Isoelectric point determination by imaged CIEF of commercially available SARS-CoV-2 proteins and the hACE2 receptor.

Finja Krebs1, Christin Scheller1, Kristina Grove-Heike1, Lena Pohl1, Hermann Wätzig1.   

Abstract

In order to contribute to the scientific research on the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), we have investigated the isoelectric points (pI) of several related proteins, which are commercially available: the receptor-binding domain (RBD) with His- and Fc-tag, the S1 subunit with His-tag, the S1/S2 subunits with His-tag and the human angiotensin-converting enzyme 2 (hACE2) with His-tag. First, the theoretical pI values, based on the amino acid (AA) sequences of the proteins, were calculated using the ProtParam tool from the Bioinformatics Resource Portal ExPASy. The proteins were then measured with the Maurice imaged CIEF system (native fluorescence detection), testing various measurement conditions, such as different ampholytes or ampholyte mixtures. Due to isoforms, we get sections with several peaks and not just one peak for each protein. The determined pI range for the RBD/Fc is 8.24-9.32 (theoretical pI: 8.55), for the RBD/His it is 7.36-9.88 (8.91) and for the S1/His it is 7.30-8.37 (7.80). The pI range of the S1/S2/His is 4.41-5.87 (no theoretical pI, AA sequence unknown) and for hACE2/His, the determined global range is 5.19-6.11 (5.60) for all experimental conditions chosen. All theoretically derived values were found within these ranges, usually close to the center. Therefore, we consider theoretical values as useful to make predictions about the isoelectric points of SARS-CoV-2 proteins. The experimental conditions had only a minor influence on the pI ranges obtained and mainly influenced the peak shapes.
© 2021 The Authors. Electrophoresis published by Wiley-VCH GmbH.

Entities:  

Keywords:  CIEF; COVID-19; Isoelectric point; SARS-CoV-2 proteins; hACE2

Mesh:

Substances:

Year:  2021        PMID: 33533060      PMCID: PMC8013610          DOI: 10.1002/elps.202100015

Source DB:  PubMed          Journal:  Electrophoresis        ISSN: 0173-0835            Impact factor:   3.535


human angiotensin‐converting enzyme 2 receptor‐binding domain severe acute respiratory syndrome coronavirus 2 The Coronavirus Disease 2019 (COVID‐19) pandemic is currently shaking up the whole world. Since the beginning of the outbreak, research has been conducted into medicines and vaccines against the virus in the field of pharmaceutical research. For research purposes, there are several significant proteins of the severe acute respiratory syndrome coronavirus 2 (SARS‐CoV‐2) on the market. Therefore, our intention was to contribute to the progress of research in this field with the means available to us. In this short communication, the results of the capillary isoelectric focusing (CIEF) measurements of different sections of the SARS‐CoV‐2 spike protein and the human angiotensin‐converting enzyme 2 (hACE2) are presented. SARS‐CoV‐2 uses the receptor‐binding domain (RBD) of its surface glycoprotein (spike protein) to enter the human body via the hACE2 receptor [1]. The spike protein is comprised of two functional subunits: S1 and S2. While the S1 subunit binds to the hACE2 receptor, the S2 subunit is implicated in the merging of the viral and human cell membranes [2]. For more information on the physicochemical properties of SARS‐CoV‐2, please refer to the review of Scheller et al. [3]. In the CIEF, the isoelectric focusing takes place in a capillary, as a charge‐based analysis via capillary electrophoresis. The CIEF method was implemented by Hjerten et al. in the 1980s and can be seen as an improved, new version of the conventional isoelectric focusing in slab gels [4, 5]. The advantages of CIEF over IEF with slab gels are, for example, a smaller sample volume required, shorter analysis times, and higher sensitivity [6]. Using CIEF, a pH gradient is built up in the capillary. The ampholytic analytes migrate in this pH gradient and then remain in the pH zone that corresponds to their pI value [7]. It is also possible in CIEF to use immobilized pH gradients as applied in slab gels. This technique is still under development, but can bring many advantages [6, 8–10]. In this case, the imaged CIEF with carrier ampholytes was used, i.e., an on‐line imaging detection system that does not require the mobilization of the analytes after focusing [11]. Since the isoelectric point of proteins is an important property, we sought to determine it using the Maurice (imaged) CIEF system from ProteinSimple, a Bio‐Techne brand. We used the SARS‐CoV‐2 proteins currently available on the market, namely the RBD with His‐ and Fc‐tag, the S1 subunit with His‐tag, and the S1/S2 subunits with His‐tag. As the virus binds to the hACE2 in the human body, this receptor was included in the study (all proteins were provided by R&D Systems, a Bio‐Techne brand, Minneapolis, Minnesota, USA). First, the theoretical pI values of the proteins were calculated using the ProtParam tool from the Bioinformatics Resource Portal ExPASy [12]. The basis for this was the given AA sequence of the corresponding protein, which can be found in the R&D Systems catalog [13] on the page of the respective protein (see catalog numbers in Table 1).
Table 1

