Nephrostomy catheter entering the right renal vein during an exchange procedure: A case report and literature review.

Jun Akatsuka1, Yasutomo Suzuki1, Yuki Endo1, Masato Yanagi1, Ichiro Matsuzawa1, Tsutomu Hamasaki1, Go Kimura1, Yukihiro Kondo1.   

Abstract

INTRODUCTION: We encountered an extremely rare case of a nephrostomy catheter entering the right renal vein during an exchange procedure. CASE
PRESENTATION: An 80-year-old man underwent radical cystectomy. Urinary diversion was achieved through right percutaneous nephrostomy. After the 15th nephrostomy catheter exchange, the patient bled heavily from the catheter. We clamped the catheter immediately, and the patient became hemodynamically stable. Emergency angiography showed the nephrostomy catheter entering the renal vein from outside the renal pelvis. Under fluoroscopy, we pulled the catheter until its tip was located in the previous penetration site of the renal pelvic wall and inserted the catheter over the guidewire into the renal pelvis.
CONCLUSIONS: Herein, we report an extremely rare case of a nephrostomy catheter inserted into the right renal vein during an exchange procedure. Inserting a nephrostomy catheter in the appropriate position and performing exchange under imaging guidance techniques could help clinicians avoid severe complications.
© 2021 The Authors. IJU Case Reports published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Urological Association.

Entities:  

Keywords:  complications; hemorrhage; nephrostomy

Year:  2021        PMID: 33977251      PMCID: PMC8088896          DOI: 10.1002/iju5.12276

Source DB:  PubMed          Journal:  IJU Case Rep        ISSN: 2577-171X


computed tomography inferior vena cava percutaneous nephrolithotomy We have reported an extremely rare case of a nephrostomy catheter entering the renal vein during an exchange procedure. Inserting a nephrostomy catheter in the appropriate position and performing exchange under imaging guidance techniques could avoid severe complications.

Introduction

Percutaneous nephrostomy is a highly effective interventional procedure most commonly performed to relieve hydronephrosis. There are various possible complications of this procedure, including hemorrhage, sepsis, and pneumothorax. One of the most severe complications is penetration into the vascular system by the catheter, of which there are few reports in the literature. , , , , , , , , Most reported complications occur following PCNL. , , , , , There are few reports of such a complication during catheter exchange. , We report an extremely rare case of a nephrostomy catheter entering the vascular system during an exchange procedure.

Case presentation

An 80‐year‐old man had undergone with a nephrostomy for post‐renal renal failure due to a bladder tumor around the right ureteral orifice with a contralateral atrophic kidney. Thereafter, the patient underwent radical cystectomy because of a diagnosis of invasive bladder cancer, and nephrostomy was chosen for urinary diversion owing to the patient’s preference. A year later, in the outpatient clinic, we performed routine exchange of his 18‐Fr silicon catheter, but the urinary tract was resistant to the exchange. We attempted insertion again by switching from the 18‐Fr to a 14‐Fr silicon catheter, while paying attention to the catheter’s depth of insertion and its resistance to insertion. All of these procedures were performed without medical equipment such as guidewire. After these procedures, we noticed a discharge that contained blood. We instructed the patient to remain in bed and, 10 min after the procedure, we noticed continuous bleeding from the catheter. The patient subsequently went into hypovolemic shock. We immediately clamped the catheter and started rapid intravenous saline infusion. After confirming the patient's hemodynamic stability, emergency renal angiography was performed. We located the inflated balloon of the nephrostomy catheter inside the renal vein (Fig. 1a,b). On conducting abdominal CT after angiography, it was evident that the nephrostomy catheter had strayed into the renal vein from the right ventral renal pelvis, with the contrast agent used in the angiography visible only in the renal pelvis. No extravasation of urine or bleeding from the renal pelvis was observed (Fig. 2a). Compared with the previous CT, the nephrostomy catheter had dislodged into the interlobar veins of the kidney (Fig. 2b). We performed a recovery procedure under general anesthesia in the operating room. Under fluoroscopy, the renal pelvis was filled with the contrast agent, which had been used during the angiography. We discovered that the nephrostomy catheter had entered the interlobar vein from the outside the renal pelvis (Fig. 3a). We pulled the catheter carefully using a guidewire until the catheter tip was within the penetration site of the renal pelvic wall (Fig. 3b). We advanced the guidewire into the renal pelvic space through the previous tract, inserted the catheter over the guidewire into the renal pelvis, and fixed it at the position of the puncture site in the renal pelvis (Fig. 3c). The inflated balloon compressed the damaged interlobar vein from the side of the renal pelvis. After releasing the clamp on the nephrostomy catheter, the flow of transparent yellow urine was confirmed. Fifteen days after the event, we replaced the nephrostomy catheter with a new one into a lower calyx through the renal papilla.
Fig. 1

Renal angiography following insertion of the catheter into the renal vein. (a) Renal arteriography. Renal arteriography demonstrated no arterial bleeding. (b) Renal venography. On the emergency renal venography, the nephrostomy catheter’s dilated balloon is observed in the renal vein.

