Intrasession and Intersession Reliability of Isometric Trunk and Hip Strength Measurements Using the Portable Tension Dynamometer.

Sakiko Oyama, Edgar Garza, Kylie Dugan.   

Abstract

CONTEXT: The trunk/pelvis is an important link between the upper- and lower-extremities. Therefore, assessing strength of the trunk and hip muscles that control the segments is clinically meaningful. While an isokinetic dynamometer can be used to measure trunk strength, the equipment is expensive and not portable.
OBJECTIVE: To test the reliability of simple trunk and hip strength measures that utilize a bar, straps, and a portable tension dynamometer.
DESIGN: Test-retest reliability study.
SETTING: Biomechanics research laboratory. Patients (or Other Participants): Twenty college-age individuals (10 males/10 females, age = 20.9 [3.7] y) participated. INTERVENTION(S): The participants attended 2 testing sessions, 1 week apart. The participants' trunk-flexion, rotation, and hip abduction strength were measured at each session. MAIN OUTCOME MEASURES: Peak trunk flexion, rotation, and hip abduction forces were normalized to the participant's body weight (BW). In addition, hip-abduction torque was calculated by multiplying the force times the leg length and normalized to BW. The trial data from both sessions were used to calculate the intrasession reliability, and the averages from the 2 sessions were used to calculate the intersession reliability. Intraclass correlation coefficients, SEM, and minimal detectable change were calculated to evaluate reliability of measures.
RESULTS: The intrasession intraclass correlation coefficients (SEM) for trunk flexion, rotation, hip abduction, and hip abduction torque were .837 (5.2% BW), .978 (1.3% BW), .955 (1.0% BW), and .969 (5.8 N·m/BW), respectively. The intersession reliability for trunk flexion, rotation, hip abduction, and hip abduction torque were .871 (4.3% BW), .801 (3.8% BW), .894 (1.5% BW), and .968 (5.9 N·m/BW), respectively.
CONCLUSIONS: The measures of trunk and hip abduction strength are highly repeatable within a session. The reliability of the measures between sessions was also good/excellent with relatively small SEM and minimal detectable change. The tests described in this study can be used to assess changes in trunk/hip strength over time.

Entities:  

Keywords:  isometric trunk strength; rotation; trunk strength

Mesh:

Year:  2021        PMID: 33785657     DOI: 10.1123/jsr.2020-0286

Source DB:  PubMed          Journal:  J Sport Rehabil        ISSN: 1056-6716            Impact factor:   1.931


  1 in total

1.  Verification of Shoulder External Rotators Strength Measurement Using a Suspension Scale.

Authors:  Atsushi Ueda; Yasuhiro Mitani; Hitoshi Koda; Toshimitsu Omine; Ryuta Inada; Naoyuki Konishi; Shunsai Mori
Journal:  Cureus       Date:  2022-06-20
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.