Experimental conditions for the determination of the isoelectric points via Maurice CIEF of the SARS‐CoV‐2 proteins and the hACE2 receptor from R&D Systems; sample Load: 55 s, anolyte: 80 mM phosphoric acid in 0.1% methyl cellulose, catholyte: 100 mM NaOH in 0.1% methyl cellulose, Vwell: 100 μL, 40% SimpleSol Protein Solubilizier (except for experiment f of hACE2/His), 0.35% methyl cellulose, 10 mM arginine (except for experiment f of hACE2/His); ρ = mass concentration of the respective protein in mg/mL

RBD/FcRBD/HisS1/HisS1/S2/HishACE2/His
Catalog number #10499‐CV#10500‐CV#10522‐CV#933‐ZV
ρinitial [mg/mL] 2.1001.8600.3500.4700.468
Buffer PBS pH 7.4PBS pH 7.4PBS pH 7.4PBS pH 7.412.5 mM Tris, NaCl, ZnCl2 and glycerol
Theoretical pI 8.558.917.805.60
Separation 1 min 1500 V1 min 1500 V1 min 1500 V1 min 1500 V1 min 1500 V
7.5 min 3000 V7.5 min 3000 V8.5 min 3000 V11 min 3000 Va‐e: 20 min, f: 15 min 3000 V
Fluorescence exposure time 50 s50 s50 s50 sa: 5 s, b: 50 s, e: 30 s, f: 10 s
ρfinal [mg/mL] 0.0360.0350.0400.080

a: 0.040

b: 0.016

e: 0.041

f: 0.098

Urea [mol/L]

e: 2

f: 7

Ampholyte a/b: 3–10 (4%)a/b: 3–10 (4%)a/b: 3–10 (4%)a: 3–10 (4%)a/b/e: 3–10 (4%)

c: 3–10 (1%),

8–10.5 (3%)

d: 8–10.5 (4%)

c: 3–10 (1%),

8–10.5 (3%)

c: 3–10 (1%),

8–10.5 (3%)

d: 8–10.5 (4%)

f: 2.5–5 (3%), 5–8 (2%)
pI markers (1% each)

