Bogdana Tăbăcaru1, Horia Tudor Stanca1, Ruxandra Angela Pîrvulescu1, Simona Stanca2, Ciprian Danielescu3, Mihnea Munteanu4, Cosmin Roșca5, Adrian Cosmin Teodoru6. 1. Department of Ophthalmology, 'Carol Davila' University of Medicine and Pharmacy, 050474 Bucharest, Romania. 2. Department of Pediatrics, 'Carol Davila' University of Medicine and Pharmacy, 050474 Bucharest, Romania. 3. Department of Ophthalmology, 'Grigore T. Popa' University of Medicine and Pharmacy, 700115 Iasi, Romania. 4. Department of Ophthalmology, 'Victor Babeş' University of Medicine and Pharmacy, 300041 Timisoara, Romania. 5. Department of Ophthalmology, Oculens Clinic, 400501 Cluj-Napoca, Romania. 6. Department of Ophthalmology, Faculty of Medicine, 'Lucian Blaga' University, 550159 Sibiu, Romania.
Currently, laser-assisted in situ keratomileusis (LASIK) is the default choice for refractive surgical procedures, as it can address a wide spectrum of high and complex ametropias (1-5). There are numerous previous reports of hyperopic LASIK showing various safety, efficiency and predictability percentages (6-32). The new femtosecond laser technology to perform LASIK flaps avoids many complications of the mechanical microkeratome (free caps, incomplete, irregular or displaced flaps), being presently preferred by many surgeons (1,2,33). In this study, we evaluated the safety, efficacy, predictability and accuracy of the refractive results of Femtosecond-LASIK (Femto-LASIK) procedure using the VisuMax®-MEL® 80 platform for hyperopia with or without astigmatism.
Patients and methods
Data collection
We performed a retrospective, noncomparative consecutive case series study on eyes with hyperopia and hyperopic astigmatism that underwent Femto-LASIK surgery. Patients were operated on by the same refractive surgeon (HTS) at the Europe Eye-Metropolitan Hospital in Bucharest, Romania between June 2011 and June 2017. All surgeries were performed using the same femtosecond laser-excimer laser platform VisuMax®-MEL® 80 (Carl Zeiss Meditec).
Inclusion and exclusion criteria
The inclusion criteria for the surgery were: Patients of age ≥22 years, no refractive change for at least 2 years before surgery, stable peripheral retina (normal or already treated by laser photocoagulation if at-risk peripheral lesions were present), central endothelial density ≥2,000 cells/mm2 and good compliance (1,3,33,34).The refractive inclusion criteria were: Manifest hyperopia up to 6.00 diopter (D) with or without astigmatism up to 5.00 D and spherical equivalent +6.00 D at most. Patients outside these limits were referred for intraocular surgery, either phakic intraocular lenses or refractive lens exchange, according to patient age and ocular biometric considerations (35-37).We considered the following exclusion criteria for surgery: Eyes with corneal inadequate parameters (evidence or suspect of ectasia, thinnest point on pachymetry ≤500 µm, estimated postoperative steep keratometry >50 D, insufficient corneal thickness for laser ablation-estimated residual thickness of the stromal bed after treatment ≤300 µm) (1,3,33), eyes with anterior segment pathology (eg. severe dry eye syndrome, ocular inflammation or infection) (1,3,33), patients with eye-related conditions which might interfere with visual acuity (eg. cataract, congenital or acquired macular pathology, optic nerve pathology or retinal vascular pathology) (1,38-44), patients with previous ocular trauma or any previous ocular procedures (eg. scleral, vitreo-retinal surgery, glaucoma laser procedures or glaucoma surgery) (1,45-47) and patients taking medication with potential ocular side effects (eg. isotretinoin, amiodarone) (1,3,33).Patients with very deep-set eyes were also excluded, as well as patients with narrow palpebral fissures or periocular tumors (1,48-50), as normal orbital anatomy is important in order to permit the proper suction cup positioning.We also excluded patients with systemic diseases with risk of postoperative low visual acuity due to possible vascular complications including ischemic optic neuropathy or vascular occlusion (eg. cardiovascular diseases, severe systemic hypertension, severe dyslipidemia) (1,51-54) and patients with systemic diseases that could interfere with the wound-healing process (eg. autoimmune disorders, diabetes mellitus) (1,33,55).Pregnancy or lactation were exclusion criteria for the surgery (1,3,33).Patients unable to understand the perioperative management, patients with unreasonable expectations or patients unable to sign the informed consent were excluded as well (1,33).