Fig. 2

Abdominal CT scan. (a) Abdominal CT scan shows the nephrostomy catheter straying into the renal vein outside the renal pelvis. The contrast agent used in the angiography is seen only in the renal pelvis, and no extravasation of urine or bleeding from the renal pelvis is observed. (b) Abdominal CT scan before this event shows the interlobar veins of the kidney at the right ventral renal pelvis.

Fig. 3

Anticipated route of the catheter into the renal vein and the recovery procedure for this event in the present case. (a) Schematic diagram of the anticipated route of the nephrostomy catheter into the renal vein: we speculated that the catheter might have entered the renal vein via the interlobar vein, through the outside of the renal pelvis. Schematic diagram of the recovery procedure for this event. (b) The catheter was drawn out along a guidewire and was stopped at the original site at which the pelvic wall had been penetrated. (c) We advanced the guidewire into the renal pelvic space through the previous tract and inserted the catheter over the guidewire into the renal pelvis. Finally, the catheter was fixed at the puncture site in the renal pelvis.

Renal angiography following insertion of the catheter into the renal vein. (a) Renal arteriography. Renal arteriography demonstrated no arterial bleeding. (b) Renal venography. On the emergency renal venography, the nephrostomy catheter’s dilated balloon is observed in the renal vein. Abdominal CT scan. (a) Abdominal CT scan shows the nephrostomy catheter straying into the renal vein outside the renal pelvis. The contrast agent used in the angiography is seen only in the renal pelvis, and no extravasation of urine or bleeding from the renal pelvis is observed. (b) Abdominal CT scan before this event shows the interlobar veins of the kidney at the right ventral renal pelvis. Anticipated route of the catheter into the renal vein and the recovery procedure for this event in the present case. (a) Schematic diagram of the anticipated route of the nephrostomy catheter into the renal vein: we speculated that the catheter might have entered the renal vein via the interlobar vein, through the outside of the renal pelvis. Schematic diagram of the recovery procedure for this event. (b) The catheter was drawn out along a guidewire and was stopped at the original site at which the pelvic wall had been penetrated. (c) We advanced the guidewire into the renal pelvic space through the previous tract and inserted the catheter over the guidewire into the renal pelvis. Finally, the catheter was fixed at the puncture site in the renal pelvis.

Discussion

Ours is the 14th reported case of a misplaced nephrostomy catheter in the vascular system and the third to have occurred during an exchange procedure (Table 1). , , , , , , , , In 10 of 14 cases, the misplacement occurred after PCNL. Mazzucchi et al. reported two cases of intravenous misplacement of the nephrostomy catheter after PCNL as uncommon complications and provided details of their management. Chen et al. reported three cases, and the incidence of intravenous misplacement of the nephrostomy catheter after PCNL was 0.5% (2/4148 cases). However, there are a few reports of this complication occurring during catheter exchange. Dias‐Filho et al. reported a case of inadvertent renal vein penetration during nephrostomy catheter exchange, with right atrial migration of the catheter. Kotb et al. reported a case of percutaneous silicon catheter insertion into the IVC. They concluded that a nephrostomy catheter should be exchanged under ultrasound or fluoroscopic guidance and argued against inserting any catheter into the kidney in a blind manner, even into a mature tract. We misplaced the nephrostomy catheter, which entered the renal vein, with a blinded technique during the 15th exchange.
Table 1

Reported cases of nephrostomy catheter misplacement

NoAuthorYearAgeSexSideLocationOccasionOperation type
1Dias‐Filho et al. 8 200563FLRenal vein/IVC/right atriumExchangingNoninvasive procedure
2Shaw et al. 4 200554MRRenal veinPCNLExploratory
3Mazzucchi et al. 2 200952MLRenal veinPCNLNoninvasive procedure
4Mazzucchi et al. 2 200935FLRenal vein/IVCPCNLNoninvasive procedure
5Li et al. 5 201332FLRenal vein/IVCPCNLNoninvasive procedure
6Kotb et al. 9 201350MLRenal vein/IVCExchangingOpen pyelotomy
7Wang et al. 6 201366FLRenal veinPCNLNoninvasive procedure
8Chen et al. 3 201442MLRenal vein/IVCPCNLNoninvasive procedure
9Chen et al. 3 201438FLRenal vein/IVCPCNLNoninvasive procedure
10Chen et al. 3 201448MLRenal veinPCNLNoninvasive procedure
11Al Zahrani et al. 10 201676FRRenal vein/IVCInsertionNoninvasive procedure
12Fu et al. 7 201768MRRenal veinPCNLExploratory laparotomy
13Fu et al. 7 201728MLRenal vein/IVCPCNLExploratory laparotomy
14Our case202180MRRenal veinExchangingNoninvasive procedure

F, female; M, male; L, left; R, right.

