Robert S Chisena1, Jordan Sengenberger2, Albert J Shih3, Hitinder Gurm4.
Abstract
The lack of suitable atherosclerotic calcification models and testing strategies inhibits preclinical efficacy testing of existing and novel percutaneous devices. The goal of this study is to develop a preclinical testing method for quantitatively and qualitatively evaluating the efficacy of noncompliant balloon angioplasty (NC BA) treatment in human ex vivo calcified plaque (CP). NC BA using a 3- and 4-mm diameter balloon was performed on an ex vivo tibial calcified vessel obtained from an amputation. Three-dimensional microcomputed tomography (μ-CT) imaging was performed pre- and post-BA to compare crack density in the CP. Comparing the pre- and posttreatment three-dimensional μ-CT images showed a glass-like cracking that occurred in the CP due to the BA procedure. Expansion of the 3-mm balloon showed little tissue deformation and no CP cracking. Although expansion of the 4-mm balloon occurred nonuniformly along balloon length and across the perpendicular projections, the balloon generated cracking throughout the CP, which allowed the surrounding elastic tissue to be dilated. This combined X-ray microscopy and μ-CT technique is a useful preclinical tool for quantifying the efficacy of percutaneous treatments for CP. Because of its nondestructive nature, the CP structure can be visualized pre- and posttreatment to determine the treatment effect.Entities:
Keywords: Ex vivo calcified plaque model; Microcomputed tomography; Noncompliant balloon angioplasty; Preclinical efficacy evaluation model
Year: 2021 PMID: 33710527 DOI: 10.1007/s11517-021-02334-w
Source DB: PubMed Journal: Med Biol Eng Comput ISSN: 0140-0118 Impact factor: 2.602