Preoperative assessment
Preoperative ocular examination included: Corrected and uncorrected distance visual acuity, manifest, fogging and cycloplegic refractions, keratometry, ultrasound corneal pachymetry, corneal topography and tomography (Scheimpflug), non-contact tonometry, pupillometry, white-to-white corneal diameter, quantification of corneal endothelial cell density, slit-lamp examination of the anterior segment and retina assessment with pharmacological mydriasis.Patients were asked not to wear soft contact lenses 2 weeks prior to preoperative investigations and then 2 weeks prior to surgery.All patients signed an informed written consent form in accordance with the Declaration of Helsinki after being fully informed about the benefits and risks of the procedure. The study was approved by the Ethics Committee of ‘Carol Davila’ University of Medicine and Pharmacy of Bucharest, Romania (no. 16-02-06-2011).
Surgical technique
All surgeries were performed by the same refractive surgeon (HTS), with the same femtosecond laser-excimer laser platform (VisuMax®-MEL® 80; Carl Zeiss Meditec), using the same protocol and technique.Surgical steps included the following: Topical anesthesia, sterilization of the eyelids with 10% povidone-iodine solution, draping of the eyelids, positioning of the eye to be treated under the femtosecond laser surgical microscope, docking of the eye and proper suction, femtosecond laser assisted cutting of the corneal flap, repositioning of the eye under the excimer laser surgical microscope, lifting the flap, drying the corneal bed, excimer laser ablation, lavage of the debris with saline solution, repositioning of the flap, bandage contact lens application and instillation of topical antibiotic and artificial tears.The flap position, the flap regularity and the interface clarity were examined before patient discharge.
Postoperative care
Postoperative treatment consisted of topical eye drops: Antibiotic, steroid and non-steroid anti-inflammatory and artificial tears. The bandage contact lens was removed on the first day of postoperative visit.The follow-up visits were carried out at 1, 3, 6 and 12 months. For each examination, a slit-lamp examination of the anterior segment and several investigations were performed which included: Manifest refraction, uncorrected distance visual acuity, corneal topography and tomography (Scheimpflug) and non-contact tonometry. Corrected distance visual acuity and cycloplegic refraction were performed for the eyes in which a residual refraction was determined or the visual acuity was uncorrelated with the manifest refraction.
Data analysis and statistics
Patient data were centralized into an Excel® (ver.14.0, Microsoft Corp.) database after being collected on case forms. Statistical analysis was performed using Statistical Package for the Social Sciences (SPSS) software (ver. 20, IBM® SPSS® Statistics; IBM Corp.). Normality of continuous variables distribution was checked by the Shapiro-Wilk test, the statistically significance being set at a P-value <0.05. Preoperative and postoperative refractive data are presented as mean ± standard deviation and range.Refractive outcome analysis was performed according to the Standard Graphs for Reporting Refractive Surgery (56,57). Thus, we compared the preoperative data with the 1-, 3-, 6- and 12-month postoperative results and we analyzed the safety, the efficacy, the predictability, the accuracy and the stability of the Femto-LASIK procedure.
Results
Patient demographics and operative data
A total of 593 eyes (290 right eyes and 303 left eyes) from 373 patients (294 males and 79 females) were available for evaluation 12 months after laser surgery and were reviewed for statistical analysis in our retrospective interventional consecutive case series study. The mean patient age at the time of surgery was 33.20±8.098 years (range, 22-49 years). A total of 224 patients (448 eyes) had bilateral Femto-LASIK, while 153 patients were anisometropic cases, being operated in one single eye. A total of 163 eyes (27.49%) were amblyopic, having a visual acuity between 20/60 and 20/30. The preoperative manifest and cycloplegic refraction data are presented in Table I.
Table I
Preoperative manifest and cycloplegic refraction data.
Parameter
Diopters (mean ± SD, range)
Manifest sphere
+3.9532±1.5282, +1.25 to +6.00
Manifest cylinder
+2.6316±1.2208, +0.50 to +5.00
Manifest SEQ
+4.0780±1.5723, +1.50 to +6.00
Cycloplegic sphere
+5.7198±1.4351, +3.25 to +6.75
Cycloplegic cylinder
+3.1320±1.4829, +0.50 to +5.00
SD, standard deviation; SEQ, spherical equivalent refraction.
Visual and refractive results
The efficacy was defined as the ratio of the uncorrected distance visual acuity (UDVA) at the end of the follow-up period (12 months) and the preoperative corrected visual acuity (CDVA). Fig. 1 shows that all eyes reached a UDVA of 20/63 or better 12 months after Femto-LASIK surgery. As shown in Fig. 2, for 84.1% of the eyes, the postoperative 12-month UDVA was the same or better than the preoperative CDVA.
Figure 1
Summary of the 12-month postoperative uncorrected distance visual acuity (UDVA) and preoperative corrected distance visual acuity (CDVA). Refractive target was emmetropia (plano target). Postop, postoperative; Preop, preoperative.
Figure 2
Efficacy, shown as gained and lost Snellen lines of postoperative uncorrected distance visual acuity (UDVA) compared to preoperative corrected distance visual acuity (CDVA), 12-months postoperative. Postop, postoperative.
