| Literature DB >> 33850283 |
Anup Katheria1, Henry C Lee2, Ronny Knol3, Leigh Irvine4, Sumesh Thomas4.
Abstract
There is a large body of evidence demonstrating that delaying clamping of the umbilical cord provides benefits for term and preterm infants. These benefits include reductions in mortality in preterm infants and improved developmental scores at 4 years of age in term infants. However, non-breathing or non-vigorous infants at birth are excluded due to the perceived need for immediate resuscitation. Recent studies have demonstrated early physiological benefits in both human and animal models if resuscitation is performed with an intact cord, but this is still an active area of research. Given the large number of ongoing and planned trials, we have brought together an international group that have been intimately involved in the development or use of resuscitation equipment designed to be used while the cord is still intact. In this review, we will present the benefits and limitations of devices that have been developed or are in use. Published trials or ongoing studies using their respective devices will also be reviewed.Entities:
Mesh:
Year: 2021 PMID: 33850283 PMCID: PMC8042840 DOI: 10.1038/s41372-021-01052-3
Source DB: PubMed Journal: J Perinatol ISSN: 0743-8346 Impact factor: 3.225
Fig. 1Life Start trolley.
Labeled diagram for setup of preterm infant (left panel) and sketch of trolley during vaginal birth (right panel).
Fig. 2Nooma cart.
Labeled computer generated mockup (left panel) and cart use during a cesarean section (right panel).
Fig. 3Inspire Bed.
Sketch of device with addtional elements labeled.
Fig. 4Concord Birth Trolley.
Labeled prototype (left panel). Sketch of trolley during delivery (right panel).
Summary of differences between different trolleys.
| Clinical trials | Commercial availability | Radiant heat | Height-adjustable | Swivel function | Ability to mount additional equipment (suctioning, humidifier, pulse oximeter) | Independent gas supply | Independent electric supply | |
|---|---|---|---|---|---|---|---|---|
| LifeStart | NRIC, Term NRIC, VentFirst Cord-Pilot, PCI-T | Yes (FDA- and CE-approved) | No, requires a chemical mattress in preterms | Yes | No | Yes | Can mount underside tanks (Europe but not in US) | No |
| NOOMA | No | No | No, requires a chemical mattress in preterms | Yes | Yes | Yes | Yes | Yes |
| Inspire | VentFirst | No | No, requires a chemical mattress in preterms | Yes | No | Yes | Yes | Yes |
| Concord | ABC1, ABC2, ABC3, PinC | Europe only (CE-approved) | Yes, radiant heater | Yes | Yes | Yes | Yes | No |