a: 3.38, 9.99

b/c/d: 7.05, 9.50

a: 3.38, 9.99

b/c: 7.05, 10.17

a: 3.38, 9.99

b/c/d: 7.05, 9.50

a: 3.38, 9.99

a: 3.38, 9.99

b/e/f: 4.05, 7.05

Experimental conditions for the determination of the isoelectric points via Maurice CIEF of the SARS‐CoV‐2 proteins and the hACE2 receptor from R&D Systems; sample Load: 55 s, anolyte: 80 mM phosphoric acid in 0.1% methyl cellulose, catholyte: 100 mM NaOH in 0.1% methyl cellulose, Vwell: 100 μL, 40% SimpleSol Protein Solubilizier (except for experiment f of hACE2/His), 0.35% methyl cellulose, 10 mM arginine (except for experiment f of hACE2/His); ρ = mass concentration of the respective protein in mg/mL a: 0.040 b: 0.016 e: 0.041 f: 0.098 e: 2 f: 7 c: 3–10 (1%), 8–10.5 (3%) d: 8–10.5 (4%) c: 3–10 (1%), 8–10.5 (3%) c: 3–10 (1%), 8–10.5 (3%) d: 8–10.5 (4%) a: 3.38, 9.99 b/c/d: 7.05, 9.50 a: 3.38, 9.99 b/c: 7.05, 10.17 a: 3.38, 9.99 b/c/d: 7.05, 9.50 a: 3.38, 9.99 b/e/f: 4.05, 7.05 The SARS‐CoV‐2 RBD/Fc is derived from the surface glycoprotein (NCBI Reference Sequence: YP_0 097 24390.1) Arg319Phe541, which is linked to human IgG1 (Pro100‐Lys330) via the sequence IEGRMD. The pI value calculated using this sequence is 8.55. The His‐tagged SARS‐CoV‐2 RBD, with a theoretical pI of 8.91, is also the surface glycoprotein Arg319Phe541, but with a C‐terminal 6‐His‐tag. The S1/His subunit is also the surface glycoprotein, but the sequence Val16‐Pro681, with a C‐terminal 6‐His‐tag. A pI of 7.80 can be calculated from this sequence. The S1/S2/His protein is not yet commercially available and therefore not included in the R&D Systems catalog. For this reason, no theoretical pI value can be calculated as the exact AA sequence is unknown. The hACE2/His receptor consists of the AA sequence of the hACE2 protein (UniProtKB/Swiss‐Prot: Q9BYF1) Gln18Ser740, with a C‐terminal 10‐His‐tag. The pI value of this sequence calculated via ExPASy is 5.60. Subsequently, it is investigated to what extent the theoretical pI values agree with the experimentally measured ones. For the Maurice CIEF system, the Maurice CIEF cartridges and the Maurice CIEF method development kit (including all reagents used, such as the anolyte, catholyte, ampholytes, pI markers, SimpleSol, etc.) were used (provided by ProteinSimple, a Bio‐Techne brand, San Jose, California, USA). With native fluorescence at an excitation wavelength of 280 nm, emitted light at 320–450 nm was used for detection [14]. The measurements were performed under the experimental conditions shown in Table 1. The preparation of the samples is now explained using the sample from experiment a of the RBD/Fc as an example. 8.5 μL of protein, dissolved in PBS buffer (ρinitial = 2.100 mg/mL), was mixed with 35 μL of 1% methyl cellulose, 40 μL SimpleSol, 4 μL ampholyte 3–10, 2 μL of 500 mM arginine, 1 μL each of pI markers 3.38 and 9.99 and 8.5 μL DI water, resulting in a total volume of 100 μL. Samples were then vortexed and centrifuged to be pipetted into a 96‐well plate. In order to get a general overview of the pI value of the respective protein, the wide range ampholyte with markers at the edges of the range was first selected (a). In a second step, pI markers whose pI values were as close as possible to those of the respective protein were chosen (b). Next step was to improve the resolution of the peaks by using ampholyte mixtures. For this purpose, wide range ampholytes and narrow range ampholytes were mixed in a ratio of 1:3 (c). The mixing ratio was chosen based on the work of Kahle et al. [15]. An additional approach was to only use narrow range ampholytes (d). Because the peak profile of the hACE2/His protein was not as reproducible, this protein was also measured with an addition of urea. For e, the two protein solubilizers SimpleSol and urea were used. Since the addition of SimpleSol and urea did not lead to a good peak shape either, SimpleSol was subsequently omitted and the concentration of urea was increased. In addition, several ampholyte mixtures were tested. Finally, a concentration of 7 M urea and a mixture of ampholytes 2.5–5 (3%) and 5–8 (2%) was used and thus well reproducible peaks were obtained (f). Since the proteins are very expensive and only available in small quantities, further optimization attempts were not conducted. Furthermore, there were no repetitions of non‐evaluable measurements, as the proteins will be used for future affinity capillary electrophoresis (ACE) experiments, in which the binding behavior of the SARS‐CoV‐2 proteins to the hACE2 receptor will be investigated. Figure 1 shows the electropherograms of each measured protein in the particular experiment where the protein yielded the most reproducible peaks with the best peak shape.
Figure 1

Electropherograms of the Maurice CIEF measurements. A: SARS‐CoV‐2 RBD/Fc experiment b; B: RBD/His experiment b; C: S1/His experiment b; D: S1/S2/His experiment a; E: hACE2/His experiment f. For better visualization, the y‐axes were displayed only in the area of fluorescence of the sample peaks.