Safety, defined as no loss of two or more Snellen lines of CDVA, was excellent. Twelve months after Femto-LASIK surgery, no eye lost 2 lines and 6 eyes (1.012%) lost one Snellen line of CDVA (Fig. 3). Contrariwise, we found several amblyopic eyes that gained at least one line of visual acuity, reflecting potential benefit of the refractive surgery, possibly explained by reduction or elimination of optical aberrations of the hyperopic magnifying correction lenses.
Figure 3
Safety, shown as gained and lost Snellen lines of postoperative corrected distance visual acuity (CDVA) compared to preoperative CDVA, 12-months postoperative. Postop, postoperative.
The scatterplot of the attempted spherical equivalent refraction (SEQ) against the achieved refractive change is shown in Fig. 4. At the postoperative 12 months follow-up visit, the mean sphere was +0.887±0.8746 D, the mean cylinder was +0.766±0.5797 and the mean SEQ was +0.5042±0.8336. The accuracy of SEQ at 12 months was 49% within ±0.50 D and respectively 74% within ±1.00 D of emmetropia (Fig. 5). Lower predictability and accuracy of SEQ was found for the eyes with high preoperative refractive errors.
Figure 4
Scattergram of attempted vs. achieved spherical equivalent refraction (SEQ), 12-months postoperative. Postop, postoperative; D, diopter.
Figure 5
Accuracy of spherical equivalent refraction (SEQ) to intended target, 12-months postoperative. Postop, postoperative; D, diopter.
Regarding the mean refractive manifest astigmatism at the last postoperative visit, 54 and 79% were <0.50 D and ≤1.00 D, respectively. Fig. 6 also shows the comparison between the preoperative and the 12-month postoperative refractive astigmatism.
Figure 6
Summary of preoperative and 12-months postoperative refractive astigmatism. Postop, postoperative; Preop, preoperative; D, diopter.
Fig. 7 shows the excellent evolution of SEQ after the surgery and its stability in time over the 12-month follow-up period. The postoperative refraction data at 1, 3, 6 and 12 months postoperative are presented in Table II. There were no intraoperative or postoperative complications during the follow-up period.
Figure 7
Stability of spherical equivalent refraction (SEQ) during the follow-up period of 12 months. Postop, postoperative.
Table II
Postoperative refraction data.
Postoperative visit
1 month
3 months
6 months
12 months
Parameter
Diopters (mean ± SD, range)
Manifest sphere
+0.233±1.1409, -1.50 to +3.25
+0.4393±1.1563, -1.25 to +3.75
+0.667±0.9742, -1.00 to +2.50
+0.887±0.8746, -1.00 to +2.75
Manifest cylinder
+0.91±0.739, +0.00 to +3.00
+0.8735±0.6021, +0.25 to +2.75
+0.7618±0.54706, +0.00 to +2.75
+0.766±0.5797, +0.25 to +3.25
Manifest SEQ
-0.222±1.1320, -2.50 to +3.00
+0.00253±1.079, -1.50 to +3.375
+0.28605±0.9382, -1.25 to +2.50
+0.5042±0.8336, -1.00 to +3.00
SD, standard deviation; SEQ, spherical equivalent refraction.
Discussion
Laser-assisted in situ keratomileusis (LASIK) is currently the most frequently practiced laser refractive procedure worldwide as it can correct large diopter limits of all types of refractive errors (1-4,6-33). However, laser refractive surgery is challenging for eyes with hyperopia or hyperopic astigmatism, especially when refractive error is high (1,6-32,58).Hyperopic correction has several difficulties. Hyperopic patients are frequently under-corrected in regards to glass prescriptions prior to preoperative evaluation and have accommodation reserve of different degrees or accommodation spasm (1,6). Therefore, the integration of manifest and cycloplegic refractions is difficult and latent hyperopia may lead to increasing values of manifest refraction with age (1,6). Accurate centration of the ablation is another difficult point as hyperopic patients have a wide angle Kappa (6). Hyperopic correction induces central corneal flattening, which may be limited by preoperative keratometry (6). Another issue concerning hyperopic and astigmaticpatients is the higher risk of de-centration of ablation due to the difficulty in fixation of the near target point during laser correction (6). Considering all the above listed features, the predictability and the accuracy of the refractive results in hyperopia, with or without astigmatism, are lower when compared to myopic refractive corrections (58).Although some excimer lasers have been approved for corrections up to +6.0 D (1,7) and current technology using models with wide ablation and high correction speed has greatly improved prognosis (1,8) [large diameter optical areas being more resistant to epithelial hyperplasia that is responsible for real