Literature DB >> 33563051

New Clamp-PID Algorithm for Automated Glucose Clamps Improves Clamp Quality.

Carsten Benesch1, Mareike Kuhlenkötter1, Leszek Nosek1, Tim Heise1.   

Abstract

BACKGROUND: In automated glucose clamp experiments, blood glucose (BG) concentrations are kept close to a predefined target level using variable glucose infusion rates (GIRs) determined by implemented algorithms. Clamp quality (ie, the ability to keep BG close to target) highly depends on the quality of these algorithms. We developed a new Clamp algorithm based on the proportional-integral-derivative (PID) approach and compared clamp quality between this and the established Biostator (BS) algorithm.
METHODS: In numerical simulations, the PID-based algorithm was optimized in silico. The optimized Clamp-PID algorithm was tested in in vitro experiments and finally validated in vivo in a small (n = 5) clinical study.
RESULTS: In silico, in vitro, and in vivo experiments showed better clamp quality for the new Clamp-PID algorithm compared with the BS algorithm: precision and absolute control deviation (ACD) decreased from 3.7% to 1.1% and from 2.9 mg/dL to 0.6 mg/dL, respectively, in the numerical simulation. The in vitro validation demonstrated reductions in precision (from 3.3% ± 0.1% (mean ± SD) to 1.4% ± 0.4%) and in ACD (from 2.3 mg/dL ± 0.4 mg/dL to 0.8 mg/dL ± 0.2 mg/dL), respectively. In the clinical study, precision and ACD improved from 6.5% ± 1.3% to 4.0% ± 1.1% and from 3.6 mg/dL ± 0.9 mg/dL to 2.2 mg/dl ± 0.6 mg/dl, respectively. The quality parameter utility did not change.
CONCLUSIONS: The new Clamp-PID algorithm improves the clamp quality parameters precision and ACD versus the BS algorithm.

Entities:  

Keywords:  ClampArt; PID algorithm; automated glucose clamp; clamp quality; pharmacodynamics; type 1 diabetes

Mesh:

Substances:

Year:  2021        PMID: 33563051      PMCID: PMC8861780          DOI: 10.1177/1932296821991514

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  10 in total

1.  Algorithms for a closed-loop artificial pancreas: the case for model predictive control.

Authors:  B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

2.  How to Assess the Quality of Glucose Clamps? Evaluation of Clamps Performed With ClampArt, a Novel Automated Clamp Device.

Authors:  Carsten Benesch; Tim Heise; Oliver Klein; Lutz Heinemann; Sabine Arnolds
Journal:  J Diabetes Sci Technol       Date:  2015-04-07

3.  Improved Algorithm for Automated Glucose Clamps.

Authors:  Mareike Kuhlenkötter; Tim Heise; Carsten Benesch
Journal:  Diabetes Technol Ther       Date:  2016-12-22       Impact factor: 6.118

4.  Randomized Crossover Comparison of Personalized MPC and PID Control Algorithms for the Artificial Pancreas.

Authors:  Jordan E Pinsker; Joon Bok Lee; Eyal Dassau; Dale E Seborg; Paige K Bradley; Ravi Gondhalekar; Wendy C Bevier; Lauren Huyett; Howard C Zisser; Francis J Doyle
Journal:  Diabetes Care       Date:  2016-06-11       Impact factor: 19.112

Review 5.  Euglycaemic glucose clamp: what it can and cannot do, and how to do it.

Authors:  Tim Heise; Eric Zijlstra; Leszek Nosek; Sascha Heckermann; Leona Plum-Mörschel; Thomas Forst
Journal:  Diabetes Obes Metab       Date:  2016-07-13       Impact factor: 6.577

6.  Fitting nonlinear regression models with correlated errors to individual pharmacodynamic data using SAS software.

Authors:  R Bender; L Heinemann
Journal:  J Pharmacokinet Biopharm       Date:  1995-02

Review 7.  Glucose clamps with the Biostator: a critical reappraisal.

Authors:  L Heinemann; F J Ampudia-Blasco
Journal:  Horm Metab Res       Date:  1994-12       Impact factor: 2.936

8.  Pharmacokinetic and Glucodynamic Responses of Ultra Rapid Lispro vs Lispro Across a Clinically Relevant Range of Subcutaneous Doses in Healthy Subjects.

Authors:  Jennifer Leohr; Mary Anne Dellva; Elizabeth LaBell; David E Coutant; Oliver Klein; Leona Plum-Moerschel; Eric Zijlstra; Helle Linnebjerg
Journal:  Clin Ther       Date:  2020-09-06       Impact factor: 3.393

9.  Development of the Biostator Glucose clamping algorithm.

Authors:  A H Clemens; D L Hough; P A D'Orazio
Journal:  Clin Chem       Date:  1982-09       Impact factor: 8.327

10.  Design and Evaluation of a Robust PID Controller for a Fully Implantable Artificial Pancreas.

Authors:  Lauren M Huyett; Eyal Dassau; Howard C Zisser; Francis J Doyle
Journal:  Ind Eng Chem Res       Date:  2015-06-09       Impact factor: 3.720

  10 in total
  1 in total

1.  How to Achieve Sufficient Endogenous Insulin Suppression in Euglycemic Clamps Assessing the Pharmacokinetics and Pharmacodynamics of Long-Acting Insulin Preparations Employing Healthy Volunteers.

Authors:  Hui Liu; Ting Li; Hongling Yu; Jiaqi Li; Huiwen Tan; Yerong Yu
Journal:  Front Pharmacol       Date:  2022-07-22       Impact factor: 5.988

  1 in total