Cross-Calibration of Prodigy and Horizon A Densitometers and Precision of the Horizon A Densitometer.

Elise Reitshamer1, Kelsey Barrett1, Kyla Shea1, Bess Dawson-Hughes2.   

Abstract

We performed this study to enable a reliable transition for clinical study participants and patients from a GE Lunar Prodigy to a Hologic Horizon A dual-energy X-ray absorptiometry (DXA) scanner and to assess the reproducibility of measurements made on the new DXA scanner. Forty-five older adults had one spine, hip, and total body scan on a Prodigy dual-energy X-ray absorptiometry (DXA) scanner and 2 spine, hip, and total body scans, with repositioning, on a new Hologic Horizon A DXA scanner. Linear regression models were used to derive cross calibration equations for each measure on the 2 scanners. Precision (group root-mean-square average coefficient of variation) of bone mineral density (BMD) of the total hip, femoral neck, and lumbar spine (L1-L4), and total body fat, bone, and lean mass, appendicular lean mass, and trabecular bone score (TBS) was assessed using the International Society of Clinical Densitometry's (ISCD's) Advanced Precision Calculation Tool. Correlation coefficients for the BMD and body composition measures on the 2 scanners ranged from 0.94 to 0.99 (p<0.001). When compared with values on the Prodigy, mean BMD on the Horizon A was lower at each skeletal site (0.136 g/cm2 lower at the femoral neck and 0.169 g/cm2 lower at the lumbar spine (L1-4)), fat mass was 0.47 kg lower, and lean mass was 4.50 kg higher. Precision of the Horizon A scans was 1.60% for total hip, 1.94% for femoral neck, and 1.25% for spine (L1-4) BMD. Precision of TBS was 1.67%. Precision of total body fat mass was 2.16%, total body lean mass was 1.26%, appendicular lean mass was 1.97%, and total body bone mass was 1.12%. The differences in BMD and body composition values on the 2 scanners illustrate the importance of cross-calibration to account for these differences when transitioning clinical study participants and patients from one scanner to another.
Copyright © 2021 The International Society for Clinical Densitometry. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  GE lunar prodigy; cross-calibration; hologic horizon A; precision

Year:  2021        PMID: 33744116     DOI: 10.1016/j.jocd.2021.02.003

Source DB:  PubMed          Journal:  J Clin Densitom        ISSN: 1094-6950            Impact factor:   2.617


  1 in total

1.  Trabecular Bone Score Reference Values for Children and Adolescents According to Age, Sex, and Ancestry.

Authors:  Heidi J Kalkwarf; John A Shepherd; Didier Hans; Elena Gonzalez Rodriguez; Joseph M Kindler; Joan M Lappe; Sharon Oberfield; Karen K Winer; Babette S Zemel
Journal:  J Bone Miner Res       Date:  2022-02-23       Impact factor: 6.390

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.