Microwave ablation of lung tumors: A probabilistic approach for simulation-based treatment planning.

Jan Sebek1,2, Pinyo Taeprasartsit3,4, Henky Wibowo3, Warren L Beard5, Radoslav Bortel2, Punit Prakash1.   

Abstract

PURPOSE: Microwave ablation (MWA) is a clinically established modality for treatment of lung tumors. A challenge with existing application of MWA, however, is local tumor progression, potentially due to failure to establish an adequate treatment margin. This study presents a robust simulation-based treatment planning methodology to assist operators in comparatively assessing thermal profiles and likelihood of achieving a specified minimum margin as a function of candidate applied energy parameters.
METHODS: We employed a biophysical simulation-based probabilistic treatment planning methodology to evaluate the likelihood of achieving a specified minimum margin for candidate treatment parameters (i.e., applied power and ablation duration for a given applicator position within a tumor). A set of simulations with varying tissue properties was evaluated for each considered combination of power and ablation duration, and for four different scenarios of contrast in tissue biophysical properties between tumor and normal lung. A treatment planning graph was then assembled, where distributions of achieved minimum ablation zone margins and collateral damage volumes can be assessed for candidate applied power and treatment duration combinations. For each chosen power and time combination, the operator can also visualize the histogram of ablation zone boundaries overlaid on the tumor and target volumes. We assembled treatment planning graphs for generic 1, 2, and 2.5 cm diameter spherically shaped tumors and also illustrated the impact of tissue heterogeneity on delivered treatment plans and resulting ablation histograms. Finally, we illustrated the treatment planning methodology on two example patient-specific cases of tumors with irregular shapes.
RESULTS: The assembled treatment planning graphs indicate that 30 W, 6 min ablations achieve a 5-mm minimum margin across all simulated cases for 1-cm diameter spherical tumors, and 70 W, 10 min ablations achieve a 3-mm minimum margin across 90% of simulations for a 2.5-cm diameter spherical tumor. Different scenarios of tissue heterogeneity between tumor and lung tissue revealed 2 min overall difference in ablation duration, in order to reliably achieve a 4-mm minimum margin or larger each time for 2-cm diameter spherical tumor.
CONCLUSIONS: An approach for simulation-based treatment planning for microwave ablation of lung tumors is illustrated to account for the impact of specific geometry of the treatment site, tissue property uncertainty, and heterogeneity between the tumor and normal lung.
© 2021 American Association of Physicists in Medicine.

Entities:  

Keywords:  ablation minimum margin; ablation treatment planning; lung ablation; microwave ablation

Mesh:

Year:  2021        PMID: 33964020      PMCID: PMC8319071          DOI: 10.1002/mp.14923

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.506


  42 in total

1.  Volumetric 3D assessment of ablation zones after thermal ablation of colorectal liver metastases to improve prediction of local tumor progression.

Authors:  Elena A Kaye; Francois H Cornelis; Elena N Petre; Neelam Tyagi; Waleed Shady; Weiji Shi; Zhigang Zhang; Stephen B Solomon; Constantinos T Sofocleous; Jeremy C Durack
Journal:  Eur Radiol       Date:  2018-11-06       Impact factor: 5.315

2.  Interactive multi-criteria planning for radiofrequency ablation.

Authors:  Christian Schumann; Christian Rieder; Sabrina Haase; Katrin Teichert; Philipp Süss; Peter Isfort; Philipp Bruners; Tobias Preusser
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-04-23       Impact factor: 2.924

3.  In vivo percutaneous microwave ablation in kidneys: Correlation with ex vivo data and ablation work.

Authors:  C Marcelin; J Leiner; A Nasri; F Petitpierre; Y Le Bras; M Yacoub; N Grenier; J C Bernhard; F Cornelis
Journal:  Diagn Interv Imaging       Date:  2017-10-21       Impact factor: 4.026

Review 4.  Bronchoscopic Ablative Therapies for Malignant Central Airway Obstruction and Peripheral Lung Tumors.

Authors:  Udit Chaddha; D Kyle Hogarth; Septimiu Murgu
Journal:  Ann Am Thorac Soc       Date:  2019-10

5.  Phenotype molding of stromal cells in the lung tumor microenvironment.

Authors:  Diether Lambrechts; Els Wauters; Bram Boeckx; Sara Aibar; David Nittner; Oliver Burton; Ayse Bassez; Herbert Decaluwé; Andreas Pircher; Kathleen Van den Eynde; Birgit Weynand; Erik Verbeken; Paul De Leyn; Adrian Liston; Johan Vansteenkiste; Peter Carmeliet; Stein Aerts; Bernard Thienpont
Journal:  Nat Med       Date:  2018-07-09       Impact factor: 53.440

6.  Accounting for Deformation in Deep Brain Stimulation Surgery With Models: Comparison to Interventional Magnetic Resonance Imaging.

Authors:  Ma Luo; Paul S Larson; Alastair J Martin; Michael I Miga
Journal:  IEEE Trans Biomed Eng       Date:  2020-02-14       Impact factor: 4.756

Review 7.  Percutaneous thermal ablation of primary lung cancer.

Authors:  T de Baere; L Tselikas; V Catena; X Buy; F Deschamps; J Palussière
Journal:  Diagn Interv Imaging       Date:  2016-09-28       Impact factor: 4.026

8.  Differences in the dielectric properties of various benign and malignant thyroid nodules.

Authors:  Shengyi Huang; Weizhen Cai; Shuai Han; Yu Lin; Yu Wang; Fei Chen; Guoli Shao; Yonghong Liu; Xuefei Yu; Zhai Cai; Zenan Zou; Shun Yao; Qiaohui Wang; Zhou Li
Journal:  Med Phys       Date:  2020-12-16       Impact factor: 4.071

9.  CT Densitometry and Morphology of Radiofrequency-Ablated Stage IA Non-Small Cell Lung Cancer: Results from the American College of Surgeons Oncology Group Z4033 (Alliance) Trial.

Authors:  Erica S Alexander; Lillian Xiong; Grayson L Baird; Hiran Fernando; Damian E Dupuy
Journal:  J Vasc Interv Radiol       Date:  2020-01-03       Impact factor: 3.464

10.  A theoretical model of the application of RF energy to the airway wall and its experimental validation.

Authors:  Jerry Jarrard; Bill Wizeman; Robert H Brown; Wayne Mitzner
Journal:  Biomed Eng Online       Date:  2010-11-27       Impact factor: 2.819

View more
  1 in total

Review 1.  Clinical application status and prospect of the combined anti-tumor strategy of ablation and immunotherapy.

Authors:  Li Yin; Xing-Yu Li; Lin-Lin Zhu; Gui-Lai Chen; Zhuo Xiang; Qing-Qing Wang; Jing-Wang Bi; Qiang Wang
Journal:  Front Immunol       Date:  2022-09-05       Impact factor: 8.786

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.