB to TBAH of 2:1 and C6NIB concentrations of 2, 1, and
0.67 mmol/L were prepared each day, and a solid-state fluorescence
spectrum was measured after 9 days to investigate the life decay law.The laboratory observed that the test paper, which was placed naturally
without any protection measures, could not be used after 2 days. Using
plastic wrap to wrap, the test paper in a plastic bag can effectively
extend its service life. On this basis, a vacuum packaging machine
is used to further improve the storage method in order to better extend
its service life.Taking 2 mmol/L as an example, the scatter
plots of two different
packaging methods and the graphs of exponential fitting based on this
are given at the bottom for analysis and comparison (Figure ). See the Supporting Information for the result analysis diagrams of
other concentrations and time periods.
Figure 6
(a) 2 mmol/L, 5 s. (b)
2 mmol/L, 10 s.
(a) 2 mmol/L, 5 s. (b)
2 mmol/L, 10 s.It can be seen from the fitted
curve diagram that the absorption
peak growth multiple has a single change relationship with time, which
conforms to the exponential curve. With the extension of the storage
time, the increase of the absorption peak under the VM condition is
greater than that under the PB condition. The long-term preservation
effect of VM is better, which greatly improves the service life of
the test paper. Enable the fluorescence detection substance to go
out of the laboratory and get a certain application in engineering
practice.
Results and Discussion
Production of the Test Prototype
The fluorescence detection
system for peroxide explosives studied
in this paper mainly includes the flow field shaping control of the
fluorescence reaction of gaseous chemical substances, the precise
optical path structure design, the high stability modulation light
source drive, the detection of weak fluorescence signals, the design
of multichannel isolated power supply, and the integration of the
system master control. On the basis of establishing a reliable flow-field
simulation model, the electrical and structural design of the excitation
light source and fluorescence detector is completed.In the
structural design, the peroxide explosive test prototype system includes
a fluorescence-generating device, a fluorescence-generating device-fixing
device, a fluorescence-receiving device, a fluorescence-receiving
device-fixing device, a detection gas chamber device, a fluorescence
detection sensor-carrying device, a fixing bracket, and a handle and
shell. The fluorescence-generating device is fixed on the fixing device
of the fluorescence-generating device and then installed on the fixing
bracket; the fluorescence-receiving device is fixed on the fixing
device of the fluorescence-receiving device and then installed on
the fixing bracket; the detection gas chamber structure is fixed on
the fixing bracket and the fluorescence detection sensor-carrying
device is matched with it; the device shell is connected with the
fluorescence-receiving device-fixing device and the fixing bracket,
and the handle is fixed on the shell. This project is combined with
device design and circuit design. A prototype that can detect H2O2 gas has been preliminarily designed, and its
appearance is shown below (Figure ).
Figure 7
Schematic diagram of the internal workflow of the testing
prototype
(upper). Schematic diagram of (a) front panel, (b) rear panel, and
(c) upper panel and test strip base (below).
Schematic diagram of the internal workflow of the testing
prototype
(upper). Schematic diagram of (a) front panel, (b) rear panel, and
(c) upper panel and test strip base (below).
Combination of the Test Prototype and Test
Paper
The prototype is designed with a self-cleaning function
to clean any residual H2O2 gas. The test strips
are disposable and need to be replaced with new ones before each test.
When H2O2 gas is detected, the buzzer sounds
long alarm; when it is not detected, the prototype remains the same
under normal working conditions. Combining reading software, mapping
software, and testing prototypes, we can preliminarily observe the
reaction of the test substrate and the working status of the device
during the testing process. In the experiment, a 458 nm light source
was used to perform an immediate reaction on the test strip, and the
reaction was performed after a period of time for comparison to explore
the working life of the test strip.We then investigate the
lifetime of the test substrate by two ways. First, we tested its stability
under the long-time irradiation to examine the photobleaching effect
(Figure ). The original
response of the freshly made sensor materials shows 3.3% signal changing
within first 5 s under 5% H2O2 vapor. After
3 h of 458 nm light to irradiate the sensor material, although the
reaction performance decreased to 2.8% within 5 s under 5% H2O2 vapor, the phenomenon was still obvious enough to analyze
the detection progress, which was in line with the expected goal.
