Inion OTPS ™ absorbable plates are not appropriate for the fixation of isolated ulnar diaphyseal fractures.
Introduction
Isolated ulnar diaphyseal fractures are relatively rare [1]; however, they have a high complication rate [1, 2]. When located in the distal two-thirds of the ulna, with <50% displacement and less than 10 degrees angulation, they may be treated non-operatively [3]. All other types are considered unstable and are treated operatively with internal fixation [1, 2, 3].Although absorbable implants have been used as fixation devices, their use for treating isolated ulnar diaphyseal fractures in adults has not yet been studied [4, 5, 6, 7, 8]. The aim of this study was to assess the outcomes of fixation with third-generation Inion OTPS™ absorbable implants consisting of L-lactide, D, L-lactide, and trimethylene carbonate in a series of patients with isolated ulnar fractures. The mechanical stability of the implants, their complications, including late-presenting foreign body reactions, and the final clinical result were evaluated.
Materials and Methods
This case series study was approved by the institution’s ethics committee (74/16-4-2005). From April 2005 to December 2006, three patients with displaced, isolated ulnar diaphyseal fractures were enrolled (mean age 45 years old) (Fig. 1). Patients with complex fractures of both radius and ulna, osteoporosis, a history of other pathological or hereditary bone disorders that affect bone metabolism, and smokers were excluded from the study. All patients provided written informed consent. In January 2007, both the orthopaedic team and the ethics committee decided discontinuation of the study, due to early implant failure and no added benefit to the patients.
Figure 1
Isolated displaced fracture of the ulnar diaphysis; (a) anteroposterior, (b) lateral.
Isolated displaced fracture of the ulnar diaphysis; (a) anteroposterior, (b) lateral.
Implants
Six-hole extended plates and 3.1-mm screws from the Inion OTPS™ Biodegradable Plating System (Inion, Tampere, Finland) were used. The implants are proprietary copolymers blends based on L-lactide, D, L-lactide, and trimethylene carbonate. According to the manufacturer, plates lose mechanical strength 18–36 weeks postoperatively; complete resorption is expected to take place in 2–4 years. These implants are licensed for the use for internal fixation and have been approved by several quality management system standards (ISO 9001, ISO 13485, EN 46001, FDA 510K).
Operative procedure
All patients were operated by the senior author under general anesthesia and pneumatic tourniquet use. A longitudinal skin incision was done over the fracture site, with the patient supine, his arm on a hand table, and the elbow flexed at 90 degrees. After reduction, the fracture was stabilized with a lag screw when feasible, and the aforementioned plate was placed between the flexor carpi ulnaris and flexor carpi radialis, with three bicortical plate screws in each main fracture fragment (Fig. 2). No screw head breakage occurred during placement of the plate, for all patients.
Figure 2
Intraoperative images. (a) The ulnar fracture is visualized. (b) Insertion of the absorbable plate and screws after reduction of the fracture. (c) The internal fixation is finished and reduction of the fracture is achieved.
Intraoperative images. (a) The ulnar fracture is visualized. (b) Insertion of the absorbable plate and screws after reduction of the fracture. (c) The internal fixation is finished and reduction of the fracture is achieved.
Post-operative protocol
Radiographs were obtained immediately postoperatively (Fig. 3). Sugar-tong splint immobilization was maintained for 6 weeks to avoid rotation of the forearm and the patients were advised to initiate hand rehabilitation exercises immediately postoperatively; elbow exercises were scheduled after removal of the aforementioned splint.
Figure 3
Plain radiographs showing the anatomical reduction of the fracture; (a) anteroposterior, (b) lateral.
Plain radiographs showing the anatomical reduction of the fracture; (a) anteroposterior, (b) lateral.
Outcomes
Follow-up assessments were scheduled at 6 weeks, 3 months, 6 months, and 9 months. The measurements were conducted by an independent investigator, who did not participate in any of the procedures. The active and passive range of motion, as well as grip strength of both arms using a hydraulic dynamometer was measured (Jamar, NexGen Ergonomics Inc, Quebec, Canada). The patients’ functional outcome was evaluated using the disabilities of the arm, Shoulder and Hand Score (DASH Score) [9]. To exclude the possibility of delayed foreign body reactions, the patients were reassessed for signs of inflammation in an annual base, for 10 consecutive years.
Results
The patients’ demographics and comorbidities are shown in Table 1. All three patients reported pain and unexpected swelling few days after the operation and the implants’ failure was confirmed radiologically (Fig. 4). Reoperation using metallic implants was proposed as a definitive treatment for all patients. However, two of them denied it and were further treated with a plaster of Paris cast. The other patient, a 55 years old woman, underwent a revision fixation using a 3.5 metallic plate and screws. The fragmented absorbable plate was removed (Fig. 5), and a 3.5 mm metallic plate and screws were used for refixation. Subsequently, all fractures healed uneventfully (Fig. 6). The patients’ grip strength and DASH score at 9 months are presented in Table 2.
Table 1
Patients’ demographics
Figure 4
Implant failure and displacement of the ulnar fractures 8 days postoperatively; (a) anteroposterior, (b) lateral.
Figure 5
Revision fixation; (a) breakage of the implant (yellow arrow), (b) fixation with a 3.5 mm metallic plate and screws.
Figure 6
Plain radiograph showing the ulnar fracture healing at the 3-month follow-up visit.
