A two minute liquid based sample preparation for rapid SARS-CoV2 real-time PCR screening: A multicentre evaluation.

Eric C J Claas1, Pieter W Smit2, Mario J A W M van Bussel3, Harold Verbakel4, Mohammed Taouil2, Jaco J Verweij4, Steven F T Thijsen5.   

Abstract

BACKGROUND: Apart from major health concerns associated to the SARS-coronavirus-2 (SARS-CoV-2) pandemic, also the diagnostic workflow encountered serious problems. Limited availability of kit components, buffers and even plastics has resulted in suboptimal testing procedures worldwide. Alternative workflows have been implemented to overcome these difficulties. Recently a liquid based sample prep has been launched as solution to overcome limitations in relation to nucleic acid extraction.
OBJECTIVE: Multicenter evaluation of the QIAprep&amp Viral RNA UM kit (QIA P&A) for rapid sample preparation and real-time PCR detection of SARS-CoV-2 in comparison to standardized laboratory testing methods. STUDY
DESIGN: Selected samples of the routine diagnostic workflow at Clinical Microbiology Laboratories of four Dutch hospitals have been subjected to the rapid QIA P&A protocol and the results have been compared to routine diagnostic data.
RESULTS: Combining results of manual and automated procedures, a total of 377 clinical samples of which 202 had been tested positive with a wide range of CT values, showed almost complete concordance in the QIA P&A assay for samples up to CT values of 33 with one exception of CT 31. Prospectively 60 samples were tested and also showed 100 % concordance with 5 positives. The method has been automated by two centres.
CONCLUSIONS: Despite an input of only 8 μL of clinical sample, the QIA P&A kit showed good performance for sample preparation and amplification of SARS-CoV-2 and can contribute as a rapid molecular testing strategy in managing the CoV-2 pandemic.
Copyright © 2021 The Authors. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Real-time PCR; SARS-CoV-2; Sample preparation

Mesh:

Year:  2020        PMID: 33418142      PMCID: PMC7774005          DOI: 10.1016/j.jcv.2020.104720

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Background

The SARS-CoV-2 pandemic has been and still is a challenge for the world. Although the initial identification [1] and subsequent implementation of diagnostic methods [2] was extremely fast, the pandemic could not be contained, in contrast to the 2002–2003 SARS-CoV-1 outbreak [3]. One of the reasons is the efficient spread of SARS-CoV-2, partly because also asymptomatically infected individuals or patients prior clinical symptoms had been found to be contagious [4] The worldwide spread of the virus also resulted in major complications for diagnostic laboratories, partly by insufficient access to reagents and even plastics provided by diagnostic companies. Initially lysis buffers, required for inactivation of the virus and crucial for nucleic acid (NA) extraction from clinical samples, were the main limitation, but later also availability of plastics and even pipet tips became problematic because of the worldwide demand [5]. As a result, most laboratories have spread the risk by investing in implementation of alternative platforms for diagnosing SARS-CoV-2. Recently, a new sample preparation solution was launched by Qiagen, the Qiaprep& Viral RNA UM (QIA P&A) kit, that combined a short, liquid based sample preparation with one-step real-time PCR amplification and detection of SARS-CoV-2, including two internal controls; one synthetic RNA IC and a sampling control amplifying transcripts of human B2M and RNaseP genes. The Research Use Only version of the kit, has been used in combination with the primers and probes developed by the CDC [6] which are included in the SARS-CoV-2 N1 + N2 assay (Qiagen) as well as the oligos published by Corman et al. [2].

Objectives

In this study, the performance of the QIA P&A kit was investigated by comparing the result to those of routine diagnostic workflows, that include NA extraction procedures, in four Dutch hospital laboratories.

Study design

Clinical samples submitted for clinical diagnosis of SARS-CoV-2 infection to the hospital laboratories, were used for comparison between the primary diagnostic results and the QIA P&A kit. All samples were anonymized, so no approval by the Medical Ethics Committee was required as only diagnostic test results have been compared. For the QIA P&A test, 8 μL of clinical sample in transport medium was added to 2 μL of sample preparation buffer in a PCR plate. After 2 min of incubation, 10 μL of reaction mixture was added and after mixing, the RNA PCR amplification protocol was started. Two centres managed to automate the workflow.