Reported cases of nephrostomy catheter misplacement F, female; M, male; L, left; R, right. Several reports have described using a relatively noninvasive procedure to prevent further complications, with a suggested management as follows: once intravenous misplacement is detected, the catheter should be clamped immediately, and the clamped catheter should be drawn back and repositioned in the collecting system under image guidance. Zahrani et al. reported a case of obstructive uropathy in a solitary kidney where placement of a nephrostomy catheter was complicated by its insertion into the IVC. They performed controlled removal of the misplaced nephrostomy catheter utilizing the percutaneous intravenous balloon tamponade technique. However, Fu et al. performed open surgery under general anesthesia, mainly to remove residual calculi and to prepare for any possible adverse events. In our case, a CT scan showed no bleeding into the renal pelvis or the retroperitoneal space. Therefore, the nephrostomy catheter was drawn outward, and the tip was adjusted to a position in the renal pelvis adjacent to the site of injury, using fluoroscopy under general anesthesia. As a result, we avoided further severe complications. We considered the following issues in our management. First, the initial insertion position was inadequate. This insertion was made not through the renal papilla but directly into the renal pelvis. This procedure is risky and not recommended due to the possibility of vessel damage between the renal parenchyma and the renal pelvis. Inserting the catheter in the appropriate site not only provides good drainage but also reduces complications during placement and the exchange procedure. Second, even when the soft silicone catheter is replaced, exchanging the catheter using the blinded technique can lead to its insertion of the catheter into the incorrect area. Due to the long‐term nephrostomy management, the renal tissues in contact with the catheter may have become fragile. We have reported an extremely rare case regarding a misplaced nephrostomy catheter extending into the renal vein during an exchange procedure. Even during an exchange procedure, intravenous misplacement of the nephrostomy catheter can occur. To avoid this severe complication, the catheter must be inserted at the appropriate position and exchanged under the guidance of imaging techniques such as ultrasound or fluoroscopy.

Conflict of interest

The authors declare no conflict of interest.
  10 in total

1.  Quality improvement guidelines for percutaneous nephrostomy.

Authors:  P Ramchandani; J F Cardella; C J Grassi; A C Roberts; D Sacks; M S Schwartzberg; C A Lewis
Journal:  J Vasc Interv Radiol       Date:  2001-11       Impact factor: 3.464

2.  Right atrial migration of nephrostomy catheter.

Authors:  Aderivaldo C Dias-Filho; Guilherme A V Coaracy; Wallace Borges
Journal:  Int Braz J Urol       Date:  2005 Sep-Oct       Impact factor: 1.541

3.  Management of renal-vein perforation during a challenging percutaneous nephrolithotomy.

Authors:  G Shaw; T M Wah; M J Kellett; S K S Choong
Journal:  J Endourol       Date:  2005 Jul-Aug       Impact factor: 2.942

Review 4.  Intravenous misplacement of nephrostomy tube following percutaneous nephrolithotomy: Three new cases and review of seven cases in the literature.

Authors:  Xiao-Feng Chen; Shan-Qun Chen; Liang-Yu Xu; Ye Gong; Zhuang-Fei Chen; Shao-Bin Zheng
Journal:  Int Braz J Urol       Date:  2014 Sep-Oct       Impact factor: 1.541

5.  Percutaneous silicon catheter insertion into the inferior vena cava, following percutaneous nephrostomy exchange.

Authors:  Ahmed Fouad Kotb; Ahmed Elabbady; Khaled Refaai Mohamed; Mohamed Adel Atta
Journal:  Can Urol Assoc J       Date:  2013 Jul-Aug       Impact factor: 1.862

Review 6.  Management of intravenous migration of urologic catheter.

Authors:  Dongjie Li; Lun Xiao; Zhengyan Tang; Lin Qi; Kuan Luo; Liang Huang; Ke Huang; Jie Zhou; Yongxiang Tang; Jiada Li
Journal:  Urology       Date:  2013-05-01       Impact factor: 2.649

7.  The use of endo-vascular balloon tamponade technique for the removal of a misplaced nephrostomy tube in the inferior vena cava: A case report.

Authors:  Yousof Al Zahrani; Sultan R AlHarbi; Daniele Wiseman
Journal:  Int J Surg Case Rep       Date:  2016-07-29

Review 8.  Intravenous misplacement of the nephrostomy catheter following percutaneous nephrostolithotomy: two case reports and literature review.

Authors:  Weijin Fu; Zhanbin Yang; Zhibin Xie; Haibiao Yan
Journal:  BMC Urol       Date:  2017-06-14       Impact factor: 2.264

9.  Intravenous misplacement of the nephrostomy catheter following percutaneous nephrostolithotomy: two case reports.

Authors:  Eduardo Mazzucchi; Anuar Mitre; Artur Brito; Marco Arap; Claudio Murta; Miguel Srougi
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365

10.  Metachronous renal vein and artery injure after percutaneous nephrostolithotomy.

Authors:  Chaojun Wang; Shanwen Chen; Fuqing Tang; Baihua Shen
Journal:  BMC Urol       Date:  2013-12-05       Impact factor: 2.264

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.