Preliminary pulsatile flow bench validation of a thermodilution right ventricular ejection fraction system.

S Eaton1.   

Abstract

A preliminary Pulsatile Flow Bench validation of the REF-1 Right Ventricular Ejection Fraction (RVEF) computer and catheter system was performed to determine the accuracy of the current REF-1 system. These results reflect the accuracy of the current REF-1 system and not the final REF-1 system that will be released at a later date. The Pulsatile Flow Bench cardiac output and ejection fraction ranges included in this study are 1 to 6 Lpm and 15% to 60%, respectively. Currently, plans are being made to increase the flow bench cardiac output range to 10 Lpm with ejection fractions up to 70%. The results of this study show a correlation coefficient of 0.94 (adjusted R-square) between flow bench ejection fractions and REF-1 system ejection fractions. Also, the correlation coefficient between flow bench cardiac output and REF-1 system cardiac output was determined to be 0.99 (adjusted R-square).

Entities:  

Mesh:

Year:  1988        PMID: 3403787     DOI: 10.1007/bf00256959

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  4 in total

1.  Bedside evaluation of right ventricular performance using a rapid computerized thermodilution method.

Authors:  J F Dhainaut; F Brunet; J F Monsallier; D Villemant; J Y Devaux; M Konno; J M De Gournay; A Armaganidis; G Iotti; M F Huyghebaert
Journal:  Crit Care Med       Date:  1987-02       Impact factor: 7.598

2.  Thermodilution measurement of right ventricular ejection fraction with a modified pulmonary artery catheter.

Authors:  J L Vincent; M Thirion; S Brimioulle; P Lejeune; R J Kahn
Journal:  Intensive Care Med       Date:  1986       Impact factor: 17.440

3.  Hybrid computer for dye-dilution curves.

Authors:  I K Daskalov; L V Gueorguiev; M G Matveev
Journal:  Med Biol Eng Comput       Date:  1978-01       Impact factor: 2.602

4.  Measurement of ejection fraction by thermal dilution techniques.

Authors:  H R Kay; M Afshari; P Barash; W Webler; A Iskandrian; C Bemis; A H Hakki; E D Mundth
Journal:  J Surg Res       Date:  1983-04       Impact factor: 2.192

  4 in total
  2 in total

1.  In vitro validation of a thermodilution right ventricular ejection fraction method.

Authors:  S E Ferris; M Konno
Journal:  J Clin Monit       Date:  1992-01

2.  In vitro validation of a right ventricular thermodilution ejection fraction system.

Authors:  R Mukherjee; F G Spinale; A F von Recum; F A Crawford
Journal:  Ann Biomed Eng       Date:  1991       Impact factor: 3.934

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.