Literature DB >> 33402939

Comparison of digital subtraction angiography and contrast-enhanced ultrasound in assessment of carotid stenosis.

Ronggui Liu1, Zhimei Yan1, Guijun Zhang1, Zhaoyan Ding1, Yunping Li1, Zhirong Jiang1.   

Abstract

INTRODUCTION: The performances of contrast-enhanced ultrasound (CEUS) and digital subtraction angiography (DSA) were used to establish an efficient as well as non-invasive clinical technique for the diagnosis of extra-cranial internal carotid artery (ICA) stenosis.
MATERIALS AND METHODS: Thirty-six successive patients (11 women and 25 men, mean age: 65.0 ± 9.2, range: 43-78 years) with internal carotid artery (ICA) stenosis were tested by CEUS and DSA. These tests were carried out by means of Hitachi Preirus ultrasound machine for CEUS and Allura Xper FD20 system (Philips Medical Systems, Nederland B.V.) for DSA. 1.2 ml SonoVue (Bracco, Switzerland) was used a s contrast agent.
RESULTS: The results clearly indicated that there were no noteworthy variations among the distributions recorded by CEUS as well as DSA for the four tested groups. The percentage of diameter stenosis calculated by CEUS was clearly in accordance with the DSA images. CEUS showed accurate results with good specificity and sensitivity at 50%, 70%, and 100%. Also, CEUS performance was relatively better than DSA in the diagnosis of ICA and suitability of CEA.
CONCLUSION: CEUS proved to be a precise non-invasive testing method for the diagnosis of carotid artery stenosis which is more feasible and well-tolerated in patients with various stages of carotid stenosis.
© 2020 Liu R et al.

Entities:  

Keywords:  CEUS; Carotid artery stenosis; contrast-enhanced ultrasound; digital subtraction angiography

Mesh:

Substances:

Year:  2020        PMID: 33402939      PMCID: PMC7750059          DOI: 10.4314/ahs.v20i1.57

Source DB:  PubMed          Journal:  Afr Health Sci        ISSN: 1680-6905            Impact factor:   0.927


Introduction

Carotid endarterectomy (CEA) or carotid artery stenting (CAS) is an operative method carried out to adjust stenosis (narrowing) in the internal carotid artery (ICA)1. The suitability of the patient to this procedure was tested on the basis of digital subtraction angiography (DSA). Consequently, the patients having an ipsilateral carotid occlusion were regarded inappropriate for this surgery. Therefore, it is a critical task as DSA is the only reference standard to assess the carotid stenosis but being an invasive procedure, it is accompanied with a minor risk of post-agiography stroke2–3 On the other hand, ultrasonography being a non-invasive procedure, affords informative indications required for the diagnosis of carotid artery stenosis. In addition, contrast-enhanced ultrasound (CEUS) imaging has been explored and progressively appeared in clinical technology over the past few decades. Furthermore, contemporary with the technological enhancements in the ultrasound scanning equipments, a wide range of other devices have been intended to encounter the requirements of this quickly growing domain of imaging. However, the rapid development of ultrasound technology is hastened by the functioning issues of the doppler procedures and ultrasound imaging. Ultrasound is employed in the investigation of sub-terranean body structures, while the spatial resolution of grayscale imaging and doppler sensitivity get affected with the impactful extent over clinical efficacy of ultrasound4. In addition, contrast agents assure the enhancement of specificity and sensitivity of present ultrasound diagnoses as well as possess the potential to extend wide scope of ultrasound applications. This innovative technique, CEUS is potentially used in the quantification of extracranial ICA stenosis in symptomatic patients with ipsilateral ICA apparent occlusion on conventional US. This study is significantly anticipated to compare CEUS with the standard reference DSA in the patients with apparent carotid stenosis by making use of conventional ultrasound as well as has explored the clinical expediency of CEUS intended for patients suffering from carotid artery stenosis.

