| Literature DB >> 34050386 |
Niels R van der Werf1,2, Ronald Booij3, Bernhard Schmidt4, Thomas G Flohr4, Tim Leiner5, Joël J de Groen3, Daniël Bos3,6, Ricardo P J Budde3, Martin J Willemink7, Marcel J W Greuter8,9.
Abstract
OBJECTIVES: The purpose of this study was twofold. First, the influence of a novel calcium-aware (Ca-aware) computed tomography (CT) reconstruction technique on coronary artery calcium (CAC) scores surrounded by a variety of tissues was assessed. Second, the performance of the Ca-aware reconstruction technique on moving CAC was evaluated with a dynamic phantom.Entities:
Keywords: Coronary artery disease; Diagnostic imaging; Radiation dosage; Tomography, x-ray computed
Mesh:
Substances:
Year: 2021 PMID: 34050386 PMCID: PMC8589753 DOI: 10.1007/s00330-021-08076-5
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Overview of phantom setup, with the robotic arm moving a coronary artery along the direction indicated by the red arrow (1), in the anthropomorphic thorax phantom (2), within the fillable compartment (3). Movement is generated with the Sim2D robot (4), which also provides an electrocardiogram output (5) to ensure data acquisition during linear motion of the artificial coronary artery
Acquisition and reconstruction parameters for the reference, reduced kVp, and tin-filtration scans
| Parameter | Reference | Reduced kVp | Tin filtration |
|---|---|---|---|
| Acquisition mode | Sequential | Sequential | Sequential |
| Ref. tube voltage (kVp) | 120 | 120 | 100Sn |
| Ref. tube current product1 (mAs/rot) | 80 | 80 | 534 |
| Tube voltage (kVp) | 120 | 1002 | 150Sn2 |
| Tube current (mAs/rotation) | 1003 | 1483 | 1363 |
| Collimation (mm) | 2 × 96 × 0.6 | 2 × 96 × 0.6 | 2 × 96 × 0.6 |
| Rotation time (s) | 0.25 | 0.25 | 0.25 |
| Temporal resolution (ms) | 66 | 66 | 66 |
| Slice thickness (mm) | 3 | 3 | 3 |
| Slice increment (mm) | 1.54 | 1.54 | 1.54 |
| Kernel | Qr36f/Sa36f5 | Qr36f/Sa36f5 | Qr36f/Sa36f5 |
| Reconstruction | FBP | FBP | FBP |
| Matrix | 512 × 512 | 512 × 512 | 512 × 512 |
| Field of view (mm) | 220 | 220 | 220 |
| CTDIvol (mGy) | 3.93 | 3.45 | 2.47 |
1Default quality reference tube current, with dose optimization slider at position 5 (calcium/bone)
2Automatically selected based on phantom size
3Based on water equivalent thickness of used phantom setup
4Standard for calcium scoring with Siemens Healthineers equipment
5Ca-aware reconstruction kernel (Sa36f) used to compare results with reference reconstruction kernel (Qr36f)
Ref reference, rot rotation, CTDIvol CT dose index volume, FBP filtered back projection
Fig. 2Flowchart which describes the methodology to account for the tube voltage dependency of the used iodine-water mixture to resemble soft tissue
