Literature DB >> 33847603

Antibacterial Activity in Iodine-coated Implants Under Conditions of Iodine Loss: Study in a Rat Model Plus In Vitro Analysis.

Ken Ueoka1, Tamon Kabata1, Masaharu Tokoro2, Yoshitomo Kajino1, Daisuke Inoue1, Tomoharu Takagi1, Takaaki Ohmori1, Junya Yoshitani1, Takuro Ueno1, Yuki Yamamuro1, Atsushi Taninaka1, Hiroyuki Tsuchiya1.   

Abstract

BACKGROUND: We developed iodine-coated titanium implants to suppress microbial activity and prevent periprosthetic joint infection (PJI); their efficacy was demonstrated in animal and in vitro models. The iodine content in iodine-coated implants naturally decreases in vivo. However, to our knowledge, the effect of reduced iodine content on the implant's antimicrobial activity has not been evaluated to date. QUESTIONS/PURPOSES: (1) How much does the iodine content on the implant surface decrease after 4 and 8 weeks in vivo in a rat model? (2) What effect does the reduced iodine content have on the antimicrobial effect of the implant against multiple bacteria in an in vitro model?
METHODS: This experiment was performed in two parts: an in vivo experiment to determine attenuation of iodine levels over time in rats, and an in vitro experiment in which we sought to assess whether the reduced iodine content observed in the in vivo experiment was still sufficient to deliver antimicrobial activity against common pathogens seen in PJI. For the in vivo experiment, three types of titanium alloy washers were implanted in rats: untreated (Ti), surface-anodized to produce an oxide film (Ti-O), and with an iodine layer on the oxidation film (Ti-I). The attenuation of iodine levels in rats was measured over time using inductively coupled plasma-mass spectrometry. Herein, only the Ti-I washer was used, with five implanted in each rat that were removed after 4 or 8 weeks. For the 4- and 8-week models, two rats and 15 washers were used. For the in vitro study, to determine the antibacterial effect, three types of washers (Ti, Ti-O, and Ti-I) (nine washers in total) were implanted in each rat. Then, the washers were removed and the antibacterial effect of each washer was examined on multiple bacterial species using the spread plate method and fluorescence microscopy. For the spread plate method, six rats were used, and five rats were used for the observation using fluorescence microscopy; further, 4- and 8-week models were made for each method. Thus, a total of 22 rats and 198 washers were used. Live and dead bacteria in the biofilm were stained, and the biofilm coverage percentage for quantitative analysis was determined using fluorescence microscopy in a nonblinded manner. Ti-I was used as the experimental group, and Ti and Ti-O were used as control groups. The total number of rats and washers used throughout this study was 24 and 213, respectively.
RESULTS: Iodine content in rats implanted with Ti-I samples decreased to 72% and 65% after the in vivo period of 4 and 8 weeks, respectively (p = 0.001 and p < 0.001, respectively). In the in vitro experiment, the Ti-I implants demonstrated a stronger antimicrobial activity than Ti and Ti-O implants in the 4- and 8-week models. Both the median number of bacterial colonies and the median biofilm coverage percentage with live bacteria on Ti-I were lower than those on Ti or Ti-O implants for each bacterial species in the 4- and 8-week models. There was no difference in the median biofilm coverage percentage of dead bacteria. In the 8-week model, the antibacterial activity using the spread plate method had median (interquartile range) numbers of bacteria on the Ti, Ti-O, and Ti-I implants of 112 (104 to 165) × 105, 147 (111 to 162) × 105, and 55 (37 to 67) × 105 of methicillin-sensitive Staphylococcus aureus (Ti-I versus Ti, p = 0.026; Ti-I versus Ti-O, p = 0.009); 71 (39 to 111) × 105, 50 (44 to 62) × 105, and 26 (9 to 31)× 105 CFU of methicillin-resistant S. aureus (Ti-I versus Ti, p = 0.026; Ti-I versus Ti-O, p = 0.034); and 77 (74 to 83) × 106, 111 (95 to 117) × 106, and 30 (21 to 45) × 106 CFU of Pseudomonas aeruginosa (Ti-I versus Ti, p = 0.004; Ti-I versus Ti-O, p = 0.009). Despite the decrease in the iodine content of Ti-I after 8 weeks, it demonstrated better antibacterial activity against all tested bacteria than the Ti and Ti-O implants.
CONCLUSION: Iodine-coated implants retained their iodine content and antibacterial activity against methicillin-sensitive S. aureus, methicillin-resistant S. aureus, and P. aeruginosa for 8 weeks in vivo in rats. To evaluate the longer-lasting antibacterial efficacy, further research using larger infected animal PJI models with implants in the joints of both males and females is desirable. CLINICAL RELEVANCE: Iodine-coated titanium implants displayed an antibacterial activity for 8 weeks in rats in vivo. Although the findings in a rat model do not guarantee efficacy in humans, they represent an important step toward clinical application.
Copyright © 2021 by the Association of Bone and Joint Surgeons.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 33847603      PMCID: PMC8208413          DOI: 10.1097/CORR.0000000000001753

