Impact of femoro-tibial size combinations and TKA design on kinematics.

Ingrid Dupraz1, Christoph Thorwächter2, Thomas M Grupp3,2, Florian Hammerschmid2, Matthias Woiczinski2, Volkmar Jansson2, Peter E Müller2, Arnd Steinbrück2.   

Abstract

INTRODUCTION: The variability in patients' femoral and tibial anatomy requires to use different tibia component sizes with the same femoral component size. These size combinations are allowed by manufacturers, but the clinical impact remains unclear. Therefore, the goals of our study were to investigate whether combining different sizes has an impact on the kinematics for two well-established knee systems and to compare these systems' kinematics to the native kinematics.
MATERIALS AND METHODS: Six fresh frozen knee specimens were tested in a force controlled knee rig before and after implantation of a cruciate retaining (CR) and a posterior-stabilized (PS) implant. Femoro-tibial kinematics were recorded using a ultrasonic-based motion analysis system while performing a loaded squat from 30° to 130°. In each knee, the original best fit inlay was then replaced by different inlays simulating a smaller or bigger tibia component. The kinematics obtained with the simulated sizes were compared to the original inlay kinematics using descriptive statistics.
RESULTS: For all size combinations, the difference to the original kinematics reached an average of 1.3 ± 3.3 mm in translation and - 0.1 ± 1.2° in rotation with the CR implant. With the PS implant, the average differences reached 0.4 ± 2.7 mm and  - 0.2 ± 0.8°. Among all knees, no size combination consistently resulted in significantly different kinematics. Each knee showed a singular kinematic pattern. For both knee systems, the rotation was smaller than in the native knee, but the direction of the rotation was preserved. The PS showed more rollback and the CR less rollback than the native knee.
CONCLUSION: TKA systems designed with a constant tibio-femoral congruency among size combinations should enable to combine different sizes without having substantial impact on the kinematics. The rotational pattern was preserved by both TKA systems, while the rollback could only be maintained by the PS design.
© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  Component size; Kinematics; Knee rig; Size combinations; TKA

Mesh:

Year:  2021        PMID: 34021794     DOI: 10.1007/s00402-021-03923-y

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  29 in total

Review 1.  Total Knee Arthroplasty Kinematics.

Authors:  Marc R Angerame; David C Holst; Jason M Jennings; Richard D Komistek; Douglas A Dennis
Journal:  J Arthroplasty       Date:  2019-05-29       Impact factor: 4.757

2.  Morphological measurement of the knee: race and sex effects.

Authors:  Pingyue Li; Tsung-Yuan Tsai; Jing-Sheng Li; Yu Zhang; Young-Min Kwon; Harry E Rubash; Guoan Li
Journal:  Acta Orthop Belg       Date:  2014-06       Impact factor: 0.500

Review 3.  Common controversies in total knee replacement surgery: Current evidence.

Authors:  Vasileios S Nikolaou; Dimitrios Chytas; George C Babis
Journal:  World J Orthop       Date:  2014-09-18

4.  Comparison of Contact Kinematics in Posterior-Stabilized and Cruciate-Retaining Total Knee Arthroplasty at Long-Term Follow-Up.

Authors:  Jordan S Broberg; Silvio Ndoja; Steven J MacDonald; Brent A Lanting; Matthew G Teeter
Journal:  J Arthroplasty       Date:  2019-08-06       Impact factor: 4.757

Review 5.  Posterior cruciate-retaining versus posterior stabilized total knee arthroplasty: a meta-analysis of randomized controlled trials.

Authors:  Ning Li; Yang Tan; Yu Deng; Liaobin Chen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-11-03       Impact factor: 4.342

6.  Three-dimensional morphology of the knee reveals ethnic differences.

Authors:  Mohamed Mahfouz; Emam Elhak Abdel Fatah; Lyndsay Smith Bowers; Giles Scuderi
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

Review 7.  Gap balancing vs. measured resection technique in total knee arthroplasty.

Authors:  Brian K Daines; Douglas A Dennis
Journal:  Clin Orthop Surg       Date:  2014-02-14

8.  Increased shape and size offerings of femoral components improve fit during total knee arthroplasty.

Authors:  Yifei Dai; Giles R Scuderi; Charles Penninger; Jeffrey E Bischoff; Aaron Rosenberg
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-07-16       Impact factor: 4.342

9.  High and Low Contrast Visual Acuity Are Not Affected in Amyotrophic Lateral Sclerosis.

Authors:  Heather E Moss; Monica Samelson; Girish Mohan; Qin Li Jiang
Journal:  PLoS One       Date:  2016-12-29       Impact factor: 3.240

Review 10.  What to Know for Selecting Cruciate-Retaining or Posterior-Stabilized Total Knee Arthroplasty.

Authors:  Sang Jun Song; Cheol Hee Park; Dae Kyung Bae
Journal:  Clin Orthop Surg       Date:  2019-05-09
View more
  1 in total

1.  Isolated effects of patellar resurfacing in total knee arthroplasty and their relation to native patellar geometry.

Authors:  Adrian Sauer; Christoph Thorwaechter; Ingrid Dupraz; Allan Maas; Arnd Steinbrueck; Thomas M Grupp; Matthias Woiczinski
Journal:  Sci Rep       Date:  2022-07-28       Impact factor: 4.996

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.