On the choice of phantom material for the dosimetry of 192Ir sources.

J A Meli1, A S Meigooni, R Nath.   

Abstract

Dosimetric characteristics of polystyrene, solid water, and polymethylmethacrylate were examined and compared to water to determine the suitability of these solid materials for the dosimetry of 192Ir. Ionization charge measured in each of the four media as a function of depth and depth-dose curves calculated by Monte Carlo simulation show that the three solids are equivalent to each other and to water under full scattering conditions. Photon energy spectra generated from the Monte Carlo simulation show little variation for the different media. Mass energy absorption coefficients and exposure-to-dose conversion factors were calculated as a function of depth for these spectra. Measured tissue attenuation factors are in excellent agreement with Meisberger's "selected" values. The radial dose function, which describes the change in dose with distance in phantom exclusive of the inverse square law, was calculated from the tissue attenuation factor and found to be in significant disagreement with Dale's Monte Carlo values. The reason for this discrepancy is discussed.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3343169     DOI: 10.1016/0360-3016(88)90280-5

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.

Authors:  Somayeh Gholami; Hamid Reza Mirzaei; Ali Jabbary Arfaee; Ramin Jaberi; Hassan Ali Nedaie; Seied Rabi Mahdavi; Eftekhar Rajab Bolookat; Ali S Meigooni
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-18

2.  Monte Carlo modeling of Co HDR brachytherapy source in water and in different solid water phantom materials.

Authors:  S Sahoo; T Palani Selvam; R S Vishwakarma; G Chourasiya
Journal:  J Med Phys       Date:  2010-01

3.  Comparative dosimetry of GammaMed Plus high-dose rate Ir brachytherapy source.

Authors:  N P Patel; B Majumdar; V Vijayan
Journal:  J Med Phys       Date:  2010-07

4.  Determination of the tissue inhomogeneity correction in high dose rate Brachytherapy for Iridium-192 source.

Authors:  Barlanka Ravikumar; S Lakshminarayana
Journal:  J Med Phys       Date:  2012-01

5.  Monte Carlo-based investigation of water-equivalence of solid phantoms at (137)Cs energy.

Authors:  Ramkrushna S Vishwakarma; T Palani Selvam; Sridhar Sahoo; Subhalaxmi Mishra; Ghanshyam Chourasiya
Journal:  J Med Phys       Date:  2013-10

6.  Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix.

Authors:  Satish C Uniyal; Umesh C Naithani; Sunil D Sharma; Anoop K Srivastava
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.