Table 2
Patients’ grip strength and DASH score at 9 months
Patients’ demographicsImplant failure and displacement of the ulnar fractures 8 days postoperatively; (a) anteroposterior, (b) lateral.Revision fixation; (a) breakage of the implant (yellow arrow), (b) fixation with a 3.5 mm metallic plate and screws.Plain radiograph showing the ulnar fracture healing at the 3-month follow-up visit.Patients’ grip strength and DASH score at 9 monthsDuring the 10 years of clinical follow-up assessments, no severe soft-tissue reaction to the absorbable polymer materials was observed. In one patient, a small, palpable, painless mass over the fracture site was apparent at the 1st-year evaluation and remained at the same size since then (Fig. 7). Debridement was not decided because we did not notice any further signs of inflammation.
Figure 7
Photograph of the patient’s hand displaying a small mass over the fracture site at the 1st-year evaluation (yellow arrow).
Photograph of the patient’s hand displaying a small mass over the fracture site at the 1st-year evaluation (yellow arrow).
Discussion
The gold standard for treating displaced isolated ulnar diaphyseal fractures is open reduction and internal fixation with compression plates and screws [1, 2, 3]. Internal fixation using absorbable implants theoretically eliminates the need for implant removal operations and their potential complications [10, 11]. Although they have not gained widespread popularity among orthopedic surgeons, they still represent an area of continuing development [4, 5, 6, 7, 8, 12]. The outcomes of fixation with Inion OTPS™ absorbable plates and screws were assessed in a series of patients with isolated unstable ulnar diaphyseal fractures.The absorbable plates failed within the 1st month in all three cases, despite the protective splinting, showing that these plates cannot withstand the internal musculoligamentous mechanical forces of this particular area. According to the manufacturer, the elasticity modulus of self-reinforced absorbable polymers is close to that of cortical bone and higher than that of cancellous bone; the one-half of the initial bending and shear strength is lost after 18 weeks, which is an adequate time frame for bone healing per se [4, 13, 14]. The hypothesis that the sinusoidal movements of the elbow generate higher forces than the average may explain these problematic failures [15]. A potential solution to this problem would be to increase the thickness of the plates, but this could increase the foreign body reactions, which can be provoked with implant volumes that exceed the capacity of the surrounding soft tissues [16]. Recommendations regarding these products suggest using the minimum possible implant size to reduce the amount of breakdown products [17].A major drawback of absorbable implants is that they cause foreign body reactions during and beyond the degradation process, which sometimes require surgical debridement [18, 19]. Studies have shown a lower rate of such complications with the use of modern third-generation L-lactide, D,L-lactide, and trimethylene carbonate implants [20, 21, 22, 23]. However, several case series still report foreign body reactions, using the same polylactic acid implants, even 2 years after the operation [24, 25]. In the current study, no significant reaction was observed during the 10 years follow-up, which exceeds the 4 years maximum degradation period. The only soft-tissue reaction was a small painless mass with no signs of inflammation in one of the patients.The strong point of the current study is that the constant failure of the implants implies that their use is inappropriate for the specific anatomical area. Another strong point is the long follow-up of the patients. On the other hand, the study has several limitations. The most important is the small number of patients, which does not allow proper analysis and statistically significant results. However, early discontinuation of the study was considered ethically correct, due to no added benefit to the patients. Another limitation is that a control group treated with metallic implants was not used. However, there is a sufficient body of published information proving that metallic plates currently constitute the gold standard [1, 2, 3]. The theoretical advantages offered by absorbable implants are the elimination of a removal operation and the reduction of the stress shielding effect, so the study outcomes focused on these issues [10, 11].
Conclusion
The results of the present study suggest that Inion OTPS™ implants are not appropriate for the fixation of isolated unstable ulnar diaphyseal fractures. Implant failure was noted in the early post-operative period in all our cases. It seems that these specific implants cannot withstand the internal musculoligamentous mechanical forces of this anatomical area despite the protective splinting. No significant soft-tissue reactions were observed during and beyond the degradation period, other than a small painless mass in one case.Inion OTPS™ absorbable plating system is not appropriate for the fixation of isolated unstable ulnar diaphyseal fractures. Early implant failure was noted in all three cases proving that the specific implants cannot withstand the internal musculoligamentous mechanical forces of this anatomical area despite the protective splinting. No severe soft-tissue reactions were observed during the follow-up period.
Authors: P U Rokkanen; O Böstman; E Hirvensalo; E A Mäkelä; E K Partio; H Pätiälä; S I Vainionpää; K Vihtonen; P Törmälä Journal: Biomaterials Date: 2000-12 Impact factor: 12.479
Authors: H Wolfgang Losken; John A van Aalst; Mark P Mooney; Virginia L Godfrey; Tripti Burt; Sumeet Teotia; Shay B Dean; Jonathan R Moss; Reza Rahbar Journal: J Craniofac Surg Date: 2008-05 Impact factor: 1.046
Authors: Tuomo Nieminen; Immo Rantala; Ilmari Hiidenheimo; Jaakko Keränen; Heikki Kainulainen; Erkki Wuolijoki; Ilkka Kallela Journal: J Mater Sci Mater Med Date: 2007-08-15 Impact factor: 3.896
Authors: C J van Manen; M L Dekker; P V van Eerten; S J Rhemrev; G D J van Olden; M van der Elst Journal: Arch Orthop Trauma Surg Date: 2008-04-01 Impact factor: 3.067