| Literature DB >> 33795987 |
Anuoluwapo Sopeyin1, Benjamin K Young1, Martha A Howard1.
Abstract
Amblyopia is the most common cause of preventable visual impairment in children and occurs as a result of unilateral or bilateral impairment in best-corrected visual acuity. Early diagnosis and proper treatment are crucial to prevent poor visual outcomes in adulthood. Advances in technology have provided more objective diagnostic tools, which can now be used by a wide range of healthcare providers. Here, we highlight tools that have gained popularity in the past two decades and compare clinically relevant parameters to guide primary care providers seeking to incorporate instrumental vision screening in pediatric patient care.Entities:
Keywords: Amblyopia; Amblyopia Risk Factors; Instrumental Vision Screening; Preventive Ophthalmology; Vision Screening
Year: 2021 PMID: 33795987 PMCID: PMC7995935
Source DB: PubMed Journal: Yale J Biol Med ISSN: 0044-0086
Figure 1American Association of Pediatric Ophthalmology and Strabismus (AAPOS) Amblyopia Risk Factors (ARFs) Guidelines.
Comparison of seven portable vision screening instruments outlining the cost, amblyopia risk factor screened, and clinically useful characteristics.
| Instrument | Spot Vision Screener | 2WIN | S12C/R | 3000 | 5500P | GoCheck Kids# | Blinq | |
| Place of Manufacture | Skaneateles Falls, NY, USA | Padova, Italy | Nuremberg, Germany | Cordova, TN, USA | Chicago, IL, USA | Scottsdale, AZ, USA | Boston, MA, USA | |
| Cost (US$) | 6750-8436 | 5000-6850 | 5495-8898 | 5000 | 3545-4716 | 169* | 7495-8995 | |
| Amblyopia Risk Factors (ARFs) | Refractive Errors | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Strabismus | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
| Cataracts (media opacity) | No | No | Yes | Yes | No | No | No | |
| EMR friendly interface§ | No | No | Yes | No | No | Yes | No~ | |
| Battery Life (minutes)§ | 240 | 240 | 180/variable** | 480 | - | 360-480 | 360 | |
| PHI Transmission§ | No | Yes | No | Yes | No | Yes | Yes | |
| Remote analysis§ | No | No | No | Yes | No | Yes | No | |
| Weight (pounds)§ | 2.55 | 1.88 | 1.76-2.20 | 4.35 | 15.50 | 0.30-0.46 | 4.00 |
# Currently only on the iPhone, * only monthly finance available, § Clinically useful characteristics, ~ EMR (electronic medical record) interface is currently being developed, ** Based on the rechargeable AA battery used.
Spot Vision Screener (Welch Allyn, Skaneateles Falls, NY), 2WIN (Adaptica, Padova, Italy), the S12C/R Mobile Vision Screener autorefractor (Plusoptix, Nuremberg, Germany), the iScreen 3000 photo screener (iScreen Vision Inc, Cordova, TN), the OPTEC 5500 vision screener (Stereo Optical Co., Inc., Chicago, IL), the GoCheck Kids smartphone photo screening application (Goquity Mobile Health, Scottsdale, AZ) on the iPhone (Apple Inc, Cupertino, CA), and the Pediatric Vision Scanner or “blinq” (Rebion, Boston, MA).
Statistical Measures of Vision screening instruments stratified by referral criteria.
| Welch Allyn | Spot Vision Screener | Manufacturer (v. 1.03) [ | 80 | 74 | 88 | 61 |
| Manufacturer (v.1.151) [ | 88.1 | 71.9 | 79.3 | 83.1 | ||
| Manufacturer (v.2.0.16) [ | 87.7 | 75.9 | 81.7 | 83.4 | ||
| 2003 AAPOS [ | 77 | 87 | ||||
| 2013 AAPOS [ | 60-92.6 (87) | 70.4-93 (84.5) | 58.1-86 (77.1) | 75-98.9 (89.3) | ||
| Adaptica | 2WIN | Manufacturer [ | 71 | 67 | ||
| 2003 AAPOS* [ | 68-91 (79.5) | 68-84 (76) | 84 | |||
| 2013 AAPOS [ | 67.4 | 83.7 | 87.9 | 59.4 | ||
| Plusoptix | S12C/R | Manufacturer ROC [ | 85 | 73 | ||
| Manufacturer ROC3 [ | 86 | 84 | ||||
| Manufacturer ROC5 [ | 69-83 (76) | |||||
| 2013 AAPOS [ | 64-100 (90.2) | 61-93 (88) | 65-76 (65) | 87-100 (98.5) | ||
| iScreen | 3000 | Manufacturer [ | 77.2 | 94.1 | ||
| 2003 AAPOS [ | 66.2-90.7 (78) | 42.9-92 (81.8) | 81.8 | 75.5 | ||
| Optec | 5500 | 2013 AAPOS [ | 77.4-81 (79.2) | 87-100 (93.5) | 91.9-100 (96.0) | 50-71.4 (60.7) |
| Goquity | Gocheck kids | 2003 AAPOS [ | 81 | 91 | 92 | |
| 2013 AAPOS [ | 65-90.5 (81) | 67.2-85 (68.1) | 50-76 (56.9) | 80-94 (88.5) | ||
| Rebion | blinq. (pediatric vision scanner) | Manufacturer [ | 41-98 (96.5) | 75-96 (87) | 38-82 (47) | 78-100 (89) |
| 2003 AAPOS [ | 75 | 68 | 81 |
* slightly modified version as outlined in Arnold, 2020 [28].