Analysis of factors determining thermal changes at osteotomy site in dental implant placement - An in-vitro study.

Radhu Raj1, V Manju2, Vinod Kumar-Gopal3, Manu Eswar4.   

Abstract

BACKGROUND: Heat generation during osteotomy site preparation is a crucial factor that determines the success of dental implant placement. Among the factors that affect the heat generation, drilling speed, hand pressure and coolant temperature are independent variables. However, a relation between these three parameters and their optimal values required for the maximum outcome has not been studied so far. This study aims at finding out a relation between these factors in order to derive the optimum balance required, using an in vitro study.
MATERIAL AND METHODS: This in vitro experiment was performed on bovine femur. A total of 72 drillings were undertaken with the aid of a physiodispenser mounted on the test apparatus. Drill diameters of 2 mm and 2.8 mm, rotated at 1500, 2000 and 2500 rpm were included for the analysis. Hand pressures included for the comparison were 1.2 kgf and 2.4 kgf. Normal saline at room temperature, and that chilled to 00C were used for external irrigation. The temperature generated during drilling was recorded by infrared thermography using a Forward-Looking Infrared (FLIR) camera.
RESULTS: The highest temperature during osteotomy was observed at 2000 rpm rotational speed, 1.2 kgf operator hand pressure and saline irrigant solution at room temperature. In contrast, the lowest temperature generated was using 2500 rpm rotational speed, 2.4 kgf operator hand pressure and chilled irrigant solution.
CONCLUSIONS: The results indicate that none of the three experimented parameters generated heat above the critical temperature for bone necrosis (47°C). Thus, a high drilling speed with high hand pressure and continuous irrigation with copious amounts of cooled saline may be the ideal combination for implant osteotomy site preparation. Key words:Heat generation, dental implant drills, drilling speed, drilling pressure, irrigation, infrared thermography, thermal necrosis, osteotomy preparation. Copyright:
© 2021 Medicina Oral S.L.

Entities:  

Year:  2021        PMID: 33680325      PMCID: PMC7920569          DOI: 10.4317/jced.57346

Source DB:  PubMed          Journal:  J Clin Exp Dent        ISSN: 1989-5488


  17 in total

1.  Measurement of thermal conductivity of bovine cortical bone.

Authors:  S R Davidson; D F James
Journal:  Med Eng Phys       Date:  2000-12       Impact factor: 2.242

Review 2.  Risk factors associated with early failure of dental implants. A literature review.

Authors:  Cristina Palma-Carrió; Laura Maestre-Ferrín; David Peñarrocha-Oltra; María A Peñarrocha-Diago; Miguel Peñarrocha-Diago
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2011-07-01

3.  Atraumatic surgical technique and implant bed preparation.

Authors:  F Sutter; G Krekeler; A E Schwammberger; F J Sutter
Journal:  Quintessence Int       Date:  1992-12       Impact factor: 1.677

4.  The effect of speed, pressure, and time on bone temperature during the drilling of implant sites.

Authors:  D L Brisman
Journal:  Int J Oral Maxillofac Implants       Date:  1996 Jan-Feb       Impact factor: 2.804

5.  The effect of irrigation on osteotomy depth and bur diameter.

Authors:  M J Yacker; M Klein
Journal:  Int J Oral Maxillofac Implants       Date:  1996 Sep-Oct       Impact factor: 2.804

6.  Influence of different parameters on bone heating and drilling time in implantology.

Authors:  Y Reingewirtz; S Szmukler-Moncler; B Senger
Journal:  Clin Oral Implants Res       Date:  1997-06       Impact factor: 5.977

7.  Effect of drill speed on bone temperature.

Authors:  M B Abouzgia; J M Symington
Journal:  Int J Oral Maxillofac Surg       Date:  1996-10       Impact factor: 2.789

8.  Temperatures measured in human cortical bone when drilling.

Authors:  L S Matthews; C Hirsch
Journal:  J Bone Joint Surg Am       Date:  1972-03       Impact factor: 5.284

9.  A 15-year study of osseointegrated implants in the treatment of the edentulous jaw.

Authors:  R Adell; U Lekholm; B Rockler; P I Brånemark
Journal:  Int J Oral Surg       Date:  1981-12

10.  Heat caused by drilling cortical bone. Temperature measured in vivo in patients and animals.

Authors:  A R Eriksson; T Albrektsson; B Albrektsson
Journal:  Acta Orthop Scand       Date:  1984-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.