Electropherograms of the Maurice CIEF measurements. A: SARS‐CoV‐2 RBD/Fc experiment b; B: RBD/His experiment b; C: S1/His experiment b; D: S1/S2/His experiment a; E: hACE2/His experiment f. For better visualization, the y‐axes were displayed only in the area of fluorescence of the sample peaks. The electropherograms show, that the proteins produce very different peak profiles. In case of the SARS‐CoV‐2 RBD/Fc (Figure 1A), the individual isoforms can be seen very clearly. SARS‐CoV‐2 RBD/His (Figure 1B), on the other hand, shows very small peaks in the approximate range of pI 7 to 8.5 and then some larger ones in the range of pI 8.5 to 10. The S1/His subunit (Figure 1C) produces three peaks, whose areas increase with pI, so the one with the highest pI is the largest at about 8.4. With the S1/S2/His subunit (Figure 1D), it was challenging to get peaks at all. Using a very high concentration, a broad peak with some spikes is finally obtained. After some optimization of the experimental conditions, the hACE2/His receptor (Figure 1E) shows a very reproducible peak profile, in which the different isoforms can be well recognized. With the peak profiles of the individual proteins in mind, we now take a look at the determined isoelectric points of the proteins. Table 2 summarizes the results of the CIEF measurements of the five proteins.
Table 2

Results of the Maurice CIEF measurements of SARS‐CoV‐2 proteins and the hACE2 receptor (experiments a‐f). Given are the mean values of the isoelectric points determined from a certain number of repetitions (n) and the relative standard deviation (RSD) in %

