Stability of performances of vascular prostheses retrospective study of 22 cases of human implanted prostheses.

E Vinard1, R Eloy, J Descotes, J R Brudon, H Guidicelli, J L Magne, P Patra, R Berruet, A Huc, J Chauchard.   

Abstract

Twenty-two cases of vascular graft failures (thrombosis, pseudoaneurysm, dilation, infection) were investigated by means of physicochemical, histological, and ultrastructural studies. A general decrease of mechanical resistance to stress of the prostheses was observed. Its magnitude ranged from 2 to 75% of the values of identical virgin prostheses and there was no relation with the duration of implantation. In addition to the breakage of yarn filaments (SEM), migration of fiber debris (histology) was observed in 7 cases. The structure of the polyester molecule had evolved after implantation. A decrease of crystallinity (x-ray diffraction and differential scanning calorimetry) was noted. These observations raise the question of the established stability of vascular prostheses and emphasize the need for further investigations in human graft retrieval programs.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3403566     DOI: 10.1002/jbm.820220705

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  6 in total

Review 1.  Design and features of the Acorn CorCap Cardiac Support Device: the concept of passive mechanical diastolic support.

Authors:  Robert G Walsh
Journal:  Heart Fail Rev       Date:  2005-06       Impact factor: 4.214

Review 2.  Mesh biocompatibility: effects of cellular inflammation and tissue remodelling.

Authors:  Karsten Junge; Marcel Binnebösel; Klaus T von Trotha; Raphael Rosch; Uwe Klinge; Ulf P Neumann; Petra Lynen Jansen
Journal:  Langenbecks Arch Surg       Date:  2011-04-01       Impact factor: 3.445

Review 3.  Biocompatibility of prosthetic meshes in abdominal surgery.

Authors:  Marcel Binnebösel; Klaus T von Trotha; Petra Lynen Jansen; Joachim Conze; Ulf P Neumann; Karsten Junge
Journal:  Semin Immunopathol       Date:  2011-01-12       Impact factor: 9.623

4.  Lipid accumulation in prosthetic vascular grafts. Experimental study.

Authors:  E Chignier; J Guidollet; C Lhopital; P Louisot; R Eloy
Journal:  Am J Pathol       Date:  1990-09       Impact factor: 4.307

5.  Comparison of polypropylene and polyethylene terephthalate (Dacron) meshes for abdominal wall hernia repair: a chemical and morphological study.

Authors:  P Bracco; V Brunella; L Trossarelli; A Coda; F Botto-Micca
Journal:  Hernia       Date:  2004-09-10       Impact factor: 4.739

6.  Efficacy of pillar implants to reduce snoring and daytime sleepiness.

Authors:  Laith Khasawneh; Haitham Odat; Basheer Y Khassawneh; Khalid A Kheirallah; Adi H Khassawneh; Ahmad Al Omari; Maisa Smadi; Firas Alzoubi; Safwan Alomari; Abdel-Hameed Al-Mistarehi
Journal:  Future Sci OA       Date:  2021-03-29
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.