It is to be noted that we found a 10 s delay time in this process,
which might be due to the slow vapor diffusion and long pipe design.
Figure 8
(a) Instant
response. (b) Reaction after 3 h of light.
(a) Instant
response. (b) Reaction after 3 h of light.Next, we further tested the reaction performance of the test paper
after a certain period of storage to examine the oxygen and environmental
effects (Figure ).
The response of the 5% H2O2 vapor within 5 s
slightly dropped to 2.5% after 1 week of storage, which demonstrate
that the long-time storage may affect the response efficiency. Similarly,
the response of the 5% H2O2 vapor within 5 s
furtherly decreased to 1.9% after 2 weeks of storage, which also meets
our expectation. The data show that the test paper can maintain most
of their performance after 2 weeks because of our storage method.
The storage life of the test paper is good in actual tests and meets
the expected goals.
Figure 9
Response effect maps after (a) 1 week of storage and (b)
2 weeks
of storage.
Response effect maps after (a) 1 week of storage and (b)
2 weeks
of storage.
Anti-interference
Performance Test
The common solvents in the laboratory, acetonitrile,
acetone, petroleum
ether, and ethyl acetate, and a certain brand of perfume were compared
with H2O2 to determine the anti-interference
performance of the test paper. The experimental results are shown
in Figure . The
results show that our substrate has good anti-interference performance.
Figure 10
Test
substrate anti-interference performance test chart.
Test
substrate anti-interference performance test chart.
Commissioning of the Testing Prototype
Based on a large number of experiments and analysis of data, the
prototype is debugged. The alarm threshold is set for peroxide detection,
and the crude data in the test are processed and combined with image
processing for subsequent analysis. The schematic diagram of the prototype
test after debugging is as follows.In Figure a, the red line represents the coarse data,
and the yellow line represents the processed data, which is also the
original data. The data collection method may be to collect one every
other segment. In Figure b, the red line is the average of 1000 points; the purple
line is an alarm line; the buzzer will alarm when the purple line
fluctuates; the blue line is the data collected at intervals; and
when the blue line intersects with the purple line and continues to
decline, the buzzer will alarm. The yellow line is another way of
judging, but a large amount of experimental data show that the blue
line is better than the yellow line; so, the criterion represented
by the blue line is ultimately adopted.
Figure 11
Schematic test of the
prototype after debugging, (a) raw data collected
by the prototype and (b) guidelines for setting the prototype alarm
domain.
Schematic test of the
prototype after debugging, (a) raw data collected
by the prototype and (b) guidelines for setting the prototype alarm
domain.The schematic diagram of the specific
detection process is shown
in Figure . A breakthrough
from laboratory mechanism exploration to practical application of
engineering was realized. Finally, 30% H2O2 solution
was diluted with ethanol 5200 times, which can be detected within
30 s. At this time, the H2O2 concentration is
0.046 mg/mL (13 ppb), which is equivalent to the concentration of
H2O2 produced by the complete decomposition
of 0.1 mg/mL TATP. The detection limit that can be achieved in the
experiment is less than 2 ppb, and the detection time is greater than
70 s. When 30% H2O2 was directly used for detection,
the detection time was less than 0.5 s.
Figure 12
Detection flow chart.
Detection flow chart.
Conclusions
In summary,
we have screened the test paper substrate, finally
selected the aluminum foil-based silica gel plate to initially prepare
the reaction test paper, screened the process conditions, and carried
out qualitative and quantitative tests on the material mixing method.
