Jacob M Wilson1, Jeremiah A Pflederer1, Andrew M Schwartz1, Kevin X Farley1, Nickolas B Reimer1.
Abstract
BACKGROUND: Intraoperative vascular injury during total hip arthroplasty represents a catastrophic complication. Acetabular screw placement represents one possible mode of injury. The purpose of this study was to evaluate the utility of various fluoroscopic views in the detection of intrapelvic screw penetration.Entities:
Keywords: Acetabular; Arthroplasty; Fluoroscopy; Radiology; Screw; Total hip
Year: 2021 PMID: 33937463 PMCID: PMC8079330 DOI: 10.1016/j.artd.2021.02.011
Source DB: PubMed Journal: Arthroplast Today ISSN: 2352-3441
Figure 1Example of the triple shot series of a screw placed in the anterior-superior quadrant with 10 mm of screw penetration. (a) Iliac oblique, (b) quad view, and (c) obturator outlet. The photos of the model pelvis taken at the trajectory of the fluoroscopy tube to obtain (d) iliac oblique, (e) quad view, and (f) obturator outlet.
Diagnostic performance of each radiographic view by penetration depth.
| Radiograph | Depth | Sensitivity | Specificity |
|---|---|---|---|
| AP | 0 | - | 1.00 |
| 5 | 0.31 | - | |
| 10 | 0.38 | - | |
| DTW | 0 | - | 1.00 |
| 5 | 0.31 | - | |
| 10 | 0.46 | - | |
| Iliac oblique | 0 | - | 1.00 |
| 5 | 0.08 | - | |
| 10 | 0.23 | - | |
| Inlet | 0 | - | 1.00 |
| 5 | 0.54 | - | |
| 10 | 0.62 | - | |
| Obturator oblique | 0 | - | 0.92 |
| 5 | 0.31 | - | |
| 10 | 0.46 | - | |
| Obturator outlet | 0 | - | 0.92 |
| 5 | 0.46 | - | |
| 10 | 0.54 | - | |
| Outlet | 0 | - | 0.92 |
| 5 | 0.38 | - | |
| 10 | 0.46 | - | |
| Quad | 0 | - | 0.85 |
| 5 | 0.62 | - | |
| 10 | 0.62 | - |
AP, anteroposterior; DTW, down the wing.
Diagnostic performance—all screw depths combined.
| Radiograph | Sensitivity | Specificity | PPV | NPV |
|---|---|---|---|---|
| AP | 0.35 | 1.00 | 1.00 | 0.43 |
| DTW | 0.38 | 1.00 | 1.00 | 0.45 |
| Iliac Oblique | 0.15 | 1.00 | 1.00 | 0.37 |
| Inlet | 0.58 | 1.00 | 1.00 | 0.54 |
| Obturator Oblique | 0.38 | 1.00 | 1.00 | 0.45 |
| Obturator Outlet | 0.50 | 0.92 | 0.93 | 0.48 |
| Outlet | 0.42 | 0.92 | 0.92 | 0.44 |
| Quad | 0.62 | 0.85 | 0.89 | 0.52 |
AP, anteroposterior; DTW, down the wing; PPV, positive predictive value; NPV, negative predictive value.
Diagnostic performance of 3-view radiographic series.
| Radiograph | Sensitivity | Specificity | PPV | NPV |
|---|---|---|---|---|
| Anterior inferior all | 1.00 | 1.00 | 1.00 | 1.00 |
| Anterior superior all | 1.00 | 1.00 | 1.00 | 1.00 |
| Posterior inferior all | 0.67 | 1.00 | 1.00 | 0.60 |
| Posterior superior all | 0.86 | 0.67 | 0.86 | 0.67 |
| Anterior inferior 5 | 1.00 | 1.00 | 1.00 | 1.00 |
| Anterior superior 5 | 1.00 | 1.00 | 1.00 | 1.00 |
| Posterior inferior 5 | 0.33 | 1.00 | 1.00 | 0.60 |
| Posterior superior 5 | 0.75 | 0.67 | 0.75 | 0.67 |
| Anterior inferior 10 | 1.00 | 1.00 | 1.00 | 1.00 |
| Anterior superior 10 | 1.00 | 1.00 | 1.00 | 1.00 |
| Posterior inferior 10 | 1.00 | 1.00 | 1.00 | 1.00 |
| Posterior superior 10 | 1.00 | 0.67 | 0.75 | 1.00 |
PPV, positive predictive value; NPV, negative predictive value.