Background material mean (mean ± SD) and noise (mean ± SD) for all combinations of tube potential, reconstruction kernel, and background material, on average for all used velocities and repetitions
| Tube potential (kVp) | Kernel | Fat | Water | Soft tissue | |||
|---|---|---|---|---|---|---|---|
| Mean | Noise | Mean | Noise | Mean | Noise | ||
| 120 | Standard | −94.4 ± 0.2 | 19.9 ± 0.7 | 0.0 ± 0.4 | 23.8 ± 0.3 | 51.3 ± 0.4 | 25.0 ± 0.4 |
| Ca-aware | −93.8 ± 0.3 | 19.9 ± 0.7 | 0.8 ± 0.4 | 23.8 ± 0.4 | 52.3 ± 0.4 | 25.0 ± 0.4 | |
| 100 | Standard | −99.0 ± 0.3 | 21.0 ± 1.1 | −1.7 ± 0.3 | 25.3 ± 0.4 | 51.8 ± 0.3 | 26.3 ± 0.6 |
| Ca-aware | −98.5 ± 0.3 | 21.2 ± 1.0 | −1.0 ± 0.3 | 25.4 ± 0.4 | 52.3 ± 0.3 | 26.2 ± 0.5 | |
| 150Sn | Standard | −84.5 ± 0.4 | 23.9 ± 0.6 | 2.4 ± 0.4 | 27.6 ± 0.6 | 49.7 ± 0.5 | 28.5 ± 0.5 |
| Ca-aware | −84.2 ± 0.3 | 23.7 ± 0.3 | 2.9 ± 0.4 | 27.8 ± 0.6 | 54.6 ± 0.6 | 31.4 ± 0.7 | |
Fig. 3Overview of representative images of the low-density calcifications for different combinations of acquisition and reconstruction parameters, and background material, for > 75 bpm
Fig. 4Overview of representative images of the medium-density calcifications for different combinations of acquisition and reconstruction parameters, and background material, for > 75 bpm
Fig. 5Deviations in low-density CAC Agatston score from the reference (for each heart rate: 120 kVp + standard reconstruction kernel) for different heart rates and combinations of tube voltage (kVp) and reconstruction. Results are shown for three background materials: fat (−100 HU, top), water (0 HU, middle), and soft tissue (50 HU, bottom). Clinically relevant differences, at ≥ ± 10%, are indicated with dashed lines
Fig. 6Deviations in medium-density CAC Agatston score from the reference (for each heart rate: 120 kVp + standard reconstruction kernel) for different heart rates and combinations of tube voltage (kVp) and reconstruction kernel. Results are shown for three background materials: fat (−100 HU, top), water (0 HU, middle), and soft tissue (50 HU, bottom). Clinically relevant differences, at ≥ ± 10%, are indicated with dashed lines
Percent difference with physical mass of the low-density calcification in mean ± SD for all acquisition, reconstruction, background material, and heart rate parameters. To calculate the percentage difference, the numerator was the mass score, and the denominator was the physical mass (38 mg)
| Background material | Heart rate | 120 kVp | 100 kVp | 150Sn kVp | |||
|---|---|---|---|---|---|---|---|
| Standard | Ca-aware | Standard | Ca-aware | Standard | Ca-aware | ||
| Fat | 0 | −47.9 ± 2.2 | −47.8 ± 2.3 | −45.1 ± 2.0 | −46.2 ± 1.8 | −73.7 ± 5.4 | −63.0 ± 4.1 |
