| Literature DB >> 33615610 |
Maartje Leemans1, Sara H Muller1,2, Maarten J A van Alphen1, Wim Vallenduuk1, Richard Dirven1, Michiel W M van den Brekel1,3,4.
Abstract
BACKGROUND: Due to the heat and moisture exchanger's (HME) breathing resistance, laryngectomized patients cannot always use an (optimal) HME during physical exercise. We propose a novel HME cassette concept with adjustable "bypass," to provide adjustment between different breathing resistances within one device.Entities:
Keywords: HME cassette; breathing resistance; heat and moisture exchanger; pulmonary rehabilitation; total laryngectomy
Year: 2020 PMID: 33615610 PMCID: PMC8048959 DOI: 10.1002/hed.26571
Source DB: PubMed Journal: Head Neck ISSN: 1043-3074 Impact factor: 3.147
Specifications of the Moisture Loss and Pressure Drop values of the Provox XtraMoist (XM) and Provox XtraFlow (XF), as provided by the manufacturer (Atos Medical AB, Malmö, Sweden) in accordance with ISO 9360‐2:2001, and the humidification performance (Water Exchange) as reported by previous studies.
| Pressure Drop (Pa) | Moisture Loss | Water Exchange (mg) Van den Boer et al. (2014a) | Water Exchange (mg) Van den Boer et al. (2014b) | |||
|---|---|---|---|---|---|---|
| HME | At 30 L/min | At 60 L/min | At 90 L/min | At | At | At |
| Provox XtraMoist | 70 | 240 | 480 | 21.5 | 3.61 | 3.63 |
| Provox XtraFlow | 40 | 130 | 290 | 24.0 | 2.89 | 1.95 |
Note: The pressure drop of the XF at a flow of 60 L/min is approximately 60% of that of the XM. The humidification performance (Water Exchange) of the XF shows relatively less decline: approximately 80% of that of the XM.
Abbreviations: AH, chosen reference value for ambient humidity; HME, heat and moisture exchanger; ISO, International Organization for Standardization; V, tidal volume.
The lower the moisture loss value, the better the HME's humidification performance.
FIGURE 1The photo shows, from left to right, the original HME cassette of both the XF and XM with speaking valve (pink lid), the 3D‐printed (FormLabs, Form2) closed cassette with inserted XF foam and the 3D‐printed (FormLabs, Form2) cassette with bypass on the tracheal side, with inserted XM foam (note the difference in pore size between the two different foams). A speaking valve was not included in the 3D printed cassette designs to simplify the prototyping and to limit the scope of this proof of principle study to only the effect of the bypass. The thicker cylinder at the base of the 3D‐printed cassettes is used to connect them to the measurement set‐up (spirometer). HME, heat and moisture exchanger; XF, lower resistance/lesser humidification HME; XM, high resistance/high humidification HME [Color figure can be viewed at wileyonlinelibrary.com]
FIGURE 2The two HME cassette types. A, Design of the closed cassette for the XF foam measurements. B, Design of the closed cassette for the XM foam measurements. The bypass on the tracheal side of the cassette is closed off with a “twist‐ring.” C, 3D‐design of the cassette with opened bypass for the XM foam measurements. The specific bypass consists of eight d = 4 mm holes at the base of the cassette and can be opened or closed by adjusting the “twist‐ring.” D, “Twist‐ring” concept as seen on salt shakers. The bar at the base and the two small holes at the top of the cassettes, intended for inserting a pin, keep the HME foam in place during the measurements. The thicker cylinder at the base of the 3D‐printed cassettes is used to connect them to the measurement set‐up (spirometer). HME, heat and moisture exchanger; XF, lower resistance/lesser humidification HME; XM, high resistance/high humidification HME
Overview of the average resistance (pressure drop) and normalized humidification performance (water exchange) of the XM and XF foams in the two different cassette types.
| HME device | Pressure Drop in Pa (SD) | Water Exchange in mg (SD) | |||
|---|---|---|---|---|---|
| HME foam type | HME cassette type | At 30 L/min | At 60 L/min | At 90 L/min | At |
| XM foam | Closed cassette | 50 (2) | 158 (7) | 325 (13) | 5.70 (0.42) |
| Cassette with bypass | 29 (1) | 95 (5) | 201 (11) | 4.77 (0.40) | |
| XF foam | Closed cassette | 26 (1) | 93 (3) | 196 (4) | 4.91 (0.35) |
Note: The tidal volume (V ) and airflow rates of the pressure drop measurements correspond to the ISO standards (see Table 1). The different airflow rates of 30, 60, and 90 L/min represent approximately breathing at rest and during light and strenuous exercise. The SDs of the Water Exchange measurements of the HME devices are comparable to those previously reported by van den Boer et al. (2013). For the XF foam, a weighted mean and SD were calculated to represent the three different batches in equal proportion.
Abbreviations: AH, reference ambient humidity; AH, reference humidity at the tracheal side of the HME; F, flow; HME, heat and moisture exchanger; V, tidal volume; XF, lower resistance/lesser humidification HMESD, standard deviation; XM, high resistance/high humidification HME.
