Literature DB >> 33620619

Active esophageal cooling for the prevention of thermal injury during atrial fibrillation ablation: a randomized controlled pilot study.

Cory M Tschabrunn1,2, Sara Attalla3, Jonathan Salas4, David S Frankel4, Matthew C Hyman4, Emily Simon4, Tiffany Sharkoski4, David J Callans4, Gregory E Supple4, Saman Nazarian4, David Lin4, Robert D Schaller4, Sanjay Dixit4, Francis E Marchlinski4, Pasquale Santangeli4.   

Abstract

BACKGROUND: Severe endoscopically detected esophageal thermal lesions (EDELs) have been associated with higher risk of progression to atrio-esophageal fistula (AEF) following radiofrequency catheter ablation (RFCA) of atrial fibrillation (AF). We sought to evaluate safety and feasibility of active esophageal cooling using the Attune Medical Esophageal Heat Transfer Device (EnsoETM) to limit frequency or severity of EDELs.
OBJECTIVE: We sought To evaluate safety and feasibility of active esophageal cooling using the Attune Medical Esophageal Heat Transfer Device (EnsoETM) to limit frequency or severity of EDELs
METHODS: Consecutive patients undergoing first-time RFCA were randomized in a 1:1 fashion to esophageal cooling (device group) or standard temperature monitoring (control group). Ablation on the posterior wall was performed with a maximum power of 30W for up to 20s. All patients underwent EGD within 48 h. Endoscopy findings were classified as 1, erythema-mild injury; 2, superficial ulceration-moderate injury; 3, deep ulceration-significant injury; and 4, fistula/perforation. Severe EDELs were defined as grade 3 or 4 lesions.
RESULTS: Forty-four patients completed the study (22 device group, 22 control group). Adjunctive posterior wall isolation was performed more frequently in the device group (11/22, 50% vs. 4/22, 18%). EDELs were detected in 5/22 (23%) control group patients, with mild or moderate injury in 2/5 patients (40%) and severe thermal injury in 3/5 patients (60%). In the device group, EDELs were detected in 8/22 (36%) patients, with mild or moderate injury in 7/8 (87%) patients and severe thermal injury in 1/8 (12%) patients. There was no acute perforation or AEF during follow-up.
CONCLUSIONS: Active esophageal cooling may reduce the occurrence of severe EDELs. A larger randomized study is warranted to further evaluate the benefit of this strategy.
© 2021. Springer Science+Business Media, LLC, part of Springer Nature.

Entities:  

Keywords:  Ablation; Atrial fibrillation; Esophageal protection

Mesh:

Year:  2021        PMID: 33620619     DOI: 10.1007/s10840-021-00960-w

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  2 in total

1.  Preventing esophageal complications from atrial fibrillation ablation: A review.

Authors:  Lisa W M Leung; Zaki Akhtar; Mary N Sheppard; John Louis-Auguste; Jamal Hayat; Mark M Gallagher
Journal:  Heart Rhythm O2       Date:  2021-09-22

2.  Atrioesophageal fistula-A focus on preventing a devastating complication.

Authors:  Christopher Joseph; Jose Nazari; Mark Metzl
Journal:  JTCVS Tech       Date:  2022-04-12
  2 in total