| < 60 | −48.9 ± 1.7 | −48.8 ± 1.5 | −46.0 ± 3.0 | −47.2 ± 2.9 | −72.8 ± 2.5 | −63.0 ± 2.7 | |
| 60–75 | −53.5 ± 0.6 | −53.5 ± 0.6 | −51.1 ± 1.6 | −52.6 ± 1.6 | −77.2 ± 8.8 | −64.9 ± 3.5 | |
| > 75 | −55.8 ± 1.8 | −55.8 ± 1.8 | −54.7 ± 3.5 | −56.0 ± 3.5 | −78.7 ± 5.3 | −69.8 ± 3.8 | |
| Water | 0 | −39.5 ± 2.9 | −38.7 ± 2.9 | −38.0 ± 2.3 | −39.0 ± 2.3 | −62.3 ± 3.7 | −49.9 ± 2.6 |
| < 60 | −42.7 ± 2.3 | −42.2 ± 2.4 | −37.7 ± 1.1 | −38.8 ± 1.0 | −59.5 ± 4.1 | −47.5 ± 3.7 | |
| 60–75 | −41.4 ± 2.5 | −40.7 ± 2.1 | −39.1 ± 1.6 | −40.7 ± 1.5 | −65.3 ± 7.4 | −53.4 ± 6.7 | |
| > 75 | −47.4 ± 2.0 | −46.8 ± 1.8 | −42.9 ± 0.7 | −44.2 ± 1.3 | −74.3 ± 8.5 | −57.7 ± 3.0 | |
| Soft tissue | 0 | −62.3 ± 3.7 | −49.9 ± 2.6 | −33.5 ± 2.3 | −32.4 ± 2.0 | −28.5 ± 2.6 | −29.8 ± 2.9 |
| < 60 | −59.5 ± 4.1 | −47.5 ± 3.7 | −39.3 ± 6.4 | −37.5 ± 5.9 | −26.3 ± 4.0 | −27.6 ± 4.1 | |
| 60–75 | −74.3 ± 8.5 | −57.7 ± 3.0 | −35.3 ± 4.5 | −33.4 ± 4.2 | −30.5 ± 3.6 | −33.6 ± 2.6 | |
| > 75 | −74.3 ± 8.5 | −57.7 ± 3.0 | −35.3 ± 4.5 | −33.4 ± 4.2 | −30.5 ± 3.6 | −33.6 ± 2.6 | |
Percent difference with physical mass of the medium-density calcification in mean ± SD for all acquisition, reconstruction, background material, and heart rate parameters. To calculate the percentage difference, the numerator was the mass score, and the denominator was the physical mass (74 mg)
| Background material | Heart rate | 120 kVp | 100 kVp | 150Sn kVp | |||
|---|---|---|---|---|---|---|---|
| Standard | Ca-aware | Standard | Ca-aware | Standard | Ca-aware | ||
| Fat | 0 | −26.7 ± 1.2 | −27.3 ± 1.1 | −28.2 ± 0.9 | −29.1 ± 1.0 | −36.1 ± 1.9 | −34.0 ± 1.5 |
| < 60 | −27.4 ± 0.4 | −28.1 ± 0.3 | −29.4 ± 1.1 | −30.3 ± 1.0 | −36.4 ± 1.1 | −34.1 ± 1.2 | |
| 60–75 | −29.3 ± 1.4 | −29.9 ± 1.4 | −30.7 ± 1.3 | −31.7 ± 1.4 | −49.9 ± 19.5 | −37.5 ± 2.6 | |
| > 75 | −34.0 ± 0.8 | −34.5 ± 0.8 | −33.1 ± 1.6 | −34.2 ± 1.5 | −42.8 ± 2.7 | −40.0 ± 2.2 | |
| Water | 0 | −17.4 ± 1.7 | −17.4 ± 1.7 | −18.3 ± 1.9 | −19.1 ± 2.1 | −21.1 ± 1.3 | −16.3 ± 0.7 |
| < 60 | −17.7 ± 1.6 | −17.6 ± 1.7 | −18.5 ± 1.3 | −19.2 ± 1.4 | −22.0 ± 1.0 | −18.4 ± 1.2 | |
| 60–75 | −18.7 ± 1.0 | −18.7 ± 1.1 | −21.2 ± 1.1 | −22.1 ± 1.1 | −26.2 ± 2.7 | −21.6 ± 2.8 | |
| > 75 | −20.1 ± 2.1 | −20.1 ± 2.0 | −18.8 ± 1.9 | −19.9 ± 1.9 | −38.9 ± 23.5 | −21.1 ± 2.4 | |
| Soft tissue | 0 | −21.1 ± 1.3 | −16.3 ± 0.7 | −10.9 ± 0.9 | −10.6 ± 0.9 | −11.8 ± 0.8 | −12.6 ± 0.7 |
| < 60 | −22.0 ± 1.0 | −18.4 ± 1.2 | −12.6 ± 1.1 | −12.2 ± 1.2 | −11.4 ± 0.5 | −12.1 ± 0.5 | |
| 60–75 | −38.9 ± 23.5 | −21.1 ± 2.4 | −7.0 ± 1.9 | −5.6 ± 2.1 | −5.5 ± 2.9 | −7.3 ± 2.2 | |
| > 75 | −38.9 ± 23.5 | −21.1 ± 2.4 | −7.0 ± 1.9 | −5.6 ± 2.1 | −5.5 ± 2.9 | −7.3 ± 2.2 | |