Literature DB >> 3405132

Focal spot size measurements with pinhole and slit for microfocus mammography units.

C Kimme-Smith1, L W Bassett, R H Gold.   

Abstract

Many mammography units now have microfocal spots ranging in size from 0.09 to 0.4 mm for magnification radiography. On site measurements of these focal spots are not possible without invasive procedures if National Equipment Manufactures Association (NEMA) specifications are to be followed. This paper describes a method to perform such measurements expeditiously (1.5 h) and noninvasively. We describe how this method differs from NEMA test specifications, and report the test results of 22 units from ten manufacturers. Six of the focal spots were larger than the manufacturer's specifications which were based on current NEMA standards. Emphasis is placed on the need for verifiable standards for mammography x-ray systems.

Mesh:

Year:  1988        PMID: 3405132     DOI: 10.1118/1.596264

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Relationship between x-ray illumination field size and flat field intensity and its impacts on x-ray imaging.

Authors:  Xue Dong; Tianye Niu; Xun Jia; Lei Zhu
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

  1 in total