How Well Does Dual-energy CT with Fast Kilovoltage Switching Quantify CT Number and Iodine and Calcium Concentrations?

Shingo Ohira1, Tsukasa Karino2, Yoshihiro Ueda3, Yuya Nitta2, Naoyuki Kanayama2, Masayoshi Miyazaki2, Masahiko Koizumi4, Teruki Teshima5.   

Abstract

RATIONALE AND
OBJECTIVES: Because it is imperative for understanding the performance of dual-energy computed tomography scanner to determine clinical diagnosis, we aimed to assess the accuracy of quantitative measurements using dual-energy computed tomography with fast kilovoltage switching.
MATERIALS AND METHODS: Quantitative measurements were performed for 16 reference materials (physical density, 0.965-1.550 g/cm3; diameter of rod, 2.0-28.5 mm; iodine concentration, 2-15 mg/mL; and calcium concentration, 50-300 mg/mL) with varying scanning settings, and the measured values were compared to their theoretical values.
RESULTS: For high-density material, the maximum differences in Hounsfield unit values in the virtual monochromatic images at 50, 70, and 100 keV were -176.2, 61.0, and -35.2 HU, respectively, and the standard deviations over short- and long-term periods were 11.1, 6.1, and 3.5 HU at maximum. The accuracy of the Hounsfield unit measurement at 50 and 70 keV was significantly higher (P < 0.05) with higher radiation output and smaller phantom size. The difference in the iodine and calcium measurements in the large phantom were up to -2.6 and -60.4 mg/mL for iodine (5 mg/mL with 2-mm diameter) and calcium (300 mg/mL) materials, and the difference was improved with a small phantom. Metal artifact reduction software improved subjective image quality; however, the quantitative values were significantly underestimated (P < 0.05) (-49.5, -26.9, and -15.3 HU for 50, 70, and 100 keV, respectively; -1.0 and -17 mg/mL for iodine and calcium concentration, respectively) compared to that acquired without a metal material.
CONCLUSIONS: The accuracy of quantitative measurements can be affected by material density and the size of the object, radiation output, phantom size, and the presence of metal materials.
Copyright © 2018 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Computed tomography; HU; calcium; dual energy; iodine

Mesh:

Substances:

Year:  2017        PMID: 29275940     DOI: 10.1016/j.acra.2017.11.002

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  2 in total

1.  Evaluation of Stopping Power Ratio Calculation Using Dual-energy Computed Tomography With Fast Kilovoltage Switching for Treatment Planning of Particle Therapy.

Authors:  Shingo Ohira; Yasuhiro Imai; Yuhei Koike; Shunsuke Ono; Yoshihiro Ueda; Masayoshi Miyazaki; Masahiko Koizumi; Koji Konishi
Journal:  In Vivo       Date:  2022 Jan-Feb       Impact factor: 2.155

2.  Structure delineation in the presence of metal - A comparative phantom study using single and dual-energy computed tomography with and without metal artefact reduction.

Authors:  Erik Pettersson; Anna Bäck; Thomas Björk-Eriksson; Ulrika Lindencrona; Karin Petruson; Anne Thilander-Klang
Journal:  Phys Imaging Radiat Oncol       Date:  2019-02-13
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.