Accuracy Assessment of SUV Measurements in SPECT/CT: A Phantom Study.

Fatin Halim1,2, Hizwan Yahya1,2, Khairul Nizam Jaafar3, Syahir Mansor4,3.   

Abstract

Advances in iterative image reconstruction enable absolute quantification of SPECT/CT studies by incorporating compensations for collimator-detector response, attenuation, and scatter. This study aimed to assess the quantitative accuracy of SPECT/CT based on different levels of 99mTc activity (low/high) using different SUV metrics (SUVmean, SUVmax, SUV0.6 max, and SUV0.75 max [the average values that include pixels greater than 60% and 75% of the SUVmax in the volume of interest, respectively]).
Methods: A Jaszczak phantom equipped with 6 fillable spheres was set up with low and high activity ratios of 1:4 and 1:10 (background-to-sphere) on background activities of 10 and 60 kBq/mL, respectively. The fixed-size volume of interest based on the diameter of each sphere was drawn on SPECT images using various metrics for SUV quantification purposes.
Results: The convergence of activity concentration was dependent on the number of iterations and application of postfiltering. For the background-to-sphere ratio of 1:10 with a low background activity concentration, the SUVmean metric showed an underestimation of about 38% from the actual SUV, and SUVmax exhibited an overestimation of about 24% for the largest sphere diameter. Meanwhile, bias reductions of as much as -6% and -7% for SUV0.6 max and SUV0.75 max, respectively, were observed. SUVmax gave a more accurate reading than the others, although points that exceeded the actual value were detected. At 1:4 and 1:10 background activity of 10 kBq/mL, a low activity concentration attained a value close to the actual ratio. Use of 2 iterations and 10 subsets without postfiltering gave the most accurate values for reconstruction and the best image overall.
Conclusion: SUVmax is the best metric in a high- or low-contrast-ratio phantom with at least 2 iterations, 10 subsets, and no postfiltering.
© 2021 by the Society of Nuclear Medicine and Molecular Imaging.

Entities:  

Keywords:  SPECT/CT; SUV; phantom; quantitative; ratio

Mesh:

Year:  2021        PMID: 33722927      PMCID: PMC8712636          DOI: 10.2967/jnmt.120.259168

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  16 in total

1.  A threshold method to improve standardized uptake value reproducibility.

Authors:  J R Lee; M T Madsen; D Bushnel; Y Menda
Journal:  Nucl Med Commun       Date:  2000-07       Impact factor: 1.690

2.  AMIDE: a free software tool for multimodality medical image analysis.

Authors:  Andreas Markus Loening; Sanjiv Sam Gambhir
Journal:  Mol Imaging       Date:  2003-07       Impact factor: 4.488

3.  A filtered backprojection algorithm with characteristics of the iterative landweber algorithm.

Authors:  Gengsheng L Zeng
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

Review 4.  A systematic review of the factors affecting accuracy of SUV measurements.

Authors:  Michael C Adams; Timothy G Turkington; Joshua M Wilson; Terence Z Wong
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

5.  Comparison of advanced iterative reconstruction methods for SPECT/CT.

Authors:  Peter Knoll; Daniela Kotalova; Gunnar Köchle; Ivan Kuzelka; Greg Minear; Siroos Mirzaei; Martin Sámal; Ladislav Zadrazil; Helmar Bergmann
Journal:  Z Med Phys       Date:  2011-07-01       Impact factor: 4.820

Review 6.  An evidence-based review of quantitative SPECT imaging and potential clinical applications.

Authors:  Dale L Bailey; Kathy P Willowson
Journal:  J Nucl Med       Date:  2013-01       Impact factor: 10.057

7.  Quantitative Assessment of Radionuclide Uptake and Positron Emission Tomography-computed Tomography Image Contrast.

Authors:  Hasford Francis; John Humphrey Amuasi; Kyere Augustine Kwame; Mboyo Di Tamba Vangu
Journal:  World J Nucl Med       Date:  2016-09

8.  Optimization of Ordered Subset Expectation Maximization Reconstruction for Reducing Urinary Bladder Artifacts in Single-photon Emission Computed Tomography Imaging.

Authors:  A M Katua; A O Ankrah; M Vorster; A van Gelder; M M Sathekge
Journal:  World J Nucl Med       Date:  2011-01

9.  Filtering in SPECT Image Reconstruction.

Authors:  Maria Lyra; Agapi Ploussi
Journal:  Int J Biomed Imaging       Date:  2011-06-23

10.  The EANM practice guidelines for bone scintigraphy.

Authors:  T Van den Wyngaert; K Strobel; W U Kampen; T Kuwert; W van der Bruggen; H K Mohan; G Gnanasegaran; R Delgado-Bolton; W A Weber; M Beheshti; W Langsteger; F Giammarile; F M Mottaghy; F Paycha
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-06-04       Impact factor: 9.236