a RBD/Fc (theoretical pI: 8.55) n = 5
mean 8.3598.5728.7828.9249.0529.1539.2509.316
RSD [%] 0.2190.2220.0640.0930.1420.0880.0850.061
b n = 8
mean 8.2548.4788.6498.7948.9159.0219.1039.164
RSD [%] 0.1570.1580.0290.0320.0480.0440.0510.049
c n = 6
mean 8.3908.6338.7578.886
RSD [%] 0.0540.1340.0360.061
d n = 6
mean 8.2358.4298.6298.877
RSD [%] 0.0500.1590.0070.069
Results of the Maurice CIEF measurements of SARS‐CoV‐2 proteins and the hACE2 receptor (experiments a‐f). Given are the mean values of the isoelectric points determined from a certain number of repetitions (n) and the relative standard deviation (RSD) in % With regard to the results, the RBD/Fc protein will be discussed first. The theoretically calculated pI value of this protein is 8.55. In all experiments (a‐d) pI values in the range of 8.24 to 9.32 were obtained. The measured values are therefore in the range around the theoretically calculated pI, the distribution is probably due to isoforms. Next, we come to the results of the SARS‐CoV‐2 RBD with His‐tag. For the two measurements in ampholyte 3–10 (a, b) we have very similar pI values in the range between about 7.36 and 9.85. The pI values in measurement c, i.e., the measurement in the ampholyte mixture, lie between 7.93 and 9.88, which is probably due to the fact that the pI gradient in the ampholyte mixture at the edges of the narrow range ampholyte is not linear. The measurement with the narrow range ampholyte (d) could not be evaluated for this protein and was not repeated due to the small amount of material available. The theoretical pI of this protein is 8.91. In the measurements, this is in the range of the large peaks of the protein. Overall, the experimentally determined pI values can be found around the theoretically determined value in the range ± 1 pH unit for this protein. Looking at the S1 subunit with His‐tag, the pI range of the experiments a and b again hardly differs, so the pI values seem to behave nearly linear over the whole pI area. The determined range in these two experiments is about 7.48 to 8.37. The differences in comparison to the measurement in the ampholyte mixture (c) are probably due to the fact that the pI value of the protein is located at the border of the narrow range ampholyte (8–10.5) and pI values in this border area are probably somewhat distorted. The pI values determined using the narrow range ampholyte in experiment d lie in the range from 7.68 to 8.13, i.e., the range is significantly smaller than in the previous measurements. The theoretically calculated pI of the S1 subunit with His‐tag is 7.80 and thus approximately in the middle of the experimentally determined ranges. When measuring the S1/S2 subunit with His‐tag, the first problem was that no peaks could be found. The concentration of the protein was then continuously increased until peaks could finally be identified. These peaks, shown in the electropherogram in Figure 1D, could also be described as very broad and flat. Because of the high sample consumption due to the high concentration and the failed experiments before, only experiment a was performed. The experimentally determined pI values lie in a range from 4.41 to 5.87. Due to the unknown AA sequence, there is no theoretically calculated pI value and therefore it is not possible to assess whether the values are in the same range. The last protein investigated is the His‐tagged hACE2 receptor, whose theoretically calculated pI value is 5.60. The pI range determined with the wide range ampholyte and the markers 3.38 and 9.99 is 5.61 to 6.11 (experiment a), but the peaks looked as if the protein was aggregated. Subsequently, a lower concentration was chosen for measurements b, c, and d to avoid aggregation, whereby only experiment b could be evaluated. The pI range determined there is 5.36 to 5.69 and is in the range determined in experiment a, but smaller. Since the peaks still looked as if the protein was aggregated, a measurement with urea and SimpleSol as protein solubilizers was conducted afterwards (e). The reproducibility of the peak profiles was slightly improved and the values (5.37–5.75) hardly differ from those of measurement b. In order to obtain a reproducible peak shape and no aggregation, an attempt was made to optimize the method. In the following, SimpleSol was omitted and the concentration of urea was increased. In addition, several ampholyte mixtures were tried. In experiment f, with 7 M urea and ampholytes 2.5–5 (3%) and 5–8 (2%), peaks with recognizable isoforms of the receptor were finally obtained, which were very reproducible. The pI values in this experiment are between 5.19 and 5.93 and therefore in the range of the previously performed experiments. Thus, the aggregation did not have a strong effect on the pI value determination. Due to the good reproducibility, this method could be well used for quality control of the receptor, since even the smallest impurities would presumably have an effect on the peak shape. In summary, the theoretical pI values calculated on the basis of the AA sequence are in the same range as those determined experimentally. It is not possible to determine an exact value experimentally for these proteins, since all electropherograms show several peaks due to isoforms. However, these ranges always coincide with the calculated values. Therefore, the model calculations for these proteins are useful to make predictions about the isoelectric points of the proteins, if they are needed and no time and/or equipment is available for own experiments regarding the pI value determination. The experimental conditions (ampholyte, pI markers, concentration) do not have a particularly large influence on the determined pI value and merely the number and shape of peaks vary. By optimizing the experimental conditions, aggregation can be prevented and the separation of the individual isoforms can be made possible. With further optimization, quality control methods using CIEF can be developed for these proteins. The authors have declared no conflict of interest.
  10 in total

1.  Protease-mediated enhancement of severe acute respiratory syndrome coronavirus infection.

Authors:  Shutoku Matsuyama; Makoto Ujike; Shigeru Morikawa; Masato Tashiro; Fumihiro Taguchi
Journal:  Proc Natl Acad Sci U S A       Date:  2005-08-22       Impact factor: 11.205

2.  Design of experiments as a valuable tool for biopharmaceutical analysis with (imaged) capillary isoelectric focusing.

Authors:  Julia Kahle; Matthias Stein; Hermann Wätzig
Journal:  Electrophoresis       Date:  2019-06-21       Impact factor: 3.535

Review 3.  Isoelectric focusing in gels.

Authors:  P G Righetti; J W Drysdale
Journal:  J Chromatogr       Date:  1974-09-25

4.  Theoretical and experimental study of high-performance electrophoretic mobilization of isoelectrically focused protein zones.

Authors:  S Hjertén; J L Liao; K Q Yao
Journal:  J Chromatogr       Date:  1987-01-30

Review 5.  Trajectory of isoelectric focusing from gels to capillaries to immobilized gradients in capillaries.

Authors:  Brooke M Koshel; Mary J Wirth
Journal:  Proteomics       Date:  2012-09-24       Impact factor: 3.984

6.  Preparation of a capillary isoelectric focusing column with monolithic immobilized pH gradient and its application on protein separation based on an online capillary isoelectric focusing platform.

