Yuechen Qiao1, My Yang2, Ian A Marabella1, Devin A J McGee1, Bernard A Olson1, Montserrat Torremorell2, Christopher J Hogan1.
Abstract
Recirculating air purification technologies are employed as potential means of reducing exposure to aerosol particles and airborne viruses. Toward improved testing of recirculating air purification units, we developed and applied a medium-scale single-pass wind tunnel test to examine the size-dependent collection of particles and the collection and inactivation of viable bovine coronavirus (BCoV, a betacoronavirus), porcine respiratory coronavirus (PRCV, an alphacoronavirus), and influenza A virus (IAV), by a commercial air purification unit. The tested unit, the Molekule Air Mini, incorporates a MERV 16 filter as well as a photoelectrochemical oxidating layer. It was found to have a collection efficiency above 95.8% for all tested particle diameters and flow rates, with collection efficiencies above 99% for supermicrometer particles with the minimum collection efficiency for particles smaller than 100 nm. For all three tested viruses, the physical tracer-based log reduction was near 2.0 (99% removal). Conversely, the viable virus log reductions were found to be near 4.0 for IAV, 3.0 for BCoV, and 2.5 for PRCV, suggesting additional inactivation in a virus family- and genus-specific manner. In total, this work describes a suite of test methods which can be used to rigorously evaluate the efficacy of recirculating air purification technologies.Entities:
Keywords: air purifier; coronavirus; filtration; influenza; photoelectrochemical oxidation; virus aerosols; wind tunnel test
Mesh:
Substances:
Year: 2021 PMID: 33960547 PMCID: PMC8242653 DOI: 10.1111/ina.12847
Source DB: PubMed Journal: Indoor Air ISSN: 0905-6947 Impact factor: 6.554
FIGURE 1A schematic diagram of the wind tunnel system utilized in penetration and log reduction measurements with the air purifier unit installed (A). Labeled photograph of the wind tunnel system within a biosafety level II room (B)
FIGURE 2Particle penetration versus particle diameter for the air purification unit, measured in a sealed wind tunnel. Solid lines denoted measurements utilizing a scanning mobility particle sizer, while dashed lines denote optical particle sizer results. The error bars are the standard deviation according to the four penetration values at each particle size determined following ASHRAE Standard 52.2 guidelines
A summary of fluorimetry, virus titer, and RT‐qPCR test results for pooled Anderson impactor stages upstream and downstream of the air purification unit in the wind tunnel
| Replicate | BCoV | PRCV | IAV | |||
|---|---|---|---|---|---|---|
| Upstream | Downstream | Upstream | Downstream | Upstream | Downstream | |
| Fluorescein Correlation Control Tests [a.u.] | ||||||
| 1 | 9.20 × 102 | 1.07 × 103 | 6.96 × 102 | 9.29 × 102 | 5.24 × 102 | 1.13 × 103 |
| 2 | 1.00 × 103 | 1.08 × 103 | 7.45 × 102 | 3.70 × 102 | 1.12 × 103 | 1.15 × 103 |
| 3 | 1.06 × 103 | 1.08 × 103 | 7.27 × 102 | 7.51 × 102 | 1.07 × 103 | 1.17 × 103 |
| Average | 9.95 × 102 | 1.08 × 103 | 7.23 × 102 | 6.83 × 102 | 9.03 × 102 | 1.15 × 103 |
| Fluorescein Penetration Tests [a.u.] | ||||||
| 1 | 1.05 × 103 | 9.00 | 7.31 × 102 | 4.00 | 9.33 × 102 | 4.00 |
| 2 | 1.24 × 103 | 10.00 | 9.97 × 102 | 6.00 | 7.53 × 102 | 3.00 |
| 3 | 1.01 × 103 | 9.00 | 8.62 × 102 | 4.00 | 7.22 × 102 | 5.00 |
| Average | 1.10 × 103 | 9.33 | 8.63 × 102 | 4.67 | 8.03 × 102 | 4.00 |
| Virus Titer Correlation Control Tests [TCID50 ml−1] | ||||||
| 1 | 1.78 × 106 | 3.16 × 106 | 1.78 × 103 | 5.62 × 103 | 3.16 × 106 | 5.62 × 106 |
| 2 | 1.48 × 106 | 1.78 × 106 | 3.16 × 103 | 5.62 × 103 | 3.16 × 106 | 3.16 × 106 |
| 3 | 3.16 × 106 | 3.16 × 106 | 1.00 × 103 | 5.62 × 103 | 1.78 × 106 | 1.00 × 106 |
