Delayed management of atrial lead dislodgment after pacemaker implantation: a case report.

Fu Guan1, Guangping Li2, Yong Liu3, Xing Gao4, Rui Zhou4.   

Abstract

BACKGROUND: Pacemaker lead dislodgement may cause malfunction in the pacing system, which may lead to severe adverse events. For patients with sick sinus syndrome but normal atrioventricular conduction, atrial lead dislocation may cause excessive unnecessary ventricular pacing, resulting in nonphysiological pacing leading to heart failure. The longer the unwanted ventricular pacing continues, the greater the chances that irreversible heart failure may occur. Ironically, we admitted a patient who had been refusing dislodged lead relocation for 7 years. The symptoms of heart failure were significantly resolved after new atrial lead implantation. We reviewed her clinical data before and after the procedure and believed the case was worthy of reflection. CASE
PRESENTATION: An 83-year-old Han Chinese woman presented with heart failure symptoms for 7 years due to the late macro-dislodgement of an atrial pacing lead. Her echocardiogram showed average left ventricular ejection fraction (LVEF) but reduced left ventricular end-diastolic volume (LVEDV) during right ventricular pacing, indicating heart failure with preserved ejection fraction (HFpEF). After 7 years of refusal, she finally agreed to implantation of a new atrial lead. She has been doing well since the operation.
CONCLUSIONS: For patients with sick sinus syndrome with dual-chamber pacemaker indication, atrial lead dislodgement should be appropriately managed if the atrioventricular function is normal. As the consequences are subtle and appear gradually, they might be overlooked by patients and even doctors. Implanting a new atrial lead is the right thing to do rather than just passively waiting or treating with symptom relief medications.

Entities:  

Keywords:  Case report; Complication; Lead dislodgement; Pacemaker

Mesh:

Year:  2021        PMID: 33441160      PMCID: PMC7807897          DOI: 10.1186/s13256-020-02626-z

Source DB:  PubMed          Journal:  J Med Case Rep        ISSN: 1752-1947


  10 in total

1.  "Reel Syndrome": a new form of Twiddler's syndrome?

Authors:  A Carnero-Varo; M Pérez-Paredes; J A Ruiz-Ros; D Giménez-Cervantes; F R Martínez-Corbalán; T Cubero-López; P Jara-Pérez
Journal:  Circulation       Date:  1999-08-24       Impact factor: 29.690

2.  Late displacement of a ventricular pacing lead after respiratory therapy.

Authors:  R Arroyo Espliguero; J F Oteo Domínguez; E Castedo Mejuto; I Antorrena Miranda; J Ortigosa Aso; A Cañas Cañas; M Artaza Andrade
Journal:  Pacing Clin Electrophysiol       Date:  2001-11       Impact factor: 1.976

3.  An unusual case of pacemaker failure: complete disconnection of connector block and battery of a subpectorally implanted dual chamber pacemaker.

Authors:  Fritz Mellert; Bahman Esmailzadeh; Christian Schneider; Marcus Haushofer; Rainer Schimpf; Christian Wolpert; Claus J Preusse; Berndt Lüderitz; Armin Welz
Journal:  Pacing Clin Electrophysiol       Date:  2002-04       Impact factor: 1.976

4.  Twiddler's Syndrome.

Authors:  William J Nicholson; Kathryn A Tuohy; Peter Tilkemeier
Journal:  N Engl J Med       Date:  2003-04-24       Impact factor: 91.245

5.  Minimizing right ventricular pacing in pacemaker patients with intact and compromised atrioventricular conduction : Results from the EVITA Trial.

Authors:  A Bauer; J Vermeulen; L Toivonen; J Voitk; C Barr; P Peytchev
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-08-28

6.  Role of transesophageal echocardiography in the evaluation of patients with clinical pacemaker syndrome.

Authors:  T M Lee; S F Su; Y J Lin; W J Chen; M F Chen; C S Liau; Y T Lee
Journal:  Am Heart J       Date:  1998-04       Impact factor: 4.749

7.  Early complications after dual chamber versus single chamber pacemaker implantation.

Authors:  A Chauhan; A A Grace; S A Newell; D L Stone; L M Shapiro; P M Schofield; M C Petch
Journal:  Pacing Clin Electrophysiol       Date:  1994-11       Impact factor: 1.976

Review 8.  Adverse effects of long-term right ventricular apical pacing and identification of patients at risk of atrial fibrillation and heart failure.

Authors:  Antonio De Sisti; Manlio F Márquez; Joelci Tonet; Aimé Bonny; Robert Frank; Françoise Hidden-Lucet
Journal:  Pacing Clin Electrophysiol       Date:  2012-03-27       Impact factor: 1.976

9.  Use of an antimicrobial pouch/envelope in the treatment of Twiddler's syndrome.

Authors:  Moses Osoro; William Lorson; Jeffrey B Hirsh; William Jeremy Mahlow
Journal:  Pacing Clin Electrophysiol       Date:  2018-01-16       Impact factor: 1.976

10.  Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction.

Authors:  Michael O Sweeney; Anne S Hellkamp; Kenneth A Ellenbogen; Arnold J Greenspon; Roger A Freedman; Kerry L Lee; Gervasio A Lamas
Journal:  Circulation       Date:  2003-06-02       Impact factor: 29.690

  10 in total
  1 in total

1.  EnSite NavX mapping system guided implantation of a dual-chamber permanent pacemaker in a 41-year-old pregnant woman with a 4-year follow-up.

Authors:  Peng Wang; Guang-Sheng Wei; Jun-Hua Wang; Yan-Jie Cao; Wei-Wei Zhu; Hang Shen; Zhi-Yue Zhang; Li Ai; Meng Wang
Journal:  BMC Cardiovasc Disord       Date:  2022-07-21       Impact factor: 2.174

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.