Abstract
BACKGROUND: Although various studies have been conducted to demonstrate the possibility of Raman spectroscopy (RS) as a diagnostic tool for prostate cancer (PC), it is difficult to use it in the real clinical area because of imitations in various research processes. Therefore, we did a systematic review and meta-analysis about the accuracy in diagnostic use of RS for PC.Entities:
Keywords: Prostate cancer (PC); Raman spectroscopy (RS); meta-analysis; optical, diagnosis
Year: 2021 PMID: 33718060 PMCID: PMC7947438 DOI: 10.21037/tau-20-924
Source DB: PubMed Journal: Transl Androl Urol ISSN: 2223-4683
Figure 1PRISMA flowchart.
Characteristics of the five studies included in the meta-analysis
| Study | Country | Total number of patients | Total number of spectra (malignant/control) | Sample type | Type of Raman spectroscopy | Main Raman algorithm type | Spectral analysis | Laser (nm) | Total sensitivity (%) | Total specificity (%) |
|---|---|---|---|---|---|---|---|---|---|---|
| Crow 2003, ( | United Kingdom | 27 | 456 (249/207) | snap frozen in liquid nitrogen-stored prostate tissue by prostate biopsy (stored at −80 °C) | Renishaw system 1000 | PCA and LDA | MATLAB (Mathwork Inc.) | 832 | 91.6 | 92.8 |
| Crow 2005, ( | United Kingdom | 37 | 197 (77/120) | snap frozen prostate tissue by TURP (stored at −80 °C) | Near-infrared Raman spectroscopy | PCA and LDA | MATLAB 6.0 (Mathwork Inc.) | 785 | 80 | 88.9 |
| Lopes 2011, ( | Brazil | N/A | 105 (14/91) | Snap frozen and stored in liquid nitrogen stored prostate tissue by radical prostatectomy, TURP, autopsy | Table-top dispersive Raman spectrometer | PCA | MATLAB 7.0 (Mathwork Inc.) | 830 | 100 | 100 |
| Silveira 2014, ( | Brazil | 32 | 152 (75/77) | Snap frozen in liquid nitrogen-stored prostate tissue by surgical procedure | Table-top dispersive Raman spectrometer (model Dimension P-1, Lambda solutions) | Ordinary least-squares analysis | MATLAB 7.0 (Mathwork Inc.) | 830 | 73.7 | 75 |
| Stone 2003, ( | United Kingdom | N/A | 520 (313/207) | Snap frozen in liquid nitrogen-stored prostate tissue by prostate biopsy (stored at −80 °C) | customized Renishaw system 1000 Raman micro-spectrometer | PCA and LDA | MATLAB (Mathwork Inc.) | 830 | 93.3 | 92.2 |
PCA, principal components analysis; LDA, linear discriminant analysis; TURP, transurethral resection of prostate.
Figure 2The pooled sensitivity and specificity of Raman spectroscopy (RS). The pooled sensitivity and specificity of RS were 0.89 (95% CI: 0.87–0.91) and 0.91 (95% CI: 0.89–0.93), respectively.
Figure 3The area under the curve (AUC) of summary receiver operating characteristic (SROC) curves. The AUC of SROC curve was 0.93.
Figure 4Risk of bias and applicability concerns graph. The risk of bias in all included studies was low in all domains except for patient selection. The applicability concerns were also almost low grade.
Figure 5Risk of bias and applicability concerns summary. The risk of bias in all included studies was low in all domains except for patient selection. The applicability concerns were also low grade except for one study.