Results (Table 1)

The Leiden University Medical Centre (LUMC) retrospectively selected 89 clinical samples of which 36 positives as determined by the Alinity-m SARS-COV-2 assay (Abbott, USA). All samples that were positive with a CT up to 35 could be confirmed. Of five positives with CT >35, three samples tested negative in the QIA P&A. A prospective run of 60 samples simultaneously to the Alinity-m showed complete concordance, including 5 positive samples. Performance of QIA P&A as compared to the routine diagnostic procedures. CT IC = 0. The Diakonessenhuis (DIAK) selected 104 samples that had been tested in real-time PCR [2]. Of the 84 positive samples 75 were positive in the QIA P&A. Of 11 samples with CT values over 35, only 2 were detected by the QIA P&A (18 %). Only two of the nine discrepant samples could be confirmed as positive in the GeneXpert® (Cepheid, USA) assay. A subset of 11 positive samples was tested in two separate runs for analysis of the reproducibility of the QIA P&A results (data not shown). The difference in CT values ranged from 0 (perfect duplicate) up to 2.8, without an association to viral load: one duplicate had CT values of 23.7 and 26.5 whereas another sample (initial CT of 32.5) resulted in CT 35.3 and 34.9 (0.4 difference) by QIA P&A. As automation of pipetting volumes of 2 μL and 8 μL may be challenging, the Elisabeth Tweesteden (ETZ) hospital managed to automate the procedure on the QIAsymphony SP/AS (Qiagen) platform, using increased volumes of 20 μL sample preparation buffer and 80 μL clinical sample. In this way they tested 64 samples, compared to diagnostic real-time PCR targeting the E gene [2] on the QIAsymphony SP/AS. The results showed that only 3 samples with CT values of 33.3, 33.7 and 35.5 were not detected. However, one sample previously tested negative was positive in the QIA P&A with a CT of 37.9. The Alinity m confirmed this sample to be true positive (CT 34.7). Maasstad hospital managed to automate the QIA P&A procedure using small pipetting volumes of 2 μL and 8 μL, using a Tecan Fluent 480 pipetting platform (Tecan, Switzerland). A total of 94 samples were tested using real-time PCR with sarbeco E primers [2] in the QIA P&A and the NeuMoDx™ 96 (Qiagen) as reference assay. The NeuMoDx™ is a fully automated microfluidics based PCR platform [7]. Out of 22 positive samples, 21 (95.5 %) were positive in the QIA P&A. One sample with an initial CT value of 31.0 was missed. This sample was confirmed in an LDT with a CT of 37.1. Out of 72 negative samples, 63 were confirmed and nine samples were inhibited based on a failed internal control, also after retesting. Three centres tested an EQA panel provided by the Dutch National Institute of Health and Environment (RIVM, Bilthoven). Table 2 shows the results of this panel for both their routine diagnostic procedure and the QIA P&A. Although some variation in the high CT value samples, very comparable results have been obtained. The observed specificity is depending on the primer set used. All assays gave negative results with other circulating human coronaviruses, but the E-gene primers are beta-coronavirus group specific and therefore also detected SARS-CoV-1 (sample 10).
Table 2

Comparative results of the 4th Dutch EQA panel for SARS-CoV-2 testing.

EQA panelHCoVLUMCMAASSTADTecanETZQS-AS
DX#QIA P&ADX#QIA P&ADX#QIA P&A
1SARS-CoV-227.128.926.933.729.125.5
2HCoV-NL63000000
3HCoV-229E000000
4SARS-CoV-239.537.500034.8
5SARS-CoV-234.233.431.1035.732.0
6HCoV-OC43000000
7SARS-CoV-236.735.231.140.834.732.3
8Negative000000
9Negative000000
10SARS-CoV-10027.7*38.5*29.0*0

DX = routine diagnostic test.

sarbeco PCR: reactive to betacoronaviruses, not specific to SARS-CoV-2.

Comparative results of the 4th Dutch EQA panel for SARS-CoV-2 testing. DX = routine diagnostic test. sarbeco PCR: reactive to betacoronaviruses, not specific to SARS-CoV-2.