Materials and methods

All the patients were provided with printed informed consent.

Patients

This study was a retrospection from April 2012 until March 2014 of 68 carotid artery samples taken from 36 patients (11 women and 25 men, mean age: 65.0 ± 9.2, range: 43–78 years) were evaluated. All the patients were investigated with color doppler flow imaging, B-mode ultrasound, CEUS, and DSA.

Equipment

All the ultrasound examinations were performed by the radiologist with about 5 years experience in vascular ultrasound. CEUS and Conventional ultrasound tests were executed by means of a Hitachi VISION Preirus System (Hitachi Medical Corporation, Tokyo, Japan) with 5–13 MHz broadband linear array probes. All the examinations were concluded within 2 weeks by means of CEUS, conventional ultrasound, and DSA. Followed by ultrasound procedure, the patients were further investigated by DSA.

Conventional ultrasound

Conventional ultrasound tests were performed for all the patients along with gray-scale as well as color doppler ultrasound imaging. Peak systolic velocity (PSV) values of the distal extra-cranial ICA were recorded. Supplementary PSV indices were analyzed.

CEUs

Hitachi HI VISION Preirus machine (Hitachi Medical Corporation, Tokyo, Japan) was utilized with the (contrast) software integrated in the Hitachi system. Contrast-enhanced imaging bearing low-mechanical-index cadence contrast pulse sequencing technology SonoVue (Bracco, Geneva, Switzerland) was employed.

DSA

Allura Xper FD20system (Philips Medical Systems, Nederland B.V.) was used to carry out intra-arterial DSA within 15 days of the sonographic tests. However, obstructive ICA stenosis were categorized as 0–50% (mild), 50–69% (moderate), 70–99% (severe), and 100% (occlusion).

Statistics

Software Statistical Product and Service Solutions 17.0 (Chicago, IL, USA) were used to perform statistical analysis. The mean ± standard deviation is used to express distribution values while median (M) and inter-quartile range (IQR) is considered to express non-normal distribution values.

Results

In the course of experimental study, 36 patients were selected for CEUS and DSA. No contrast agents used have instigated undesirable effects. Sixty-eight carotid arteries (11 women and 25 men, mean age: 65.0 ± 9.2, range: 43–78 years) were evaluated. Comparison of CEUS and DSA in the detection of carotid artery stenosis Distribution of the four groups detected by DSA and CEUS were tabulated in Table-1. No noteworthy variation in the distribution values were so far noticed both by DSA and CEUS for all the four groups (p=0.01). In addition, the percent diameter stenosis values calculated by CEUS imaging intensely correlated with the DSA measurements (r = 0.97, p =0.01).
Table 1

Contrast between the performance results of CEUS and digital subtraction angiography

Total numbers in the four degrees of carotid artery stenosis groupsCEUS
0–49%50–69%70–99%100%
DSA0–49%23000
50–69%1540
70–99%02240
100%0016
Contrast between the performance results of CEUS and digital subtraction angiography Depending upon the four stages of carotid artery stenosis, 50%, 70%, and 100% were taken as standard values to record the specificity and sensitivity DSA and CEUS for evaluating the extent of carotid artery stenosis while the DSA findings were presumed to represent the actual condition. CEUS recordings were in accordance for 7 of 8 vessels (86% sensitivity) for total occlusion, 31 of 33 vessels (92% sensitivity) for ≥ 70% stenosis and 41 of 42 vessels (96% sensitivity) for ≥ 30% stenosis (as shown in Table-2).
Table 2

The sensitivity, specificity, positive, and negative predicted values for CEUS at various cut-off points with the DSA results being the reference standard for actual condition. Note: +ve refers to positive; -ve refers to negative

Degree of carotid artery stenosis observed by CEUSsensitivityspecificity+ve predicted value (%)-ve predicted value (%)
For occlusion8610010098
For 70–100%92909194
For 50–100%969910096
The sensitivity, specificity, positive, and negative predicted values for CEUS at various cut-off points with the DSA results being the reference standard for actual condition. Note: +ve refers to positive; -ve refers to negative

