| Literature DB >> 33458424 |
Erik Pettersson1,2,3, Anna Bäck2,3, Thomas Björk-Eriksson4,5, Ulrika Lindencrona2,3, Karin Petruson6, Anne Thilander-Klang1,3.
Abstract
BACKGROUND ANDEntities:
Keywords: Computed tomography; Delineation uncertainty; Dual-energy computed tomography; Metal artefact reduction; Radiotherapy; Treatment planning
Year: 2019 PMID: 33458424 PMCID: PMC7807554 DOI: 10.1016/j.phro.2019.01.001
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Left: A photograph of the phantom with the central cylinder containing five structures denoted 1 to 5 in homogeneous gelatine. Tissue surrogate inserts with six different mass densities, lung exhale (0.50 g/cm3), adipose (0.96 g/cm3), breast (0.99 g/cm3), plastic water (1.03 g/cm3), liver (1.07 g/cm3) and trabecular bone (1.16 g/cm3) were inserted in the six frontal and dorsal positions of the phantom. Ten cm of plastic water slabs were placed in both the cranial and caudal direction of the central electron density phantom during the CT scans. The caudal slabs are not present in the photograph. Right: A SECT image of the phantom acquired with PMMA inserts in the lateral positions (A10) with display window level and width set to 40 HU and 400 HU, respectively. The two PMMA cylinders are marked with asterisks (*).
The scanning and reconstruction parameters used for the SECT and DECT scanners.
| Scanner | Tube voltage (kV) | Scan mode (Bow-tie filter) | Helical pitch | Tube current [mA] | Rotation time [s] | SFOV [cm] | DFOV [cm] | Beam collima-tion [mm] | Slice thickness [mm] | Convolu-tion kernel | CTDIvol,32cm [mGy] |
|---|---|---|---|---|---|---|---|---|---|---|---|
| SECT | 120 | LL | 0.938 | 125 | 0.5 | 55 | 55 | 32 | 3.0 | FC17 | 14.5 |
| DECT | 80/140 | Body | 0.969 | 225 | 0.6 | 50 | 50 | 40 | 2.5 | Standard | 13.84 |
| Ref. | 120 | Body | 0.969 | 525 | 1 | 50 | 25 | 40 | 2.5 | Standard | 47 |
SFOV = scan field of view, DFOV = display field of view, CTDIvol, 32cm = volumetric CT dose index in the 32 cm phantom.
The high-dose SECT scan acquired with the DECT scanner used to reconstruct the low-noise reference images.
The sequence of image series in which the four radiation oncologists delineated the structures in the phantom. The image series were presented in descending degree of metal artefacts. This sequence were used for both phantom set-up A and B.
| Image series | Scanner | Properties | Metal artefact reduction | Lateral insert material |
|---|---|---|---|---|
| 1 | SECT | 120 kV | – | Steel |
| 2 | DECT | 70 keV | No | Steel |
| 3 | DECT | 110 keV | No | Steel |
| 4 | DECT | 110 keV | No | Titanium |
| 5 | SECT | 120 kV | – | Titanium |
| 6 | DECT | 110 keV | Yes | Steel |
| 7 | DECT | 70 keV | Yes | Steel |
| 8 | DECT | 110 keV | Yes | Titanium |
| 9 | DECT | 70 keV | Yes | Titanium |
| 10 | SECT | 120 kV | – | PMMA |
| 11 | DECT | 110 keV | No | PMMA |
| 12 | DECT | 70 keV | No | PMMA |
| 13 | Ref. | 120 kV, MBIR | – | PMMA |
kV indicates polychromatic X-ray spectra, while keV indicates virtually monoenergetic energy resulting from dual-energy CT scanning.
The Ref. image series was a high dose SECT scan acquired with the DECT scanner.
MBIR = model-based iterative reconstruction.
Fig. 2The central slice in each of the image series of phantom set-up A (A1-A13). The image series were ordered in a sequence of decreasing severity of metal artefacts. Image series number 13 does not depict the whole phantom as it was reconstructed with a reduced display field of view of 25 cm. The same sequence was used for both phantom set-ups.
Fig. 3The contours delineated by the four radiation oncologists (green) and the reference contours (red) in the central image slice of each of the 13 image series in phantom set-up A. For each phantom set-up, image series 13 was also used to create the red reference contours. Only the central part of the phantom is displayed in the figure, but the oncologists were shown the whole images. For the corresponding figure with set-up B, see Fig. B.1 in the supplementary material. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 4Delineation of all structures (1–5) by all four radiation oncologists, for the 13 image series with the set-up A (a) and B (b). The numbering of the structures is given in the image in the top left corners. The colours in the cells indicate how many of the four radiation oncologists that delineated each structure in each image series (white = 0, red = 1, orange = 2, yellow = 3, green = 4). The minimum and maximum values of the corresponding Dice similarity coefficients and Hausdorff distances for each structure and image series, are given with the colour scheme described above for each phantom set-up. Reference denotes the SECT scans performed with the DECT scanner, these image series (A13 and B13) were used to create a reference structure sets. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)