Florian J Mayer1, Franz Ratzinger2, Ralf L J Schmidt1, Georg Greiner1, Olfert Landt3, Alexander Am Ende3, Victor M Corman4, Nicole Perkmann-Nagele1, Thomas Watkins-Riedel1, Dagmar Petermann1, Karoline Abadir5, Josef Zweimüller-Mayer5, Robert Strassl1.
Abstract
Currently, testing for coronavirus is performed with time aene">nd personnel consuming PCR assays. The aim of this study was to evaluate the sensitivity, specificity aene">nd capacity of a fully automated, raene">ndom access high-throughput real-time PCR-based diagnostic platform for the detection ofEntities:
Keywords: COVID-19; Coronavirus; SARS-CoV-2
Mesh:
Substances:
Year: 2020 PMID: 32726172 PMCID: PMC7549918 DOI: 10.1080/21505594.2020.1798041
Source DB: PubMed Journal: Virulence ISSN: 2150-5594 Impact factor: 5.882
Ct-Values of the LightCycler480 system and the NeuMoDx N96 for the detection of in-vitro transcribed RNA containing the amplicon region of SARS-COV2 (three measurements for each sample), cell culture heat inactivated SARS-COV2 (two measurements for each sample) and respiratory specimen of a patient with confirmed COVID-19 (one measurement for each sample), according to a dilution row series. Every given Ct-value indicates one PCR run. Bottom section shows the assessment of the specificity. #The 89 bp long artificial RNA fragment was too short for the extraction with the “MagNa Pure compact system”; §The precision of Ct-values are attenuated at extremely low virus concentrations. *refers to the Ct-value of the LightCycler 480II.
| RNA concentration | Ct-Values Lightcycler 480II | Ct-Values NeuMoDxN96 |
|---|---|---|
| In-vitro transcribed RNA | ||
| 10e5 (cp/ml) | not performed | 23.3/24.2/24.2 |
| 10e4 (cp/ml) | not performed | 27.4/27.4/27.4 |
| 10e3 (cp/ml) | not performed | 30.6/31.6/30.7 |
| 10e2 (cp/ml) | not performed | 34.5/33.5/33.9 |
| 10e1 (cp/ml) | not performed | neg/41.6/neg |
| Inactivated cell culture SARS-COV2 | ||
| 1.6e2 (cp/µl) | 29.78/29.7 | 24.1/25.8 |
| 1.6e1 (cp/µl) | 32.5/32.5 | 29.9/28.9 |
| 1.6e°.1 (cp/µl) | 34.47/34.9 | 32.1/32.6 |
| 1.6e°.01 (cp/µl) | Neg/neg | 37.3/34.6 |
| Clinical sample (respiratory specimen, undiluted Ct 23.4*) | ||
| 1:10 | 25.5 | 22.8 |
| 1:100 | 28.8 | 24.6 |
| 1:1,000 | 32.7 | 29.3 |
| 1:10,000 | 36.8 | 34.8 |
| 1:100,000 | neg | 34.6 |
| Assessment of Specificity | ||
| Influenza A | neg | neg |
| Influenza B | neg | neg |
| RSV | neg | neg |