The study was to find the optimal values of priority in the inverse planning module of Leksell GammaPlan which would give better treatment plan indices in GammaKnife SRS. The study showed that the best optimised setting of the weighting or priority in the inverse planning module of Leksell GammaPlan were 0.6 for coverage, 0.3 for gradient index and 0.5 for beam on time. Inverse plans (Hybrid Inverse Plan, HIP) which were made using this optimal priority setting were compared with forward plans (FP) with all 95% coverage. The results showed that the average selectivity index (SI) was 83.05±9.68 for FP and 85.35±8.03 for HIP. So, SI improved in the HIP technique by about 2.3% compare to FP. Similarly, average gradient index (GI) for FP and HIP were respectively 2.82±0.23 and 2.76±0.33. And the average beam on time (BT) of FP and HIP were, respectively, 48.15±23.14 min and 48.35±18.09 min. So, all plan indices show improvement in the hybrid inverse planning technique over forward plans. Consequently, this will improve the quality of patient treatment in GammaKnife.
© 2021 Old City Publishing, Inc.

Entities:  

Keywords:  GammaKnife; forward planning; gradient index; inverse planning; selectivity index

Year:  2021        PMID: 33898088      PMCID: PMC8055234     

Source DB:  PubMed          Journal:  J Radiosurg SBRT


  4 in total

1.  Inverse treatment planning for Gamma Knife radiosurgery.

Authors:  D M Shepard; M C Ferris; R Ove; L Ma
Journal:  Med Phys       Date:  2000-12       Impact factor: 4.071

2.  Real-time inverse planning for Gamma Knife radiosurgery.

Authors:  Q Jackie Wu; Vira Chankong; Suradet Jitprapaikulsarn; Barry W Wessels; Douglas B Einstein; Boonyanit Mathayomchan; Timothy J Kinsella
Journal:  Med Phys       Date:  2003-11       Impact factor: 4.071

Review 3.  Comparison of full width at half maximum and penumbra of different Gamma Knife models.

Authors:  Sepideh Asgari; Nooshin Banaee; Hassan Ali Nedaie
Journal:  J Cancer Res Ther       Date:  2018 Jan-Mar       Impact factor: 1.805

4.  Stereotactic radiosurgery using the 201 cobalt-60 source gamma knife.

Authors:  R J Coffey; L D Lunsford
Journal:  Neurosurg Clin N Am       Date:  1990-10       Impact factor: 2.509

  4 in total
  1 in total

1.  Role of gamma angle in treatment planning of vestibular schwannoma in Gamma Knife: A retrospective study.

Authors:  Ngangom Robert; Manjul Tripathi; Budhi Singh Yadav
Journal:  J Radiosurg SBRT       Date:  2021
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.