CEUS versus DSA in the patency estimation of ICA along with suitability for CEA

From the patency assessment findings (shown in Table-3 and Table-4) by CEUS versus DSA, the DSA was diagnosed as carotid artery occlusion for one patient while the CEUS evaluation of the same patient was also carotid artery stenosis > 90% (Shown in Figure-1).
Table 3

Performance of CEUS versus DSA in evaluating patency of the ICA

Number of carotid arteriesCEUS
patentoccluded
DSApatent01
occluded111
Table 4

Comparison between CEUS and DSA in the assessment of CEA suitability

Number of carotid arteriesCEUS
suitableunsuitable
DSA (suitability for CEA)suitable01
unsuitable111
Figure 1

CEUS analysis of a patient with ICA stenosis. (A) An image displaying luminal stenosis and plaque ulceration (arrow) followed by the use of contrast agent. (B) Lateral DSA images of the corresponding patient.

Performance of CEUS versus DSA in evaluating patency of the ICA Comparison between CEUS and DSA in the assessment of CEA suitability CEUS analysis of a patient with ICA stenosis. (A) An image displaying luminal stenosis and plaque ulceration (arrow) followed by the use of contrast agent. (B) Lateral DSA images of the corresponding patient.

Discussion

Severe carotid artery stenosis diagnosed patients are subjected to extreme chances of stroke. Increasing obstruction of the ICA apparently increases these chances of ischemic stroke while the degree of stenosis being one of the significant factors to decide the patient selectivity for surgical CEA8. DSA is considered as a golden standard in the diagnosis of vascular diseases, however possess several complications. CEUS is a latest technological innovation that has the potential to perform microvascular imaging7,8. However, numerous benefits of CEUS turn it to be a more effective technology in contrast to magnetic resonance imaging (MRI), computed tomography (CT) and digital subtraction angiography (DSA). Microbubble contrast agents along with various other acquisition techniques are available to considerably improve signals from the blood displaying the vascular lumen. Especially, SonoVue is a chemical substance that is injected in to the blood pool for clearer visibility of the ultrasound pictures, which is stabilized by a phospholipid core filled with sulfur hexafluoride gas9. In the present study, we have monitored stenosis as well as carotid plaques inside the carotid artery repeatedly. There were no side effects identified instantaneously or even after 24-hour follow-up duration. Images of carotid stenosis and carotid plaque identified in the CEUS and DSA images respectively were represented in Figure 1. In carotid imaging, sonographic contrast agents were employed to increase signals in the blood stream as well as to increase delineation of vessel walls facilitating the investigation of carotid plagues and intima media thickness (IMT)10. From Fig. 2, it was clear that DSA diagnosed as occlusion of the internal carotid artery while CEUS also diagnosed the presence of slight occlusion in the bloodstream. Prospective comparison of contrast-enhanced ultrasound and digital subtraction angiography was brought about by the ultrasound imaging modalities.
Fig. 2

(A) An image displaying slender flow (arrow) after the use of (SonoVue) a contrast agent. (B) An image displaying slender flow (arrow) before the use of SonoVue. (C) DSA images of the corresponding patient displaying a small stump of the proximal ICA (arrow) with no noticeable flow away from it, representing a proximal ICA occlusion.

(A) An image displaying slender flow (arrow) after the use of (SonoVue) a contrast agent. (B) An image displaying slender flow (arrow) before the use of SonoVue. (C) DSA images of the corresponding patient displaying a small stump of the proximal ICA (arrow) with no noticeable flow away from it, representing a proximal ICA occlusion. In this experiment, we have employed CEUS and DSA to diagnose stenosis in 68 ICA and compared these methods against conventional DSA as a standard reference. A better specificity and sensitivity was observed for these two techniques at the values of 50%, 70%, and 100%. As per our observation, the possible reason for CEUS results showing stenosis rate lower than that of DSA in one patient might be the presence of acute stenosis portion in the distal internal carotid artery. Consequently, extended research would be desirable to assess the promising clinical expediency of contrast enhanced ultrasound imaging (CEUS) in the determination and proper diagnosis of ICA stenosis of carotid artery.

