Flávio de Mendonça Copello1, Daniel Paludo Brunetto2, Carlos Nelson Elias3, Matheus Melo Pithon4, Raildo Silva Coqueiro4, Amanda Cunha Regal de Castro1, Eduardo Franzotti Sant'anna1.
Abstract
OBJECTIVE: Assess the influence of mono- and bicortical anchorage and diameter of mini-implants (MIs) on the primary stability of these devices.Entities:
Mesh:
Substances:
Year: 2021 PMID: 33759966 PMCID: PMC8018750 DOI: 10.1590/2177-6709.26.1.e211967.oar
Source DB: PubMed Journal: Dental Press J Orthod ISSN: 2176-9451
Figure 1:Flowchart showing distribution of specimens into the groups/subgroups according to their diameter and type of insertion.
Figure 5:Graph showing pull-out strength results. (D= Diameter; BI=Bone Insertion). Only mini-implant diameter influenced pull-out strength values.
Figure 3:Graph showing insertion torque ( IT ) analysis (D=Diameter; BI=Bone Insertion). Higher IT values were found in mini-implants with greater diameter and bicortical insertion. * ANOVA two-way: a, b, c distinct letters indicate statistical difference (p ≤ 0.05).
Figure 4:Graph showing Periotest assessment (D= Diameter; BI=Bone Insertion). The results indicated influence of diameter and insertion type on mobility. Mini-implant mobility was statistically lower for mini-implants with larger diameters regardless of anchorage insertion type. * Kruskal-Wallis test: a, b, c distinct letters indicate statistical difference between the diameters (Mann-Whitney test); † Mann-Whitney test.
Descriptive statistics for Insertion Torque and Pull-out Strength.
| Primary Stability Tests | BI | Diameter (D) | Effect | *p-valor | ||
|---|---|---|---|---|---|---|
| 1.5 mm | 1.8 mm | 2.0 mm | ||||
| Insertion Torque (Ncm) | D | < 0.001 | ||||
| MC | 16.66 ± 0.45a | 18.95 ± 0.33b | 29.67 ± 0.34c | BI | < 0.001 | |
| BC | 24.61 ± 0.47a† | 28.13 ± 0.18b† | 37.00 ± 0.19c† | D x BI | < 0.001 | |
| Pull-out Strength (N) | a | b | c | D | < 0.001 | |
| MC | 124.23 ± 4.10 | 182.78 ± 2.87 | 268.40 ± 5.05 | BI | 0.288 | |
| BC | 125.58 ± 4.84 | 181.87 ± 3.98 | 271.41 ± 3.70 | D x BI | 0.335 | |
D, Diameter; BI, Bone Insertion; D x BI, effect between diameter and bone insertion; MC, monocortical; BC, bicortical. Results are expressed as mean ± standard deviation. * ANOVA two-way: a, b, c distinct letters indicate statistical difference (p ≤ 0.05) between the diameters, according to the comparisons for interaction effect (insertion torque) and Tukey test (pull-out strength) ; † indicates statistical difference (p ≤ 0.05) between types of bone insertion.
Descriptive statistics for mini-implant micromobility.
| Bone Insertion | Diameter | *p-valor | ||
|---|---|---|---|---|
| 1.5 mm | 1.8 mm | 2.0 mm | ||
| Monocortical | 17.75 ± 1,10a | 14.25 ± 0,60b | 4.50 ± 0.60c | < 0.001 |
| Bicortical | 14.00 ± 0,06a | 9.00 ± 0,80b | 1.00 ± 1.00c | < 0.001 |
| †p-valor | < 0.001 | < 0.001 | < 0.001 | |
The results are expressed as median ± interquartile amplitude. * Kruskal-Wallis test: a, b, c distinct letters indicate statistical difference between the diameters (Mann-Whitney test); † Mann-Whitney test.