Validation of the 2WIN Corneal Reflexes App in children.

Elisabetta Racano1, Giulia Di Stefano2, Salvatore Alessi2, Riccardo Pertile3, Federica Romanelli2.   

Abstract

PURPOSE: To test the reliability and accuracy of the 2Win Refractometer Corneal Reflexes App (CR App) in detecting manifest strabismus in pediatric patients.
METHODS: Prospective study involving 167 children with suspected strabismus (mean age 7.6 years; SD = 3.0, range 2-14 years) undergoing the CR App ocular alignment assessment (Pediatric Ophthalmologist) versus the alternate cover test with prism ocular alignment assessment for distance (Orthoptist) as the gold standard. The AAPOS 2013 guidelines for the detection of manifest strabismus in primary position (> 8 PD) were used.
RESULTS: Total sensitivity, specificity, PPV, and NPV for the CR App were 79.2%, 86.2%, 86.4%, and 78.9%, respectively. The overall inconclusive rate was 17.9%, but was 36.3% in children younger than 5. Sensitivity and PPV for vertical deviations were poor (33.3% and 12.5%, respectively). The accuracy of the CR App regarding the degree (in prism diopters) of manifest deviations was tested with the Wilcoxon signed rank sum test: correlation with the gold standard was good for esodeviations (p value = 0.765, not statistically significant) and poorer for exodeviations (p value = 0.056, still not statistically significant), whereas a significant difference (p value = 0.0001) was observed for vertical deviations.
CONCLUSION: The CR App showed good sensitivity, specificity, PPV, and NPV for manifest strabismus > 8PD in accordance with the AAPOS 2013 guidelines; sensitivity and PPV were poor for vertical deviations. The accuracy of the CR App was good for horizontal deviations, but poor for vertical deviations. The inconclusive result rate was high in younger children.

Entities:  

Keywords:  2Win refractometer (Adaptica); CR App (Corneal Reflexes App); Manifest strabismus; Nonrefractive amblyopia risk factors

Year:  2021        PMID: 33404678     DOI: 10.1007/s00417-020-05066-z

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  7 in total

1.  Preschool vision screening: what should we be detecting and how should we report it? Uniform guidelines for reporting results of preschool vision screening studies.

Authors:  Sean P Donahue; Robert W Arnold; James B Ruben
Journal:  J AAPOS       Date:  2003-10       Impact factor: 1.220

2.  Detection of strabismus by the Spot Vision Screener.

Authors:  Mae Millicent W Peterseim; Jennifer D Davidson; Rupal Trivedi; M Edward Wilson; Carrie E Papa; Edward W Cheeseman
Journal:  J AAPOS       Date:  2015-12       Impact factor: 1.220

3.  Instrument-based pediatric vision screening policy statement.

Authors:  Joseph M Miller; Herschel R Lessin
Journal:  Pediatrics       Date:  2012-10-29       Impact factor: 7.124

4.  Self-assessment of angles of strabismus with photographic Purkinje I and IV reflection pattern evaluation.

Authors:  R Effert; J C Barry; R Colberg; A Kaupp; G Scherer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1995-08       Impact factor: 3.117

5.  Guidelines for automated preschool vision screening: a 10-year, evidence-based update.

Authors:  Sean P Donahue; Brian Arthur; Daniel E Neely; Robert W Arnold; David Silbert; James B Ruben
Journal:  J AAPOS       Date:  2013-01-27       Impact factor: 1.220

6.  Sensitivity, Specificity, and Predictive Values: Foundations, Pliabilities, and Pitfalls in Research and Practice.

Authors:  Robert Trevethan
Journal:  Front Public Health       Date:  2017-11-20

7.  Comparative AAPOS Validation of the Birefringent Amblyopia Screener with Isolated Small-Angle Strabismus.

Authors:  Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2020-01-31