Conclusion

CEUS has the ability to enhance better visibility of the blood vessels thus providing a precise interpretation of plaques, ulcers, and carotid stenosis. CEUS is a precise noninvasive imaging technique for the diagnosis of carotid artery stenosis which is feasible as well as highly tolerable in patients with various extents of carotid stenosis.
  10 in total

1.  Carotid arteries: contrast-enhanced US angiography--preliminary clinical experience.

Authors:  Yuko Kono; Sean P Pinnell; Claude B Sirlin; Steven R Sparks; Bassem Georgy; Wade Wong; Robert F Mattrey
Journal:  Radiology       Date:  2003-12-29       Impact factor: 11.105

2.  Management of atherosclerotic carotid artery disease: clinical practice guidelines of the Society for Vascular Surgery.

Authors:  Robert W Hobson; William C Mackey; Enrico Ascher; M Hassan Murad; Keith D Calligaro; Anthony J Comerota; Victor M Montori; Mark K Eskandari; Douglas W Massop; Ruth L Bush; Brajesh K Lal; Bruce A Perler
Journal:  J Vasc Surg       Date:  2008-08       Impact factor: 4.268

3.  Randomised trial of endarterectomy for recently symptomatic carotid stenosis: final results of the MRC European Carotid Surgery Trial (ECST)

Authors: 
Journal:  Lancet       Date:  1998-05-09       Impact factor: 79.321

Review 4.  Advances in the management of carotid artery disease: focus on recent evidence and guidelines.

Authors:  John G T Augoustides
Journal:  J Cardiothorac Vasc Anesth       Date:  2012-02       Impact factor: 2.628

5.  Neurologic complications of cerebral angiography. A retrospective study of complication rate and patient risk factors.

Authors:  A M Leffers; A Wagner
Journal:  Acta Radiol       Date:  2000-05       Impact factor: 1.990

6.  Identification of early complications following liver transplantation using contrast enhanced ultrasound (CEUS). First results.

Authors:  Janine Rennert; Christian Dornia; Martina Georgieva; Stefanie Roehrl; Claudia Fellner; Stephan Schleder; Christian Stroszczynski; Ernst-Michael Jung
Journal:  J Gastrointestin Liver Dis       Date:  2012-12       Impact factor: 2.008

7.  CEUS: a new imaging approach for postoperative vascular complications after right-lobe LDLT.

Authors:  Yan Luo; Yu-Ting Fan; Qiang Lu; Bo Li; Tian-Fu Wen; Zhong-Wei Zhang
Journal:  World J Gastroenterol       Date:  2009-08-07       Impact factor: 5.742

8.  BR1: a new ultrasonographic contrast agent based on sulfur hexafluoride-filled microbubbles.

Authors:  M Schneider; M Arditi; M B Barrau; J Brochot; A Broillet; R Ventrone; F Yan
Journal:  Invest Radiol       Date:  1995-08       Impact factor: 6.016

Review 9.  Liver metastases in cancer: detection with contrast-enhanced ultrasonography.

Authors:  J Hohmann; T Albrecht; A Oldenburg; J Skrok; K-J Wolf
Journal:  Abdom Imaging       Date:  2004-06-08

10.  Detection of total occlusion, string sign, and preocclusive stenosis of the internal carotid artery by color-flow duplex scanning.

Authors:  M A Mansour; M A Mattos; D B Hood; K J Hodgson; L D Barkmeier; D E Ramsey; D S Sumner
Journal:  Am J Surg       Date:  1995-08       Impact factor: 2.565

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.