| Average | 2.14 × 106 | 2.70 × 106 | 1.98 × 103 | 5.62 × 103 | 2.70 × 106 | 3.26 × 106 |
| Virus Titer Penetration Tests [TCID50 ml−1] | ||||||
| 1 | 1.00 × 106 | 5.62 × 102 | 1.78 × 104 | 3.16 × 101 | 1.78 × 106 | 1.78 × 102 |
| 2 | 3.16 × 105 | 5.62 × 102 | 1.78 × 104 | 5.62 × 101 | 1.00 × 106 | 5.62 × 101 |
| 3 | 5.62 × 105 | 5.62 × 102 | 5.62 × 103 | 5.62 × 101 | 1.00 × 105 | 1.00 × 102 |
| Average | 6.26 × 105 | 5.62 × 102 | 1.37 × 104 | 4.80 × 101 | 9.59 × 105 | 1.11 × 102 |
| RT‐qPCR Correlation Control Tests [Copies ml−1] | ||||||
| 1 | 3.60 × 108 | 4.26 × 108 | 1.58 × 107 | 2.13 × 107 | 2.88 × 109 | 2.62 × 109 |
| 2 | 4.39 × 108 | 4.82 × 108 | 1.40 × 107 | 1.73 × 107 | 2.66 × 109 | 2.39 × 109 |
| 3 | 4.36 × 108 | 4.70 × 108 | 1.36 × 107 | 1.58 × 107 | 2.50 × 109 | 3.08 × 109 |
| Average | 4.12 × 108 | 4.59 × 108 | 1.45 × 107 | 1.81 × 107 | 2.68 × 109 | 2.70 × 109 |
| RT‐qPCR Penetration Tests [Copies ml−1] | ||||||
| 1 | 3.51 × 108 | 3.50 × 105 | 3.53 × 107 | 4.99 × 104 | 1.64 × 109 | 1.62 × 106 |
| 2 | 4.90 × 108 | 6.10 × 105 | 5.35 × 107 | 7.87 × 104 | 8.23 × 108 | 5.79 × 105 |
| 3 | 2.60 × 108 | 4.23 × 105 | 3.80 × 107 | 7.31 × 104 | 1.33 × 109 | 1.26 × 106 |
| Average | 3.67 × 108 | 4.61 × 105 | 4.22 × 107 | 6.72 × 104 | 1.27 × 109 | 1.15 × 106 |
Measurements were carried out in triplicate, both without the air purification unit installed (control/correlation tests) and with the air purification unit installed (penetration tests).
Abbreviations: BCoV, bovine coronavirus; IAV, influenza A virus; PRCV, porcine respiratory coronavirus.
Values of 3.16 × 101 correspond to virus titer levels below the limit of detection, with this value utilized as the upper limit of the possible virus titer.
FIGURE 3Log reductions in virus concentrations (log10[upstream concentration/downstream concentration] with concentration based upon fluorescein tracer (physical removal), RT‐qPCR (RNA physical removal and damage), and virus titration (viable viruses) for bovine coronavirus (BCoV), porcine respiratory coronavirus (PRCV), influenza A virus (IAV). The error bars are the standard deviation of the log reduction calculated by root sum square error propagation method applying the standard deviations of the upstream and downstream triplicate measurements
Summary of RT‐qPCR, virus titer results of bovine coronavirus (BCoV), porcine respiratory coronavirus (PRCV), and influenza A virus (IAV) for extraction from inside and outside of the photoelectrochemical oxidation (PECO) filters during operation of the air purification unit
| Virus | RT‐qPCR [Copies ml−1] | Titer [TCID50 ml−1] | ||
|---|---|---|---|---|
| Inside | Outside | Inside | Outside | |
| BCoV | 9.38 × 104 | 8.49 × 108 | 3.16 × 101
| 1.78 × 106 |
| PRCV | 9.25 × 103 | 2.15 × 108 | 3.16 × 101
| 5.62 × 104 |
| IAV | 3.47 × 104 | 5.31 × 109 | 3.16 × 101
| 5.62 × 105 |
The filters were loaded for 90 min prior to sampling for each virus.
Values of 3.16 × 101 correspond to virus titer levels below the limit of detection, with this value utilized as the upper limit of the possible virus titer.
FIGURE 4Images of 1 in × 2 in cutout strips of photoelectrochemical oxidation (PECO) filter laminates and HEPA filter media used in virus surface inactivation studies (A). Log reduction over time on various filter media based upon bovine coronavirus titer (B). Log reduction over time on various filter media based upon bovine coronavirus RT‐qPCR (C). Log reduction was calculated based on the average values of three replicates at different sampling times, comparing with initial time (t = 0) in both titer and RT‐qPCR measurements. Error bars were calculated using the root sum square error propagation method with the standard deviations for triplicate measurements at the initial time and the sampling time in titer and RT‐qPCR considered, respectively