| Literature DB >> 33814586 |
Swati Jindal1, Ankita Mittal1, Lakesh K Anand1, Manpreet Singh1, Dheeraj Kapoor1.
Abstract
BACKGROUND AND AIMS: The Air-Q blocker (Cook gas LLC, Mercury Medical, Clearwater, FL, USA) is a relatively new supraglottic airway device (SAD) with capability to serve as a conduit for intubation. As there is limited data on Air-Q blocker, the present study was performed to compare the efficacy of Air-Q blocker and Proseal laryngeal mask airway (PLMA) in patients undergoing elective surgery.Entities:
Keywords: Air-Q blocker; general anaesthesia; laryngeal masks; proseal laryngeal mask airway
Year: 2021 PMID: 33814586 PMCID: PMC7993039 DOI: 10.4103/ija.IJA_1254_20
Source DB: PubMed Journal: Indian J Anaesth ISSN: 0019-5049
Figure 1Consort Diagram
Demographic data
| Group A Air- Q blocker ( | Group P PLMA ( | ||
|---|---|---|---|
| Age (years) | 40.7±12.3 | 40.1±12.2 | 0.810 |
| Sex (M/F), | 6/37 (13.9%/86.1%) | 6/38 (13.6%/86.4%) | 0.966 |
| ASA (I/II) | 25/18 (58.1%/41.9%) | 33/11 (75%/25%) | 0.060 |
| Weight (kg) | 62.6±11.4 | 61.9±12.6 | 0.802 |
| Height (cm) | 157.1±5.2 | 157±5.9 | 0.923 |
| BMI (kg/m2) | 25.3±4.2 | 25±4.4 | 0.763 |
| Surgical time (min) | 82±32 | 76±31 | 0.58 |
| Anaesthesia time (min) | 106±37 | 104±38 | 0.24 |
Values are expressed as mean±standard deviation or number of patients (%); ASA - American Society of Anesthesiologists; BMI - Body mass index
SAD (Air-Q blocker and Proseal LMA) performance parameters
| Air-Q blocker Group A ( | Proseal LMA Group P ( | ||
|---|---|---|---|
| OLP (cm H2O) | 27.5±5.8 | 31.9±6.5 | 0.002 |
| Insertion Attempts | |||
| 1 | 39 (90.7%) | 42 (95.5%) | 0.434 |
| 2 | 4 (9.3%) | 2 (4.5%) | |
| Insertion time (s) | 16.4±3.4 | 18.1±3.3 | 0.019 |
| Ease of OGT insertion | 0.055 | ||
| Easy | 39 (90.7%) | 44 (100%) | |
| Difficult | 4 (9.3%) | 0 | |
| Impossible to insert | 0 | 0 | |
| Blood stain on device | |||
| Yes | 1 (2.3%) | 2 (4.5%) | 1.00 |
| No | 42 (97.7%) | 42 (95.5%) | |
| Postoperative (0,2,24 h) | |||
| Sore throat | 3,1,0 (6.9,2.32%, 0) | 4,2,0 (9.1, 4.5%, 0) | 1.000/1.000/- |
| Dysphagia | 1,0,0 (2.3%, 0,0) | 1,0,0 (2.3%, 0,0) | 1.000/-/- |
| Hoarseness | - | - | - |
Values are presented as Mean±Standard Deviation or number (%); LMA - Laryngeal mask airway; OLP - Oropharyngeal leak pressure; OGT - Orogastric tube
Fibreoptic scoring system involving a four-point assessment of the view of the vocal cords and epiglottis
| Score | Air-Q blocker | PLMA | |
|---|---|---|---|
| 4, only vocal cords visible | 27 (62.8%) | 17 (38.6%) | 0.038 |
| 3, vocal cords plus posterior epiglottis visible | 12 (27.9%) | 21 (47.7%) | |
| 2, vocal cords plus anterior epiglottis visible | 2 (4.7%) | 4 (9.1%) | |
| 1, vocal cords not visible | 2 (4.6%) | 2 (4.5%) |
Figure 2Airway compliance measured after device insertion at 5, 10, 15 min in PLMA and Air-Q blocker groups