M-mode echocardiograms for determination of optimal left atrial timing in patients with dual chamber pacemakers.

M Wish1, J S Gottdiener, A I Cohen, R D Fletcher.   

Abstract

To determine if the A wave of the mitral valve echocardiogram can be used as a marker for left atrial (LA) activity and assist in the programming of dual chamber pacemakers, 156 echocardiograms with the mitral A wave present were obtained from 23 patients with dual chamber pacemakers, all of whom had bipolar esophageal recordings of LA depolarization. Twelve of these patients also underwent hemodynamic study with cardiac function determined at 5 different pacemaker settings: ventricular demand pacing and dual chamber sequential pacing at 0 or 25, 150, 200 and 250 ms programming atrioventricular (AV) delay. The time delay from right atrial pacing artifact to onset and peak of mitral A wave was linearly related to the time from atrial pacing artifact to LA depolarization on the esophageal lead (p less than 0.001). As pacing mode changed from dual chamber sequential pacing (DVI) mode to atrial synchronous-ventricular pacing (VDD), the A wave came earlier relative to the ventricular pacing spike, linearly related to the LA to ventricular extension with mode change determined with the esophageal lead (r = 0.94, p less than 0.001). The time from atrial pacing to peak of A wave was shorter in patients whose optimal programmed AV delay was 150 ms compared with those whose optimal AV delay was 200 or 250 ms (p less than 0.02). At the optimal programmed delay for cardiac output, the peak of the A wave was an average of 13 +/- 36 ms after the ventricular pacing spike.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1988        PMID: 3341208     DOI: 10.1016/0002-9149(88)90937-x

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  1 in total

1.  Difference in mechanical atrioventricular delay between atrial sensing and atrial pacing modes in patients with hypertrophic and dilated cardiomyopathy: an electrical hemodynamic catheterization study.

Authors:  Yong-Mei Cha; Rick A Nishimura; David L Hayes
Journal:  J Interv Card Electrophysiol       Date:  2002-06       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.