| Literature DB >> 33627984 |
Sweta Gupta1, Ashish Fauzdar1, Vikram Jeet Singh1, Ajay Srivastava1, Kamlesh Sharma1, Sabina Singh1.
Abstract
BACKGROUND: Electronic witness system (EWS) is one of the recent advancements in the field of in vitro fertilization (IVF) that uses radiofrequency identification (RFID) technology to monitor all critical work carried out in each stage of IVF procedures cycle.Entities:
Keywords: Assisted reproductive technology; biological mismatch; electronic witness system; in vitro fertilization; infertility; radio frequency identification technology
Year: 2020 PMID: 33627984 PMCID: PMC7879839 DOI: 10.4103/jhrs.JHRS_66_20
Source DB: PubMed Journal: J Hum Reprod Sci ISSN: 1998-4766
Figure 1Elaborate workflow chart as all defined fertilization procedures followed in the laboratory table corrections
Questionnnaire to assess the satisfaction level of clinician, clinical embryologist, and staff working at the center before and after installation of electronic witness system
| Serial number | Question | Electronic witnessing system |
|---|---|---|
| 1 | With the implementation of witnessing system, the possibility of human error in minimized, due you think it is going to help in your patient counseling and satisfaction? | |
| 2 | Since witnessing system can track each step of gamete manipulation in the IVF laboratory. How much do you think it can contribute to the success rates of your ART treatment? | |
| 3 | Now that you are aware of witness system, does it increase your preference or trust in an IVF center as a professional (consultant/embryologist) that have implemented this technological innovation? | |
| 4 | The witnessing system helps to identify if there is a sample mix up or another patient sample. Given this advantage in the process, how much confidence do you have in system? | |
| 5 | Since witnessing system track and check every step of sample manipulation in the IVF laboratory and hence there is extremely low possibility of omission of procedural steps. Does the implementation of such a system provide peace of mind to as a professional (consultant/embryologist)? |
Score: 0=No viewpoint, 1=Not at all, 2=A little, 3=Moderately, 4=A lot, 5=Extremely. IVF=In vitro fertilization, ART=Assisted reproduction technology
Summary of witness points and patient included in the study
| Measurement | Value |
|---|---|
| Number of tags used | 751 |
| Number of witness points | 803 |
| Number of patients seen | 187 |
| Number of cycle | 128 |
| Number of barcodes | 77 |
| Number of discards | 415 |
| Number of mismatches | 21 |
| Number of admin assign | 7 |
| Use/proximity (U) mismatches | 12 |
| True mismatches (T) | 2 |
| Mismatch identified by manual double witnessing | 2 |
Figure 2Pie chart with details of electronic witness system analytics of fertilization process done as per witness point according to defined workflow
Figure 3Details of number of procedures done per location
Data on secondary outcomes to evaluate the efficacy and safety of electronic witness system for maturation rate, fertilization rate, and cleavage rate followed formation of good quality day 3 embryo development rate and day 5 blastocyst development rate (day 5) compared to control groups
| Groups/cases ( | Number. of OCC retrieved | OCC/patient | Maturation rate (%) | Fertilization rate (%) | Cleavage rate (%) | Day-3 (Grade A) embryo development rate (%) | Day 5, blastocyst good-grade development rate (%) |
|---|---|---|---|---|---|---|---|
| Control (185) | 2263 | 12.23 | 86.04 | 75.60 | 100 | 60.73 | 28.60 |
| Positive control (408) | 5309 | 13.01 | 75.87 | 76.32 | 99.61 | 59.11 | 22.17 |
| EWS (56) | 695 | 12.41 | 83.16 | 74.39 | 93.95 | 64.60 | 26.48 |
EWS=Electronic witness system
Clinical pregnancy data of frozen embryo transfer cycles including of both day 3 and day 5 embryos in electronic witness cases in comparison to other groups
| FET ( | Number of embryos thawed | Embryo survival rate (%) | Implantation rate (%) Sac/embryo | Biochemical pregnancy rate (%) | Ongoing pregnancy rate (%) |
|---|---|---|---|---|---|
| Control (346) | 846 | 98.46 | 20.85 | 44.8 | 41.04 |
| Positive control (536) | 1364 | 96.92 | 22.94 | 48.88 | 44.78 |
| EWS (14) | 30 | 100 | 23.33 | 57.14 | 50 |
EWS=Electronic witness system, FET=Frozen embryo transfer
Figure 4Satisfaction graph of clinician, clinical embryologist, and staff personnel working at the center after installation of electronic witness system along with mean satisfaction index after installation of electronic witness system (4.5 ± 0.31)