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.755


  37 in total

Review 1.  The management of peri-prosthetic infection in total joint arthroplasty.

Authors:  A D Toms; D Davidson; B A Masri; C P Duncan
Journal:  J Bone Joint Surg Br       Date:  2006-02

2.  Antimicrobial susceptibility of Staphylococcus aureus and Staphylococcus epidermidis biofilms isolated from infected total hip arthroplasty cases.

Authors:  Seisuke Nishimura; Toshiyuki Tsurumoto; Akihiko Yonekura; Koichi Adachi; Hiroyuki Shindo
Journal:  J Orthop Sci       Date:  2006-01       Impact factor: 1.601

3.  Biomaterial-centered infection: microbial adhesion versus tissue integration.

Authors:  A G Gristina
Journal:  Science       Date:  1987-09-25       Impact factor: 47.728

4.  The feasibility of iodine-supported processing for titanium with different surfaces.

Authors:  Ken Ueoka; Yoshitomo Kajino; Tamon Kabata; Daisuke Inoue; Junya Yoshitani; Takuro Ueno; Yuki Yamamuro; Toshiharu Shirai; Hiroyuki Tsuchiya
Journal:  J Orthop Sci       Date:  2020-03-03       Impact factor: 1.601

5.  Prosthetic joint infection risk after TKA in the Medicare population.

Authors:  Steven M Kurtz; Kevin L Ong; Edward Lau; Kevin J Bozic; Daniel Berry; Javad Parvizi
Journal:  Clin Orthop Relat Res       Date:  2009-08-08       Impact factor: 4.176

6.  Susceptibility of metallic magnesium implants to bacterial biofilm infections.

Authors:  Muhammad Imran Rahim; Manfred Rohde; Bushra Rais; Jan-Marten Seitz; Peter P Mueller
Journal:  J Biomed Mater Res A       Date:  2016-02-24       Impact factor: 4.396

Review 7.  Review of titanium surface modification techniques and coatings for antibacterial applications.

Authors:  H Chouirfa; H Bouloussa; V Migonney; C Falentin-Daudré
Journal:  Acta Biomater       Date:  2018-10-26       Impact factor: 8.947

8.  Inhibition of biofilm formation on iodine-supported titanium implants.

Authors:  Daisuke Inoue; Tamon Kabata; Kaori Ohtani; Yoshitomo Kajino; Toshiharu Shirai; Hiroyuki Tsuchiya
Journal:  Int Orthop       Date:  2017-04-07       Impact factor: 3.075

9.  Higher risk of failure of methicillin-resistant Staphylococcus aureus prosthetic joint infections.

Authors:  Cassandra D Salgado; Sangita Dash; J Robert Cantey; Camelia E Marculescu
Journal:  Clin Orthop Relat Res       Date:  2007-08       Impact factor: 4.176

10.  Innovative antimicrobial coating of titanium implants with iodine.

Authors:  Hiroyuki Tsuchiya; Toshiharu Shirai; Hideji Nishida; Hideki Murakami; Tamon Kabata; Norio Yamamoto; Koji Watanabe; Junsuke Nakase
Journal:  J Orthop Sci       Date:  2012-07-18       Impact factor: 1.601

View more
  1 in total

1.  CORR Insights®: Antibacterial Activity in Iodine-coated Implants Under Conditions of Iodine Loss: Study in a Rat Model Plus In Vitro Analysis.

Authors:  Curtis W Hartman
Journal:  Clin Orthop Relat Res       Date:  2021-07-01       Impact factor: 4.755

  1 in total