Conclusions

Nucleic extraction is a crucial part of molecular diagnostic procedures and results in highly purified DNA or RNA, which may take from 30 min. up to several hours. This study shows the evaluation of a two-minute, liquid based, sample preparation method for SARS-CoV-2 testing in comparison to different diagnostic workflows at four hospital laboratories in the Netherlands. The overall conclusion is that the QIA P&A procedure provided excellent results, given the small amount of sample that is being used. Performing the QIA P&A on 185 positive samples with CT values up to 35 samples as determined with different platforms, resulted in 182 positive results. Only three samples with CT values of 31.0, 33.3 and 33.7 from the two centres using automated procedures were not detected in QIA P&A. Of 18 samples with CT >35, only 5 (28 %) were detected, but the clinical relevance of this kind of results is an ongoing point of discussion. It should be noted that the sample preparation buffer does not inactivate the SARS-CoV-2, so for proper sample handling a BSL-2 facility is required. This can be overcome by heat inactivation of the sample in transport buffers, for which 50 μL sample at 70 °C for 10 min. would be sufficient according to the manufacturer’s instructions. The QIA P&A can handle a broad range of non-fixating buffers and transport media, but is incompatible with inactivating lysis buffers containing detergents or guanidinium thiocyanate. Potentially, the inhibited samples observed in one centre, were performed on swab in lysis buffer. Unfortunately, the original sample tubes were unavailable for reinspection. The QIA P&A can be used in combination with lab developed primer sets for other viruses as well. Early 2021, the company will launch a CE/IVD version of the kit containing specific viral target primers and probes. Altogether the QIA P&A can be considered a valuable addition to the diagnostic laboratory, either as back -up when extraction reagents or plastics are poorly delivered or, when automated, to further increase the diagnostic capacity.

Declaration of Competing Interest

None of the authors reports a conflict of interest. The QIA P&A kit was provided by Qiagen, without further involvement in experimental design and data analysis.
Table 1

Performance of QIA P&A as compared to the routine diagnostic procedures.

Manual procedure
LUMCQIA P&A vs Alinity m
N=ConcordanceMissed CT
POS Cq <303333100 %
POS Cq 30−3533100 %
POS Cq >355240 %35.2, 35.9, 38.8
NEGATIVE7575100 %
DIAKQIA P&A vs LDT/GeneXpert®
N=ConcordanceMissed CT
POS Cq <306464100 %
POS Cq 30−3599100 %
POS Cq >3511218 %range 36.0−39.9
NEGATIVE2020100 %



Automated procedures
ETZQIAP&A vs LDT on QIASymphony
QS SP/ASN=ConcordanceMissed CT
POS Cq <303838100 %
POS Cq 30−35161487.5 %33.7, 34.1
POS Cq >352150 %35.5
NEGATIVE8787,5 %37.9



MAASSTADQIA P&A vs LDT on NeuMoDx96
TecanN=ConcordanceMissed CT
POS Cq <302020100 %
POS Cq 30−352150 %31.0
NEGATIVE726387.5 %9x Inhibition*

CT IC = 0.

  7 in total

1.  A novel coronavirus associated with severe acute respiratory syndrome.

Authors:  Thomas G Ksiazek; Dean Erdman; Cynthia S Goldsmith; Sherif R Zaki; Teresa Peret; Shannon Emery; Suxiang Tong; Carlo Urbani; James A Comer; Wilina Lim; Pierre E Rollin; Scott F Dowell; Ai-Ee Ling; Charles D Humphrey; Wun-Ju Shieh; Jeannette Guarner; Christopher D Paddock; Paul Rota; Barry Fields; Joseph DeRisi; Jyh-Yuan Yang; Nancy Cox; James M Hughes; James W LeDuc; William J Bellini; Larry J Anderson
Journal:  N Engl J Med       Date:  2003-04-10       Impact factor: 91.245

Review 2.  Pathophysiology, Transmission, Diagnosis, and Treatment of Coronavirus Disease 2019 (COVID-19): A Review.

Authors:  W Joost Wiersinga; Andrew Rhodes; Allen C Cheng; Sharon J Peacock; Hallie C Prescott
Journal:  JAMA       Date:  2020-08-25       Impact factor: 56.272

3.  Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.

