Accuracy of a single-lead mobile smartphone electrocardiogram for QT interval measurement in patients undergoing maintenance methadone therapy.

Erika N Titus-Lay1,2, Heather A Jaynes2, Elena Tomaselli Muensterman2, Carol A Ott1,2,3, Todd A Walroth1, Gabriela Williams1, Paul R Moe4, Michelle Wilbrandt4, James E Tisdale2,3.   

Abstract

STUDY
OBJECTIVE: Methadone is associated with QT interval prolongation and torsades de pointes. Expert panel recommendations advocate a pre-methadone electrocardiogram (ECG) and another ECG at 30 days of therapy in patients with risk factors. Some guidelines recommend a pre-methadone ECG and routine ECG monitoring in all methadone patients, but this is controversial due to the resources required. Availability of a convenient, less resource-intensive method of ECG monitoring for patients taking methadone is desirable. The objective of this study was to assess the accuracy of a handheld smartphone ECG (iECG) for QT measurement in patients on maintenance methadone therapy in an urban opioid treatment program.
DESIGN: Prospective study.
SETTING: Urban opioid treatment program. PATIENTS: n = 115 patients in normal sinus rhythm who were on steady-state maintenance methadone therapy INTERVENTION: Patients (n = 115) underwent a simultaneous 12-lead ECG and a single-lead iECG.
MEASUREMENTS AND MAIN RESULTS: The first three QT and RR intervals from lead II of the 12-lead ECG and simulated lead I from the iECG were compared using the Bland-Altman analysis of measurement agreement. Mean [± standard deviation) age was 34 ± 11 years; 71% were female, 75% were white. Compared to the 12-lead ECG, the iECG was associated with a QTc bias of - 0.14 ms (SD = 12 ms, 95% CI = -2.4 to 2.1 ms). The absolute mean difference in QTc between the two methods was 9.5 ± 7.1 ms. For identification of patients with methadone-associated QTc prolongation, the iECG performed moderately well [c-statistic 0.97 (95% CI 0.91-0.99); sensitivity and specificity 75% (95% CI 43-95%) and 99% (95% CI 94-99%), respectively]. The positive and negative likelihood ratios of the iECG for identifying patients with methadone-associated QTc prolongation were 77.25 (95% CI 10.69 to 558.18) and 0.25 (95% CI 0.09 to 0.67), respectively, while the positive and negative predictive values were 90% (95% CI 56-99%) and 97% (95% CI 92-99%), respectively. The accuracy of the iECG for identifying patients with QTc prolongation was 97% (95% CI 91-99%).
CONCLUSION: A handheld smartphone ECG is accurate for QT interval measurement in patients taking maintenance methadone therapy, and its performance is moderately good for identifying patients with methadone-associated QTc prolongation.
© 2021 Pharmacotherapy Publications, Inc.

Entities:  

Keywords:  QT interval; QT interval prolongation; electrocardiography; methadone; smartphone

Mesh:

Substances:

Year:  2021        PMID: 33772822     DOI: 10.1002/phar.2521

Source DB:  PubMed          Journal:  Pharmacotherapy        ISSN: 0277-0008            Impact factor:   4.705


  1 in total

Review 1.  Remote and wearable ECG devices with diagnostic abilities in adults: A state-of-the-science scoping review.

Authors:  Zeineb Bouzid; Salah S Al-Zaiti; Raymond Bond; Ervin Sejdić
Journal:  Heart Rhythm       Date:  2022-03-09       Impact factor: 6.779

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.