Literature DB >> 33646355

Metallosis after using distal fibular locking plate for lateral malleolar fractures: a retrospective study.

Sung Jun Park1, Gi Cheol Bae1, Dae Gyu Kwon2.   

Abstract

INTRODUCTION: Studies regarding the development of metallosis following open reduction and internal fixation (ORIF) of fractures are rarely found in orthopedic literature. The aim of the current study was to assess metallosis following ORIF using distal fibular locking plates to treat distal fibular fractures.
MATERIALS AND METHODS: 69 patients who underwent surgery using locking compression plates to treat lateral malleolar fractures, with a minimum 1-year follow-up period and subsequent hardware removal were enrolled in our study. We divided the patients into 2 groups, to compare the complications and demographics: 38 patients, treated with ZPLP plate; 31 patients, treated with other plates.
RESULTS: During 1 year of postoperative follow-up, 20 complications developed: 6 superficial infections at the operative site, 1 case of nonunion, 3 cases of osteitis, 4 cases of hypaesthesia, 2 cases of peripheral neuropathy, and 4 cases of metallosis. No statistical difference was found in the rate of complications when comparing the treatment groups (Mann-Whitney U test, p < .05) except for metallosis. All 4 patients who developed metallosis were treated using a ZPLP plate, and metallosis did not develop at all in patients who underwent surgery using other plates.
CONCLUSION: In our study, metallosis developed more than was previously known, particularly after using LCPs to treat lateral malleolar fractures. Our findings and those in recent publications support the possibility that metallosis can occur not only in patients with arthroplasties, but also in patients with open reduction and internal fixation with LCPs. Surgeons should be aware of such risk of metallosis and be careful to select proper plates for internal fixation.
© 2021. Springer-Verlag GmbH, DE part of Springer Nature.

Entities:  

Keywords:  Complications; Lateral malleolar fracture; Locking plate; Metallosis

Mesh:

Year:  2021        PMID: 33646355     DOI: 10.1007/s00402-020-03713-y

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   2.928


  17 in total

1.  The use of titanium and stainless steel in fracture fixation.

Authors:  J S Hayes; R G Richards
Journal:  Expert Rev Med Devices       Date:  2010-11       Impact factor: 3.166

2.  Massive metallosis due to metal-on-metal impingement in substitutive long-stemmed knee prosthesis.

Authors:  G Ottaviani; M A Catagni; L Matturri
Journal:  Histopathology       Date:  2005-02       Impact factor: 5.087

Review 3.  Metal implants and surface reactions.

Authors:  S G Steinemann
Journal:  Injury       Date:  1996       Impact factor: 2.586

4.  Diagnostic pitfalls in malignant gliomas: the analysis of misdiagnosis and current recommendations.

Authors:  Yu Wang; Yi Guo; Jun Gao; Kan Deng; Gui-lin Li; Min Feng; Jun-ji Wei; Zhi-qin Xu; Yong Yao; Wen-bin Ma; Yong-ning Li; Yi Yang; Ren-zhi Wang
Journal:  Chin Med J (Engl)       Date:  2012-12       Impact factor: 2.628

5.  Epidemiology of ankle fractures. A prospective population-based study of 212 cases in Aalborg, Denmark.

Authors:  S L Jensen; B K Andresen; S Mencke; P T Nielsen
Journal:  Acta Orthop Scand       Date:  1998-02

6.  Difference in metallic wear distribution released from commercially pure titanium compared with stainless steel plates.

Authors:  G D Krischak; F Gebhard; W Mohr; V Krivan; A Ignatius; A Beck; N J Wachter; P Reuter; M Arand; L Kinzl; L E Claes
Journal:  Arch Orthop Trauma Surg       Date:  2004-01-16       Impact factor: 3.067

Review 7.  Locking plates for extremity fractures.

Authors:  Jeffrey Anglen; Richard F Kyle; John Lawrence Marsh; Walter W Virkus; William C Watters; Michael Warren Keith; Charles M Turkelson; Janet L Wies; Kevin M Boyer
Journal:  J Am Acad Orthop Surg       Date:  2009-07       Impact factor: 3.020

8.  A Specialized Fibular Locking Plate for Lateral Malleolar Fractures.

Authors:  Eui Dong Yeo; Hak Jun Kim; Woo In Cho; Young Koo Lee
Journal:  J Foot Ankle Surg       Date:  2015-07-26       Impact factor: 1.286

Review 9.  Intraosseous metal implants in orthopedics: A review.

Authors:  L Rony; R Lancigu; L Hubert
Journal:  Morphologie       Date:  2018-10-19

10.  Orthogonal Double Plating and Autologous Bone Grafting of Postoperative Humeral Shaft Nonunion - A Rare Case Report and Review of Literature.

Authors:  Sreenivasulu Metikala; Prabhudheer Bhogadi
Journal:  J Orthop Case Rep       Date:  2015 Oct-Dec
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.