Authors:  Victor M Corman; Olfert Landt; Marco Kaiser; Richard Molenkamp; Adam Meijer; Daniel Kw Chu; Tobias Bleicker; Sebastian Brünink; Julia Schneider; Marie Luisa Schmidt; Daphne Gjc Mulders; Bart L Haagmans; Bas van der Veer; Sharon van den Brink; Lisa Wijsman; Gabriel Goderski; Jean-Louis Romette; Joanna Ellis; Maria Zambon; Malik Peiris; Herman Goossens; Chantal Reusken; Marion Pg Koopmans; Christian Drosten
Journal:  Euro Surveill       Date:  2020-01

4.  Under-allocation: Critical Supply Chain Hurdles Negatively Impact the Ability of Community Hospitals To Perform Repeat SARS-CoV-2 Testing.

Authors:  Susan M Butler-Wu; Noah Wald-Dickler; Paul Holtom; Kenneth M Zangwill; Tam T Van
Journal:  J Clin Microbiol       Date:  2020-08-24       Impact factor: 5.948

5.  Development of a fully automated high throughput PCR for the detection of SARS-CoV-2: The need for speed.

Authors:  Florian J Mayer; Franz Ratzinger; Ralf L J Schmidt; Georg Greiner; Olfert Landt; Alexander Am Ende; Victor M Corman; Nicole Perkmann-Nagele; Thomas Watkins-Riedel; Dagmar Petermann; Karoline Abadir; Josef Zweimüller-Mayer; Robert Strassl
Journal:  Virulence       Date:  2020-12       Impact factor: 5.882

6.  US CDC Real-Time Reverse Transcription PCR Panel for Detection of Severe Acute Respiratory Syndrome Coronavirus 2.

Authors:  Xiaoyan Lu; Lijuan Wang; Senthilkumar K Sakthivel; Brett Whitaker; Janna Murray; Shifaq Kamili; Brian Lynch; Lakshmi Malapati; Stephen A Burke; Jennifer Harcourt; Azaibi Tamin; Natalie J Thornburg; Julie M Villanueva; Stephen Lindstrom
Journal:  Emerg Infect Dis       Date:  2020-05-15       Impact factor: 6.883

7.  A Novel Coronavirus from Patients with Pneumonia in China, 2019.

Authors:  Na Zhu; Dingyu Zhang; Wenling Wang; Xingwang Li; Bo Yang; Jingdong Song; Xiang Zhao; Baoying Huang; Weifeng Shi; Roujian Lu; Peihua Niu; Faxian Zhan; Xuejun Ma; Dayan Wang; Wenbo Xu; Guizhen Wu; George F Gao; Wenjie Tan
Journal:  N Engl J Med       Date:  2020-01-24       Impact factor: 91.245

  7 in total
  4 in total

Review 1.  Strategies for Scaling up SARS-CoV-2 Molecular Testing Capacity.

Authors:  Sanchita Das; Karen M Frank
Journal:  Clin Lab Med       Date:  2022-03-08       Impact factor: 2.172

2.  SARS-CoV-2 and RT-PCR Testing in Travelers: Results of a Cross-sectional Study of Travelers at Iraq's International Borders.

Authors:  Doaa Abdulkhaleq Salih; Jivan Qasim Ahmed; Marwan Khalil Qader; Mohammed S Shukur; Bayar K A Zeebaree; Mateen Bahjat Sadeq; Sirwan Najeeb Mohammed; Hushank Nadhmi Masiha; Ibrahim Mohammed Abdullah; Omar Mohammed; Luqman Saleh Taha; Rawand Al-Qadi
Journal:  Disaster Med Public Health Prep       Date:  2022-06-07       Impact factor: 5.556

3.  Rapid Extraction of Viral Nucleic Acids Using Rotating Blade Lysis and Magnetic Beads.

Authors:  Minju Bae; Junsoo Park; Hyeonah Seong; Hansol Lee; Wonsuk Choi; Jiyun Noh; Woojoo Kim; Sehyun Shin
Journal:  Diagnostics (Basel)       Date:  2022-08-17

4.  A single-center report of COVID-19 disease course and management in liver transplanted pediatric patients.

Authors:  Muhammed Yuksel; Hacer Akturk; Ozlem Mizikoglu; Ertug Toroslu; Cigdem Arikan
Journal:  Pediatr Transplant       Date:  2021-06-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.