regression (1,7)], latent hyperopia is often the reason why refractive surgeons avoid approaching cases of hyperopia and hyperopic astigmatism or limit the surgeries to refractive errors below +3.0 D (1,8). Laser refractive treatment is difficult to choose because ablation of manifest refraction in hyperopia or hyperopic astigmatism may lead postoperatively to recurrence of a degree of hyperopia, falsely interpreted as regression of the laser procedure (1,6,7), while ablation of cycloplegic refraction may lead to myopic refraction in the immediate postoperative period, causing an unsatisfactory UDVA (1). In order to compensate in part for the latent hyperopia, Kanellopoulos proposed a nomogram that involves full treatment in the case of the cylinder and the ablation of the manifest refraction sphere with an addition of +0.25 D for the dominant eye and up to +0.75 D for the non-dominant eye (1,9).There have been several reports of hyperopic-LASIK and hyperopic astigmatism-LASIK in the past, using different excimer laser platforms and using for the flap creation either the microkeratome or the femtosecond laser (1,6-32). Table III shows the follow-up periods and the refractive reports of all of these studies either with high (1,6-23,25-29) and lower predictability and efficacy rates for high refractive errors (1,6,24,30). The safety of the procedure in these reports was reported as good; the number of eyes that lost two lines of CDVA postoperatively being either zero (1,6,8,25,29) or ranging between 1.09 and 4.34% (1,20,27,28,30,32). The reports in the literature mention that the improvement of the ametropic error allowed the gain in CDVA of one line in 44.6% of cases (1,9) and of two lines in 3.7 to 22.8% of cases in various studies (1,8,28).
Table III
Predictability of the SEQ between ±0.5 D and ±1.0 D of the intended target, in several published reports of LASIK interventions for hyperopia or hyperopic astigmatism (1).
Authors (Ref.)
Refractive error of the treated eyes
Follow-up period (months)
±0.5 D of the intended target
±1.0 D of the intended target
Zadok et al (20)
Hyperopia (SEQ <+3)
6
88.9%
Salz and Stevens (32)
Hyperopia
6
65.0%
87.4%
Pineda-Fernández et al (30)
Low hyperopia
6
54.5%
Zadok et al (20)
Hyperopia (SEQ <+5)
6
51.8%
Pineda-Fernández et al (30)
Medium hyperopia
6
40.0%
Kanellopoulos et al (25)
Hyperopia (SEQ <+3)
12
92.0%
Salz and Stevens (32)
Hyperopia
12
74.1%
91.4%
Roesler and Kohnen (31)
Hyperopia
12
88.0%
Kanellopoulos et al (25)
Hyperopia (SEQ <+5)
12
79.0%
Ditzen et al (27)
Hyperopia
12
78.0%
Kanellopoulos (9)
Hyperopia and Hyperopic astigmatism
6
77.9%
88.9%
El-Naggar and Hovaghimian (6)
Hyperopia and Hyperopic astigmatism
12
95%
100.0%
Kanellopoulos et al (25)
Hyperopia (SEQ >+5) and Hyperopic astigmatism
12
71.0%
Lian et al (28)
Hyperopia and Hyperopic astigmatism
12
61.1%
83.3%
Kanellopoulos (9)
Hyperopia and Hyperopic astigmatism
24
75.5%
94.4%
Pineda-Fernández et al (30)
Hyperopic astigmatism (SEQ <+3)
6
50.0%
Pineda-Fernández et al (30)
Hyperopic astigmatism (SEQ <+6)
6
38.8%
Salz and Stevens (32)
Hyperopic astigmatism
6
60.5%
88.7%
Salz and Stevens (32)
Hyperopic astigmatism
12
73.0%
89.2%
Ditzen et al (27)
Hyperopic astigmatism
12
42.0%
SEQ, spherical equivalent refraction.
In the present study, we assessed the refractive outcomes of femtosecond assisted LASIK performed for the correction of hyperopia up to 6 D without or with astigmatism up to 5 D, with a spherical equivalent not exceeding +6 D. The refractive target was emmetropia in all cases. The treatment plan was chosen corroborating the manifest, fogging and cycloplegic refraction, considering both the accommodative reserve and patient age.Femtosecond-LASIK using the VisuMax®-MEL® 80 platform was demonstrated to be a suitable option to correct selected cases of hyperopia and hyperopic astigmatism. The postoperative results after one year demonstrated the Femto-LASIK procedure to be safe and effective. The predictability at 12 months was 74% within ±1 D of emmetropia. As both the sphere and the cylinder plus values were reduced and the need for hyperopic correction lenses was minimal, the optical aberrations and distortion of the retinal image were smaller, allowing us to achieve a better postoperative CDVA with at least one line in 9% of the eyes.The retrospective pattern of this report and the short period of follow-up of one year are limiting factors in our study. A future prospective study and a longer follow-up period are necessary for a better understanding of the procedure stability and for elaboration of nomogram ablation profiles.