Ying Ping1, Xuchu Wang1, Yibei Dai1, Danhua Wang1, Weiwei Liu1, Pan Yu1, Zhihua Tao1.
Abstract
BACKGROUND: Cardiotrophin-1 (CT-1) is a cytokine that could induce cardiomyocytes hypertrophy and dysfunction. Plasma CT-1 might serve as a cardiac biomarker both in diagnosis, staging, and prognostic assessment of heart failure.Entities:
Keywords: Cardiotrophin-1; chemiluminescence immunoassay; chronic heart failure; paramagnetic particles; quantitative detection
Year: 2021 PMID: 33713510 PMCID: PMC8059751 DOI: 10.1002/jcla.23570
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Clinical and echocardiographic characteristics of the study group
| Variables | HF patients | Healthy subjects |
|---|---|---|
| Age(years) | 22‐84 | 66 ± 13 |
| Sex(male/female) | 72/28 | 27/13 |
| BMI (kg/m2) | 18.9 ± 3.4 | 19.5 ± 3 |
| HF (beats/min) | 74 ± 5 | 72 ± 5 |
| Systolic pressure | 118 ± 23 | 116 ± 30 |
| Diastolic pressure | 71 ± 17 | 72 ± 10 |
| LVEF (%) | <50% | >70% |
| Treatments | ||
| β blockers, n, (%) | 41(41%) | |
| Diuretic, n, (%) | 100(100%) |
Data are presented as mean ± standard deviation or as absolute numbers (n) with percentage (% of the referring group) in brackets.
Abbreviations: BMI, Body Mass Index; HF, heart rate; LVEF, left ventricular ejection fraction.
Figure 1Schematic illustration of the quantitative detection of CT‐1 by MPs‐CLIA. 50 μL bio‐Ab (capture antibody) (1.6 μg/ mL), 50 μL AE ‐Ab (detector antibody) (1.6 μg/mL) was incubated with 100 μL CT‐1 in PBS or plasma at 37°C for 30 min. Thereafter, MPs‐SA was incubated with bio‐Ab‐CT‐1‐AE‐Ab sandwich immunocomplexes and then unreacted antibodies were removed by magnetic separation. After washing three times, The luminescence signal was generated by the trigger solution containing 200 μL acid buffer with oxidants hydrogen peroxide (H2O2) and 200 μL sodium hydroxide (NaOH). Then the relative light unit (RLU) was measured with a photomultiplier instrument to quantify CT‐1 concentration
Figure 2Optimization of crucial parameters for CT‐1 detection by MPs‐CLIA. A, antibody concentrations. Color represents relative light unit, the darker the color, the higher the relative light unit. B, incubation time, closed circles represent the incubation times of 5, 10, 15, 30, 45 min, and 1 h; C, capture time, closed circles represent the incubation times of 5, 10, 15, 30, 45 min, and 1 h; D, magnetic particle concentration, closed circles represent streptavidin‐modified paramagnetic particles concentrations of 0.14, 0.28, 0.43, 0.72, and 0.86 μg/μL
Figure 3Calibration curve for the determination of CT‐1 by MPs‐CLIA. MPs‐CLIA was conducted using bio‐Ab and AE‐Ab at concentration of 1.6 μg/mL, respectively. Reaction time and capture time were 15 and 30 min, respectively. The dotted line represents the linearized curve ranging from 7.8 to 2000 pg/mL. RLU: relative light unit; CCT‐1: the concentration of CT‐1
Comparison between MPs‐CLIA and the previous reported detection methods
| Methods | Time (hour) | LOD | linear range | reference |
|---|---|---|---|---|
| ELISA | ~4 | 5.6pg/mL | 16‐2000pg/mL |
|
| ILAM | >24 | 1.1 ± 0.8fmol/mL | N.F |
|
| RIA | >16 | N.F | 120‐8300fmol/mL |
|
| CIA | >24 | 2.9fmol/mL | N.F |
|
| MPs‐CLIA | ~1 | 1pg/ mL | 7.8‐2000pg/mL | Present study |
Abbreviations: CIA, Non‐competitive immunochemiluminometric assay; ELISA, Enzyme‐linked immunosorbent assay; ILAM, Competitive immunoluminometric assay; N.F, No reference; RIA, Competitive radioimmunoassay.
Results of CT‐1 determination in spiked serum/plasma samples Calibration equation: = 0.9996 Results are the mean of n = 10 measurements
|
Spiked CT1 concentration (pg/mL) |
Calculated CT1 concentration (pg/mL) |
Mean recovery (%) | RSD (%) |
|---|---|---|---|
| 10 | 9.6 | 96% | 8.1 |
| 100 | 104 | 104 | 6.3 |
| 800 | 885 | 110 | 4.2 |
Intra‐assay and inter‐assay of MPs‐CLIA
|
CT‐1 concentration (pg/mL) | CVs (%) | |
|---|---|---|
|
inter‐assay (n = 30) |
intra‐assay (n = 10) | |
| 10 | 9.25% | 8.92% |
| 100 | 10.9% | 6.69% |
| 800 | 4.30% | 3.54% |
Coefficient of variation (CV)=100 × mean/standard deviation.
Figure 4CT‐1 plasma level in 100 chronic heart failure patients and 40 controls. A, Comparison between healthy people and CHF patients, (P < .05). B, Receiver operating characteristic curves for the ability of cardiotrophin‐1 (CT‐1) to diagnosis in CHF patients. The area under the ROC curve gave a value of 0.66. A threshold value of 52.06 pg/ mL gives a sensitivity of 63.5% and a specificity of 65%
Figure 5Correlation between plasma levels of CT‐1 and pro‐BNP in 44 CHF patients