Authors:  Rangdong Liu; Hong Pan; Yiting Zheng; Jing Hu; Soumia Cheddah; Weiwei Wang; Yan Wang; Chao Yan
Journal:  Electrophoresis       Date:  2019-04-29       Impact factor: 3.535

7.  On-line combination of monolithic immobilized pH gradient-based capillary isoelectric focusing and capillary zone electrophoresis via a partially etched porous interface for protein analysis.

Authors:  Tingting Wang; Junfeng Ma; Shuaibin Wu; Liangliang Sun; Huiming Yuan; Lihua Zhang; Zhen Liang; Yukui Zhang
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2011-02-21       Impact factor: 3.205

8.  Repeatedly usable immobilized pH gradient in a monolithic capillary column.

Authors:  Chun Yang; Guijie Zhu; Lihua Zhang; Weibing Zhang; Yukui Zhang
Journal:  Electrophoresis       Date:  2004-06       Impact factor: 3.535

9.  Structure, Function, and Antigenicity of the SARS-CoV-2 Spike Glycoprotein.

Authors:  Alexandra C Walls; Young-Jun Park; M Alejandra Tortorici; Abigail Wall; Andrew T McGuire; David Veesler
Journal:  Cell       Date:  2020-03-09       Impact factor: 41.582

Review 10.  Physicochemical properties of SARS-CoV-2 for drug targeting, virus inactivation and attenuation, vaccine formulation and quality control.

Authors:  Christin Scheller; Finja Krebs; Robert Minkner; Isabel Astner; Maria Gil-Moles; Hermann Wätzig
Journal:  Electrophoresis       Date:  2020-06-08       Impact factor: 3.595

  10 in total
  6 in total

1.  Biofunctionalization of Graphene-Based FET Sensors through Heterobifunctional Nanoscaffolds: Technology Validation toward Rapid COVID-19 Diagnostics and Monitoring.

Authors:  Esteban Piccinini; Gonzalo E Fenoy; Agustín L Cantillo; Juan A Allegretto; Juliana Scotto; José M Piccinini; Waldemar A Marmisollé; Omar Azzaroni
Journal:  Adv Mater Interfaces       Date:  2022-03-18       Impact factor: 6.389

2.  Amomum tsao-ko essential oil, a novel anti-COVID-19 Omicron spike protein natural products: A computational study.

Authors:  Ju-Zhao Liu; Hong-Chang Lyu; Yu-Jie Fu; Qi Cui
Journal:  Arab J Chem       Date:  2022-04-18       Impact factor: 6.212

3.  Adsorption of SARS CoV-2 spike proteins on various functionalized surfaces correlates with the high transmissibility of Delta and Omicron variants.

Authors:  Daniela Dobrynin; Iryna Polishchuk; Lotan Portal; Ivan Zlotver; Alejandro Sosnik; Boaz Pokroy
Journal:  Mater Today Bio       Date:  2022-04-19

4.  Carbon nanotube field-effect transistor (CNT-FET)-based biosensor for rapid detection of SARS-CoV-2 (COVID-19) surface spike protein S1.

Authors:  Mazin A Zamzami; Gulam Rabbani; Abrar Ahmad; Ahmad A Basalah; Wesam H Al-Sabban; Saeyoung Nate Ahn; Hani Choudhry
Journal:  Bioelectrochemistry       Date:  2021-10-15       Impact factor: 5.373

5.  Sensitive detection of SARS-CoV-2 spike protein using vertically-oriented silicon nanowire array-based biosensor.

Authors:  Bingtao Gao; Anthony A Rojas Chavez; Walla I Malkawi; Daniel W Keefe; Rasheid Smith; Hillel Haim; Aliasger K Salem; Fatima Toor
Journal:  Sens Biosensing Res       Date:  2022-03-22

6.  Design and simulation of a millifluidic device for differential detection of SARS-CoV-2 and H1N1 based on triboelectricity.

Authors:  Tara Ghafouri; Negin Manavizadeh
Journal:  Bioelectrochemistry       Date:  2022-03-16       Impact factor: 5.760

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.