| Literature DB >> 34040201 |
Samantha Kennedy1, Brooke Smith1, Sima Sobhiyeh1, Marcelline E Dechenaud2, Michael Wong3, Nisa Kelly3, John Shepherd3, Steven B Heymsfield4.
Abstract
OBJECTIVE: Three-dimensional optical (3DO) imaging devices for acquiring anthropometric measurements are proliferating in healthcare facilities, although applicability in young children has not been evaluated; small body size and movement may limit device accuracy. The current study aim was to critically test three commercial 3DO devices in young children.Entities:
Mesh:
Year: 2021 PMID: 34040201 PMCID: PMC8617044 DOI: 10.1038/s41430-021-00938-x
Source DB: PubMed Journal: Eur J Clin Nutr ISSN: 0954-3007 Impact factor: 4.016
Subject Characteristics
|
|
| |
|---|---|---|
|
| ||
|
| 29 | 35 |
| African American | 10 | 9 |
| White | 10 | 9 |
| Hispanic | 4 | 2 |
| Asian | 1 | 6 |
| NHOPI | 4 | 9 |
|
| ||
| Mean | 6.9 ± 1.1 | 6.6 ± 1.2 |
| Range | 5-8 | 5-8 |
|
| ||
| Mean | 126.0 | 125.2 |
| Range | 112.8 - 142.6 | 106.4 - 146.2 |
|
| ||
| Mean | 29.6 ± 10.8 | 28.7 ± 9.2 |
| Range | 19.2 - 66.5 | 17.4 - 55.0 |
|
| ||
| Mean | 18.4 ± 5.6 | 17.9 ± 3.1 |
| Range | 13.9 - 39.5 | 13.1 - 25.7 |
Abbreviations: NHOPI, Native Hawaiian and Other Pacific Islanders; BMI, body mass index. Results are X ± SD
Figure 1.Scan image errors present on a representative avatar produced by the SizeStream SS20 scanner. The image shows circumferences that do not align with the shape of the individual’s body and thus cannot be directly compared with reference circumferences. One example is the right mid-upper arm circumference, shown at the arrow, which not only surrounds the mid-upper arm but the chest as well.
Processed scans by age
| N | Fit3D | Size Stream | Styku |
|---|---|---|---|
|
| 64 | 53 | 64 |
|
| 43 | 52 | 26 |
|
| |||
|
|
|
|
|
| 5 | 38.5 | 90.0 | 0.0 |
| 6 | 63.6 | 100 | 45.5 |
| 7 | 84.2 | 100 | 57.9 |
| 8 | 71.4 | 100 | 47.6 |
|
| 67.2 | 98.1 | 40.6 |
Results summary of processed scans
|
|
| ||
|---|---|---|---|
|
| |||
|
| |||
| Fit3D | 65.5 ± 11.9 | −1.5 | |
| Styku | 64.2 ± 10.3 | 5.5 | |
| Size Stream | 62.5 ± 10.2 | −1.6 | |
|
| |||
| Fit3D | 75.4 ± 9.9 | −4.2 | |
| Styku | 77.5 ± 9.0 | −1.0 | |
| Size Stream | 70.0 ± 10.4 | −1.7 | |
|
| |||
| Fit3D | 22.3 ± 4.1 | −1.8 | |
| Styku | 22.3 ± 4.2 | 1.7 | |
| Size Stream | 22.3 ± 4.3 | −1.9 | |
|
| |||
| Fit3D | 21.2 ± 3.2 | −1.4 | |
| Styku | 20.6 ± 1.8 | 1.5 | |
| Size Stream | 21.8 ± 3.6 | −2.5 | |
|
| |||
| Fit3D | 37.2 ± 5.1 | 1.2 | |
| Styku | 44.0 ± 5.6 | −2.9 | |
| Size Stream | 33.8 ± 5.7 | 3.0 | |
|
| |||
| Fit3D | 37.4 ± 3.9 | 0.3 | |
| Styku | 45.1 ± 4.7 | −3.8 | |
| Size Stream | 33.3 ± 4.8 | 3.4 | |
Abbreviation: Δ mean = tape measurement mean - device measurement mean. Results are X ± SD
P < 0.01.
Figure 2.Circumferences estimated by the Fit3D Proscanner versus circumference tape measurements and associated Bland-Altman plots. For all regression plots, solid lines denote the line of regression while a dashed line indicates the line of identity. Regression equations with significant intercepts (p < 0.01) are marked *; equations with non-significant intercepts are set to the origin.
Figure 4.Circumferences estimated by the Styku S100 versus circumference tape measurements and associated Bland-Altman plots. For all regression plots, solid lines denote the line of regression while a dashed line indicates the line of identity. Regression equations with significant intercepts (p < 0.01) are marked *; equations with non-significant intercepts are set to the origin.
Figure 3.Circumferences estimated by the Size Stream SS20 versus circumference tape measurements and associated Bland-Altman plots. For all regression plots, solid lines denote the line of regression while a dashed line indicates the line of identity. Regression equations with significant intercepts (p < 0.01) are marked *; equations with non-significant intercepts are set to the origin.