Bernadette C Young1, David W Eyre2, Katie Jeffery3.
Abstract
Entities:
Keywords: COVID-19; Infection control; Point-of-care testing
Mesh:
Year: 2021 PMID: 33662408 PMCID: PMC7917467 DOI: 10.1016/j.jinf.2021.02.025
Source DB: PubMed Journal: J Infect ISSN: 0163-4453 Impact factor: 6.072
Sensitivity of Lateral flow device (LFD) compared with Polymerase Chain Reaction (PCR), depending on Cycle threshold (Ct) value (mean Ct of all detected targets).
| All PCR | Mean Ct across detected PCR targets | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Positive | Negative | Invalid | <15 | 15–19.9 | 20–24.9 | 25–29.9 | ≥30 | ||
| LFD result | Positive | 133 | 0 | 1 | 59 | 47 | 23 | 4 | 0 |
| Negative | 80 | 572 | 15 | 12 | 15 | 31 | 19 | 3 | |
| Invalid | 1 | 1 | 0 | 0 | 0 | 1 | 0 | 0 | |
| Cumulative sensitivity (95% CI) | 83% | 80% | 69% (62–76) | 63% (56–70) | 62% (56–69) | ||||
Fig. 1Lateral flow device (LFD) results and Mean Cycle threshold (Ct) value of Polymerase Chain Reaction (PCR) target detection in 214 patients with SARS-CoV-2 detected. The median (central line), inter-quartile range (box) and range (whiskers) of Ct values are shown.