FIGURE 3Resistance (Pressure Drop at 60 L/min) against normalized humidificationperformance (Water Exchange at V = 1 L) of the different HME devices. The horizontal and vertical error bars indicate the standard deviations from the average Resistance and Water Exchange, respectively. Abbreviations: HME, heat and moisture exchanger; XF, lower resistance/lesser humidification HME; XM, high resistance/high humidification HME; V, tidal volume [Color figure can be viewed at wileyonlinelibrary.com]
Normalized input and verification data of different HMEs for the determination of the conversion from the normalized Water Exchange (WE) to Moisture Loss (ML).
| All values in mg/L |
|
|
| Ref. for | |
|---|---|---|---|---|---|
| Year | HME type | @32; 0 mg/L ( | @44; 0 mg/L ( | @44; 0 mg/L ( | |
| 2014 | Hiflow | 4.49 | 24.4 | 23.9 | Van den Boer et al. (2014b)2 |
| 2014 | Normal | 4.39 | 23.7 | 23.9 | Van den Boer et al. (2014b)2 |
| 2014 | XtraFlow | 4.49 | 24 | 23.9 | Van den Boer et al. (2014b)2 |
| 2014 | XtraMoist | 7.58 | 21.5 | 22.1 | Van den Boer et al. (2014b)2 |
| 2014 | Hiflow | 4.44 | 24.4 | 23.9 | Van den Boer et al. (2014a)1 |
| 2014 | Normal | 5.37 | 23.7 | 23.4 | Van den Boer et al. (2014a)1 |
| 2014 | XtraFlow | 5.92 | 24 | 23.1 | Van den Boer et al. (2014a)1 |
| 2014 | XtraMoist | 7.09 | 21.5 | 22.4 | Van den Boer et al. (2014a)1 |
| 2016 | XtraFlow | 5.01 | 24 | 23.6 |
|
| 2016 | XtraMoist | 6.79 | 21.5 | 22.6 |
|
| 2017 | XtraFlow | 4.41 | 24 | 23.9 |
|
| 2017 | XtraMoist | 5.58 | 21.5 | 23.3 |
|
Note: Normalized WEV () values at V = 1 L were calculated (see Appendix B.1 and B.2) from the values as measured by van den Boer et al. (2014a, 2014b).1,2 ML values were provided by the manufacturer (Atos Medical, Malmö, Sweden) in accordance with ISO 9360‐2:2001.3 ML was calculated from WE using Equation (21) and = 17.8 mg. For, abbreviations, see nomenclature in Appendix B.3.
Note: 1. van den Boer C, Muller SH, Vincent AD, van den Brekel MW, Hilgers FJ. Ex vivo assessment and validation of water exchange performance of 23 heat and moisture exchangers for laryngectomized patients. Respiratory Care. 2014; 59(8): 1161‐1171.
Note: 2. van den Boer C, Muller SH, Vincent AD, Züchner K, van den Brekel MWM, Hilgers FJM. Ex vivo water exchange performance and short‐term clinical feasibility assessment of newly developed heat and moisture exchangers for pulmonary rehabilitation after total laryngectomy. European Archives of Oto‐Rhino‐Laryngology. 2014;271(2):359‐366.
Note: 3. International Standards Organization. Anesthetic and respiratory equipment—heat and moisture exchangers (HMEs) for humidifying respired gases in humans. HMEs for use with tracheostomized patients having minimal tidal volume of 250 mL. Geneva: ISO; 9360‐2:2001.
The observations in 2016 and 2017 were made following the protocol of van den Boer (internal communication).
Comparison of the data measured in this study with the Water Exchange and Moisture Loss values (in accordance with ISO 9360‐2:2001)6 of the HMEs.
| All values in mg/L, at | Water Exchange/ | Moisture Loss, normalized to 44/0 mg/L | |||||
|---|---|---|---|---|---|---|---|
| This study | Van den Boer et al. (2014a)3 | Van den Boer et al. (2014b)2 | NKI‐AVL unpublished 2016 | NKI‐AVL unpublished (averaged over 2016, 2017 and 2018) | This study, calculated | Atos Medical | |
| Cassette foam | Narrow fitting, no speaking valve | with speaking valve | with speaking valve | with speaking valve | Narrow fitting, no speaking valve | Narrow fitting, no speaking valve | with speaking valve |
| XtraMoist | 5.70 | 5.98 | 6.40 | 5.73 | 5.47 | 22.6 | 21.5 |
| XtraFlow | 4.91 | 4.99 | 3.79 | 4.23 | n.a. | 23.2 | 24.0 |
Note: All Water Exchange data from Van den Boer et al.2,3 were normalized to AH = 5 mg/L and AH32 mg/L (Appendix B.1) and converted to V = 1 L (Appendix B.2). Van den Boer et al. measured in volunteers, so the actual AH during the measurements was about 32 mg/L. However, they performed their data normalization with an AH of 44 mg/L (AH in the alveoli of the lungs). Using the appropriate AH value, only has a minor impact on the results of van den Boer et al. (2014a and 2014b); the WE increases with approximately 4% and the rating of HMEs stays the same.]) Moisture Loss data for our HME devices were determined using Equation (21) (Appendix B.3.4). For comparison with the ML values, the table shows WE/V values (at V = 1 L, numerically equal to WE).
Abbreviations: NKI‐AVL, Netherlands Cancer Institute – Antoni van Leeuwenhoek (Amsterdam, The Netherlands); also see nomenclature in Appendix B.
Note: 1. International Standards Organization. Anesthetic and respiratory equipment—heat and moisture exchangers (HMEs) for humidifying respired gases in humans. HMEs for use with tracheostomized patients having minimal tidal volume of 250 mL. Geneva: ISO; 9360‐2:2001.
Note: 2. van den Boer C, Muller SH, Vincent AD, Züchner K, van den Brekel MWM, Hilgers FJM. Ex vivo water exchange performance and short‐term clinical feasibility assessment of newly developed heat and moisture exchangers for pulmonary rehabilitation after total laryngectomy. European Archives of Oto‐Rhino‐Laryngology. 2014;271(2):359‐366.
Note: 3. van den Boer C, Muller SH, Vincent AD, van den Brekel MW, Hilgers FJ. Ex vivo assessment and validation of water exchange performance of 23 heat and moisture exchangers for laryngectomized patients. Respiratory Care. 2014; 59(8): 1161‐1171.