Literature DB >> 33957798

A comparison between the CS-TOF and the CTA/DSA for WEB device management.

Oktay Algin1,2,3, Gokhan Yuce3, Ural Koc3,4, Gıyas Ayberk5.   

Abstract

PURPOSE: There is no study on the role of three-dimensional compressed sensing time of flight MR angiography (3D-CS-TOF) in the management of the WEB device. We evaluated the efficacy of 3-tesla 3D-CS-TOF for the management and follow-up of the WEB device implantations.
MATERIALS AND METHODS: Seventy-three aneurysms of 69 patients treated with the WEB device were retrospectively examined. Morphological parameters and embolization results of the aneurysms were assessed and compared on 3D-CS-TOF, CTA, and DSA images.
RESULTS: Occluded, neck remnant, and recurrent aneurysms were observed in 61 (83.6%), 7 (9.6%), and 5 (6.8%) aneurysms, respectively. Inter- and intra-reader agreement values related to aneurysm size measurements were perfect. Aneurysms size, age, and proximal vessel tortuosity were negatively correlated with the visibility of the aneurysms and parent vessels on 3D-CS-TOF images (p = 0.043; p = 0.032; p < 0.001, respectively). Subarachnoid hemorrhage and age are associated with 3D-CS-TOF artifacts (p = 0.031; p = 0.005, respectively). 3D-CS-TOF findings are in perfect agreement with DSA or CT angiography (CTA) results (p < 0.001).
CONCLUSION: According to our results, 3D-CS-TOF can be an easy, fast, and reliable alternative for the management or follow-up of WEB assisted embolization.

Entities:  

Keywords:  DSA; TOF; WEB; cerebral aneurysm; compressed sensing; non-contrast MR angiography

Mesh:

Year:  2021        PMID: 33957798      PMCID: PMC8905082          DOI: 10.1177/15910199211014708

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  15 in total

1.  Usefulness of Silent MR Angiography for Intracranial Aneurysms Treated with a Flow-Diverter Device.

Authors:  H Oishi; T Fujii; M Suzuki; N Takano; K Teranishi; K Yatomi; T Kitamura; M Yamamoto; S Aoki; H Arai
Journal:  AJNR Am J Neuroradiol       Date:  2019-05-02       Impact factor: 3.825

2.  The New Low-Profile WEB 17 System for Treatment of Intracranial Aneurysms: First Clinical Experiences.

Authors:  S B T van Rooij; J P Peluso; M Sluzewski; H G Kortman; W J van Rooij
Journal:  AJNR Am J Neuroradiol       Date:  2018-03-22       Impact factor: 3.825

3.  Clinical feasibility study of 3D intracranial magnetic resonance angiography using compressed sensing.

Authors:  Zhiyong Lin; Xiaodong Zhang; Li Guo; Ke Wang; Yuan Jiang; Xiaoyu Hu; Yong Huang; Juan Wei; Shuai Ma; Yi Liu; Lina Zhu; Zhizheng Zhuo; Jing Liu; Xiaoying Wang
Journal:  J Magn Reson Imaging       Date:  2019-04-13       Impact factor: 4.813

4.  WEB-only treatment of ruptured and unruptured intracranial aneurysms: a retrospective analysis of 47 aneurysms.

Authors:  Yigit Ozpeynirci; Michael Braun; Andrej Pala; Melanie Schick; Bernd Schmitz
Journal:  Acta Neurochir (Wien)       Date:  2019-06-25       Impact factor: 2.216

Review 5.  Diagnostic value of 3D time-of-flight magnetic resonance angiography for detecting intracranial aneurysm: a meta-analysis.

Authors:  Liu HaiFeng; Xu YongSheng; Xun YangQin; Dou Yu; Wang ShuaiWen; Lu XingRu; Lei JunQiang
Journal:  Neuroradiology       Date:  2017-09-08       Impact factor: 2.804

6.  Preoperative Digital Subtraction Angiography in Incidental Unruptured Intracranial Aneurysms : How Much is too Much?

Authors:  Moriz Herzberg; Robert Forbrig; Christian Schichor; Hartmut Brückmann; Franziska Dorn
Journal:  Clin Neuroradiol       Date:  2018-04-24       Impact factor: 3.649

7.  Woven Endobridge (WEB) device for endovascular treatment of complex unruptured aneurysms-a single center experience.

Authors:  Stephanie Lescher; Richard du Mesnil de Rochemont; Joachim Berkefeld
Journal:  Neuroradiology       Date:  2016-01-15       Impact factor: 2.804

Review 8.  How to WEB: a practical review of methodology for the use of the Woven EndoBridge.

Authors:  Nitin Goyal; Daniel Hoit; Julie DiNitto; Lucas Elijovich; David Fiorella; Laurent Pierot; Saleh Lamin; Laurent Spelle; Isil Saatci; Saru Cekirge; Adam S Arthur
Journal:  J Neurointerv Surg       Date:  2020-01-31       Impact factor: 5.836

9.  Treatment of Acutely Ruptured Cerebral Aneurysms With the Woven EndoBridge Device: Experience Post-FDA Approval.

Authors:  Fadi Al Saiegh; David Hasan; Nikolaos Mouchtouris; Mario Zanaty; Ahmad Sweid; Omaditya Khanna; Nohra Chalouhi; Ritam Ghosh; Stavropoula Tjoumakaris; M Reid Gooch; Robert Rosenwasser; Pascal Jabbour
Journal:  Neurosurgery       Date:  2020-07-01       Impact factor: 4.654

10.  3D Rotational Angiography in Follow-Up of Clipped Intracranial Aneurysms.

Authors:  Subhash Kumar; Shailesh B Gaikwad; Nalini Kant Mishra
Journal:  ISRN Radiol       Date:  2014-01-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.