The optimal concentration of C6NIB in the test paper, the ratio of
C6NIB to TBAH, excitation wavelength, test paper storage life, and
test paper anti-interference performance were tested.By designing
the peroxide explosive fluorescence detection system
and the prototype, we carried out the peroxide explosive fluorescence
detection experiment. The influencing factors of the fluorescence
detection sensor, including excitation light source wavelength, illumination
time, detection substance concentration, detection substance composition,
inhalation fan switch, H2O2 reagent concentration,
other organic solvents, sensor aging, and so forth were analyzed.
We gradually determined the optimized parameters of the fluorescence
detection sensor and the parameter settings of the prototype. Finally,
the detection capability of the prototype of the peroxide explosive
fluorescence detector developed in this paper was verified through
fluorescence detection experiments. The prototype can realize fast,
efficient, and accurate detection of peroxide explosives with a certain
accuracy.This work initially realized the transition from chemical
molecules
to engineering applications. This paper mainly focuses on the practical
application of engineering and explores the preparation, packaging,
storage, and detection of the reaction test paper to a certain extent.
A breakthrough has been achieved from laboratory mechanism exploration
to practical engineering application. Eventually, the detection limit
for H2O2 gas is 2 ppb, which is better than
the laboratory data fitting results to a certain extent.
Synthesis of the Substance C6NIB
The synthesis of the substance
C6NIB used for detection was performed
following the modified steps in the literature.[20] As shown in Scheme , the synthesis contains two steps. 4-Bromo-1,8-naphthalic
anhydride, hexylamine, and triethylamine were added into anhydrous
ethanol and refluxed to obtain the intermediate. Afterward, the intermediate
was mixed with anhydrous potassium acetate, bis(pinacolato)diboron,
[PdCl2(dppf)], and dppf in dioxane. The product was then
obtained as a white powder. In this study, C6NIB is only weakly fluorescent
in the UV region, where the quantum yield is only 0.6% under basic
conditions. However, upon reaction with H2O2, the aryl boranate group of C6NIB is transformed to phenol. The
weak blue-fluorescent naphthalimide backbone transition was converted
to the electron donor–acceptor (push–pull) C6NIO structure,
which turns on the charge transfer transition and fluorescent emission
in the longer wavelength band. Because the pristine C6NIB molecule
is close to zero emission in the charge transfer band, an extremely
high turn-on ratio will be obtained if the reaction with H2O2 is monitored in the long wavelength domain (Scheme ).
Scheme 1
Synthesis Route of
Sensor Molecule C6NIB
Scheme 2
Mechanism of the Reaction between C6NIB and H2O2
Materials and General Instrumentations
All raw materials and reagents were obtained from commercial suppliers
(Alfa Aesar, Aldrich, Weiss). C6NIB refers to the steps in Xu’s
article for synthesis and subsequent processing and verifies it with
NMR. The 1H NMR spectra and 13C NMR spectra
were recorded using a Bruker Avance 500 MHz superconducting nuclear
magnetic resonance spectrometer.Both the aluminum foil-based
silicone plate and the vacuum laminator were purchased from commercial
manufacturers and used as received. C6NIB is dissolved with chloroform
as the solvent, and TBAH is dissolved with ethanol as the solvent.
The experimental reagents (the concentration of C6NIB and the ratio
of C6NIB to TBAH) are all measured with a pipette and are prepared
in a 20 mL centrifuge bottle, and the additional solvent is ethanol.The aluminum foil-based silica gel plate is cut into a size of
20 mm × 20 mm as the test paper base, and a pipette is used to
suck 30 μL of liquid in the center of the test paper base and
let it to spread evenly and dry to complete the test paper preparation.
The sealing method of the plastic wrap and disposable plastic bag
mentioned in the article is to wrap the prepared test paper in the
plastic wrap, then put it into a disposable plastic bag, and store
it at room temperature away from light. The abovementioned storage
method of the cling film plus disposable plastic bag and vacuum laminator
is to wrap the prepared test paper with the cling film, then put it
in a disposable plastic bag, then put it in a vacuum packaging bag,
use vacuum plastic sealing, and store it at room temperature away
from light.The fluorescence emission spectrum was tested using
a RF6000 fluorescence
spectrophotometer.