Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data.

J Greffier1, S Si-Mohamed2,3, D Dabli4, H de Forges4, A Hamard4, P Douek2,3, J P Beregi4, J Frandon4.   

Abstract

OBJECTIVES: To compare the spectral performance of dual-energy CT (DECT) platforms using task-based image quality assessment based on phantom data.
MATERIALS AND METHODS: Two CT phantoms were scanned on four DECT platforms: fast kV-switching CT (KVSCT), split filter CT (SFCT), dual-source CT (DSCT), and dual-layer CT (DLCT). Acquisitions on each phantom were performed using classical parameters of abdomen-pelvic examination and a CTDIvol at 10 mGy. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated from 40 to 140 keV of virtual monoenergetic images. A detectability index (d') was computed to model the detection task of two contrast-enhanced lesions as function of keV.
RESULTS: The noise magnitude decreased from 40 to 70 keV for all DECT platforms, and the highest noise magnitude values were found for KVSCT and SFCT and the lowest for DSCT and DLCT. The average NPS spatial frequency shifted towards lower frequencies as the energy level increased for all DECT platforms, smoothing the image texture. TTF values decreased with the increase of keV deteriorating the spatial resolution. For both simulated lesions, higher detectability (d' value) was obtained at 40 keV for DLCT, DSCT, and SFCT but at 70 keV for KVSCT. The detectability of both simulated lesions was highest for DLCT and DSCT.
CONCLUSION: Highest detectability was found for DLCT for the lowest energy levels. The task-based image quality assessment used for the first time for DECT acquisitions showed the benefit of using low keV for the detection of contrast-enhanced lesions. KEY POINTS: • Detectability of both simulated contrast-enhanced lesions was higher for dual-layer CT for the lowest energy levels. • The image noise increased and the image texture changed for the lowest energy levels. • The detectability of both simulated contrast-enhanced lesions was highest at 40 keV for all dual-energy CT platforms except for fast kV-switching platform.

Keywords:  Comparative study; Phantoms, imaging; Radiography, dual-energy scanned projection; Tomography, X-ray computed

Year:  2021        PMID: 33449188     DOI: 10.1007/s00330-020-07671-2

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  1 in total

Review 1.  Dual-energy CT revisited with multidetector CT: review of principles and clinical applications.

Authors:  Muşturay Karçaaltıncaba; Aykut Aktaş
Journal:  Diagn Interv Radiol       Date:  2010-11-14       Impact factor: 2.630

  1 in total
  2 in total

1.  Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance.

Authors:  Davide Cester; Matthias Eberhard; Hatem Alkadhi; André Euler
Journal:  Quant Imaging Med Surg       Date:  2022-01

2.  Comparison of virtual monoenergetic imaging between a rapid kilovoltage switching dual-energy computed tomography with deep-learning and four dual-energy CTs with iterative reconstruction.

Authors:  Joël Greffier; Salim Si-Mohamed; Boris Guiu; Julien Frandon; Maeliss Loisy; Fabien de Oliveira; Philippe Douek; Jean-Paul Beregi; Djamel Dabli
Journal:  Quant Imaging Med Surg       Date:  2022-02
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.