Karol Julieth García1, Julián David Ocampo2, María Del Pilar Pardo3, Tatiana Aguilar4, Carlos Alberto Ruiz5, Andrés Castaño6.
Abstract
Introduction: Mammography quality is directly related to the ability to detect an abnormality and, therefore, quality control is necessary for diagnostic imaging centers. Objective: To evaluate image quality, reading, and mammography service in some diagnostic imaging centers in Manizales, Colombia. Materials and methods: Four diagnostic imaging centers participated voluntarily in the study under confidentiality agreements. Out of 520 women attending the centers, 318 had a mammography. The infrastructure, technology, and human resources of each unit were evaluated based on visual inspections. A radiologist expert in reading and clinical interpretation of mammary images evaluated the quality of the image and the reading. We made the statistical analysis using anova, the kappa index, and the percentage of disagreement.Entities:
Keywords: Mammography; quality management; diagnostic imaging; quality assurance; health care; maintenance and engineering; hospital; biomedical engineering
Year: 2021 PMID: 33761189 PMCID: PMC8055586 DOI: 10.7705/biomedica.5135
Source DB: PubMed Journal: Biomedica ISSN: 0120-4157 Impact factor: 0.935
Valoración del índice kappa. Valores de referencia tomados para determinar la concordancia entre los observadores
| Menor de 0,00 | Sin acuerdo |
| 0,00 a 0,20 | Insignificante |
| 0,21 a 0,40 | Media |
| 0,41 a 0,60 | Moderada |
| 0,61 a 0,80 | Sustancial |
| 0,81 a 1,00 | Casi perfecta |
Descripción de la relación de los resultados de cada una de las pruebas estadísticas utilizadas en el estudio: valor kappa, índice del anova de un factor e índice de desacuerdo
| Moderado | p mayor de 0,05: se acepta la hipótesis nula. | Porcentaje menor de 10% |
| Sin acuerdo | p menor de 0,05: se rechaza la hipótesis nula. | Porcentaje mayor de 10% |
Resultados de cada centro de imágenes según los principales criterios de calidad de la lectura de la imagen mamográfica establecidos en el estudio
| C1 | 0,724 | 0,6019 | 4,5 |
| C2 | 0,318 | 0,0960 | 4,1 |
| C3 | 0,403 | 0,0740 | 25,0 |
| C4 | 0,334 | 0,0013 | 21,9 |
| C1 | 0,327 | 0,000260633 | 20,5 |
| C2 | 0,602 | 0,999999819 | 10,8 |
| C3 | 0,175 | 0,238135452 | 31,9 |
| C4 | 0,516 | 0,295821213 | 15,6 |
| C1 | -0,135 | 0,155614956 | 85,71 |
| C2 | -0,111 | 1 | 62,5 |
| C3 | 0,146 | 0,002121124 | 71,4 |
| C4 | -0,286 | 0,188998593 | 66,6 |
| C1 | 0,063 | 0,007974464 | 85,71 |
| C2 | 0,429 | 0,316979253 | 25,0 |
| C3 | -0,2 | 0,599867373 | 85,7 |
| C4 | -0,2 | 0,130932801 | 66,6 |
| C1 | 0,097 | 0,013243586 | 18,2 |
| C2 | -0,21 | 0,313958542 | 5,4 |
| C3 | 0,179 | 0,299705705 | 19,8 |
| C4 | 0,122 | 0,00036924 | 30,8 |
| C1 | 0,044 | 4,53146E-13 | 17,0 |
| C2 | 0 | 8,0187E-14 | 48,7 |
| C3 | 0,017 | 0,314597456 | 4,4 |
| C4 | NA | NA | 0,0 |
| C1 | 0,131 | 3,19121E-11 | 46,6 |
| C2 | 0,456 | 0,080078509 | 24,3 |
| C3 | 0,131 | 7,7032E-06 | 22,0 |
| C4 | 0,618 | 0,853522694 | 18,5 |
| C1 | 0,476 | 0,309552041 | 6,1 |
| C2 | NA | NA | 0,0 |
| C3 | 0,549 | 0,700602226 | 4,3 |
| C4 | 0 | 0,155863942 | 5,3 |
| C1 | 0,05 | 1,04253E-34 | 77,3 |
| C2 | 0,04 | 6,66932E-22 | 68,1 |
| C3 | 0,009 | 2,63482E-33 | 76,9 |
| C4 | 0,227 | 6,71366E-07 | 46,2 |
| C1 | 0,093 | 3,95035E-11 | 38,4 |
| C2 | 0,275 | 0,008858293 | 16,2 |
| C3 | 0,269 | 0,000312655 | 36,3 |
| C4 | 0,073 | 0,001912098 | 38,5 |
NA: no aplica