Literature DB >> 3400853

Pulse oximetry and postoperative hypothermia. An evaluation of the Nellcor N-100 in a cardiac surgical intensive care unit.

M R Gabrielczyk1, R J Buist.   

Abstract

The accuracy of the Nellcor N-100 pulse oximeter was evaluated in hypothermic patients (core temperature less than or equal to 35.0 degrees C) after cardiac surgery. The pulse oximeter overestimated oxygen saturation in comparison to values obtained by direct in vitro oximetry with a mean bias of 0.6% saturation. The 95% predictability limits for individual measurements were +/- 3.9% oxygen saturation. No demonstrable loss of accuracy was caused by skin pigmentation, the concurrent administration of low doses of dopamine or vasodilators, or the intra-operative intravascular administration of Patent Blue V dye.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3400853     DOI: 10.1111/j.1365-2044.1988.tb09025.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  7 in total

1.  Arterial line for monitoring SpO2 in patients with ischemic peripheries.

Authors:  Goneppanavar Umesh; Kaur Jasvinder; Shetty Nanda; Shetty Ranjan
Journal:  J Clin Monit Comput       Date:  2010-07-16       Impact factor: 2.502

Review 2.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

Review 3.  Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health.

Authors: 
Journal:  CMAJ       Date:  1992-03-01       Impact factor: 8.262

Review 4.  Pulse oximetry and capnography in intensive and transitional care units.

Authors:  F Bongard; D Sue
Journal:  West J Med       Date:  1992-01

Review 5.  Skin Pigmentation Influence on Pulse Oximetry Accuracy: A Systematic Review and Bibliometric Analysis.

Authors:  Ana M Cabanas; Macarena Fuentes-Guajardo; Katina Latorre; Dayneri León; Pilar Martín-Escudero
Journal:  Sensors (Basel)       Date:  2022-04-29       Impact factor: 3.847

6.  Weaning from ventilation after cardiopulmonary bypass: evaluation of a non-invasive technique.

Authors:  D E Withington; J G Ramsay; A T Saoud; J Bilodeau
Journal:  Can J Anaesth       Date:  1991-01       Impact factor: 5.063

7.  The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis.

Authors:  Chunhu Shi; Mark Goodall; Jo Dumville; James Hill; Gill Norman; Oliver Hamer; Andrew Clegg; Caroline Leigh Watkins; George Georgiou; Alexander Hodkinson; Catherine Elizabeth Lightbody; Paul Dark; Nicky Cullum
Journal:  BMC Med       Date:  2022-08-16       Impact factor: 11.150

  7 in total