Transition From Manual to Automated Planning and Delivery of Volumetric Modulated Arc Therapy Stereotactic Radiosurgery: Clinical, Dosimetric, and Quality Assurance Results.

Richard A Popple1, Matthew H Brown2, Evan M Thomas2, Christopher D Willey2, Rex A Cardan2, Elizabeth L Covington2, Kristen O Riley3, James M Markert3, Markus Bredel2, John B Fiveash2.   

Abstract

PURPOSE: Properly planned single isocenter volumetric modulated arc therapy (VMAT) radiosurgery plans exhibit high quality and efficiency. We report here the largest clinical experience to date, to our knowledge, comparing manual planning with a new automated platform designed to standardize and simplify radiosurgery planning and delivery processes.
METHODS: We treated 693 patients with single isocenter VMAT radiosurgical plans generated by either our conventional manual (mVMAT) or a recently implemented automated (HyperArcTM) technique. All plans targeted the gross tumor volume without margin. Radiochromic film was used for patient-specific quality assurance (PSQA). We evaluated local control and toxicity data for a subgroup of 107 patients having 377 metastatic tumors that were treated with HyperArc.
RESULTS: The median Radiation Therapy Oncology Group (RTOG) conformity index was 1.14 and was not different between the 2 techniques. The median Paddick gradient index was 5.42 for HyperArc versus 7.09 for mVMAT (P < .001). The median mean brain doses were 4.6% and 5.1% for HyperArc and mVMAT, respectively (P = .04). The PSQA for both techniques met clinical criteria, but 97% of the HyperArc plans satisfied the gamma tolerance limit recommended by the American Association of Physicists in Medicine Task Group No. 218, compared with 94% of the mVMAT plans (P = .02). The median treatment-planning times were not significantly different. The median treatment times were 10.5 and 11.4 minutes for HyperArc and mVMAT, respectively (P < .001). The Kaplan-Meier estimate of local control was 90.1% at 1 year.
CONCLUSIONS: HyperArc produces high-quality radiosurgical plans that are at least as good as mVMAT plans created by an expert manual planner with easier planning and more efficient delivery workflow. A less experienced planner can produce very high-quality radiosurgical plans even for patients with more than 10 targets. The use of a single-isocenter technique for multiple targets with no PTV margin did not compromise clinical outcomes, and 1-year local control for treated targets remained congruent with historical series.
Copyright © 2020 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2021        PMID: 33632630     DOI: 10.1016/j.prro.2020.10.013

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  6 in total

1.  How to estimate R50% for cranial SRS/SRT cases with overlapping 50% isodose volumes: A proposed system.

Authors:  Dharmin D Desai; Ivan L Cordrey
Journal:  J Appl Clin Med Phys       Date:  2022-05-02       Impact factor: 2.243

2.  A measure of SRS/SRT plan quality: Quantitative limits for intermediate dose spill (R50%) in linac-based delivery.

Authors:  Dharmin D Desai; Ivan L Cordrey; E L Johnson
Journal:  J Appl Clin Med Phys       Date:  2022-03-02       Impact factor: 2.243

3.  Knowledge-based planning for the radiation therapy treatment plan quality assurance for patients with head and neck cancer.

Authors:  Wenhua Cao; Mary Gronberg; Adenike Olanrewaju; Thomas Whitaker; Karen Hoffman; Carlos Cardenas; Adam Garden; Heath Skinner; Beth Beadle; Laurence Court
Journal:  J Appl Clin Med Phys       Date:  2022-04-30       Impact factor: 2.243

4.  Analysis of a surface imaging system using a six degree-of-freedom couch.

Authors:  Xiaodong Zhao; Elizabeth L Covington; Richard A Popple
Journal:  J Appl Clin Med Phys       Date:  2022-07-12       Impact factor: 2.243

5.  HyperArcTM Dosimetric Validation for Multiple Targets Using Ionization Chamber and RT-100 Polymer Gel.

Authors:  Lucia Zirone; Elisa Bonanno; Giuseppina Rita Borzì; Nina Cavalli; Alessia D'Anna; Rosaria Galvagno; Andrea Girlando; Anna Maria Gueli; Martina Pace; Giuseppe Stella; Carmelo Marino
Journal:  Gels       Date:  2022-07-31

6.  Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc.

Authors:  Richard A Popple; Rodney J Sullivan; Yuan Yuan; Xingen Wu; Elizabeth L Covington
Journal:  J Appl Clin Med Phys       Date:  2021-11-01       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.