Repeatability of proton magnetic resonance spectroscopy of the brain at 7 T: effect of scan time on semi-localized by adiabatic selective refocusing and short-echo time stimulated echo acquisition mode scans and their comparison.

Tomohisa Okada1, Hideto Kuribayashi2, Lana G Kaiser3, Yuta Urushibata2, Nouha Salibi4, Ravi Teja Seethamraju4, Sinyeob Ahn4, Dinh Ha Duy Thuy1, Koji Fujimoto1, Tadashi Isa1.   

Abstract

BACKGROUND: Proton magnetic resonance spectroscopy (MRS) provides a unique opportunity for in vivo measurements of the brain's metabolic profile. Two methods of mainstream data acquisition are compared at 7 T, which provides certain advantages as well as challenges. The two representative methods have seldom been compared in terms of measured metabolite concentrations and different scan times. The current study investigated proton MRS of the posterior cingulate cortex using a semi-localized by adiabatic selective refocusing (sLASER) sequence and a short echo time (TE) stimulated echo acquisition mode (sSTEAM) sequence, and it compared their reliability and repeatability at 7 T using a 32-channel head coil.
METHODS: Sixteen healthy subjects were prospectively enrolled and scanned twice with an off-bed interval between scans. The scan parameters for sLASER were a TR/TE of 6.5 s/32 ms and 32 and 48 averages (sLASER×32 and sLASER×48, respectively). The scan parameters for sSTEAM were a TR/TE of 4 s/5 ms and 32, 48, and 64 averages (sSTEAM4×32, sSTEAM4×48, and sSTEAM4×64, respectively) in addition to that with a TR/TE of 8 s/5 ms and 32 averages (sSTEAM8×32). Data were analyzed using LCModel. Metabolites quantified with Cramér-Rao lower bounds (CRLBs) >50% were classified as not detected, and metabolites quantified with mean or median CRLBs ≤20% were included for further analysis. The SNR, CRLBs, coefficient of variation (CV), and metabolite concentrations were statistically compared using the Shapiro-Wilk test, one-way ANOVA, or the Friedman test.
RESULTS: The sLASER spectra for N-acetylaspartate + N-acetylaspartylglutamate (tNAA) and glutamate (Glu) had a comparable or higher SNR than sSTEAM spectra. Ten metabolites had lower CRLBs than prefixed thresholds: aspartate (Asp), γ-aminobutyric acid (GABA), glutamine (Gln), Glu, glutathione (GSH), myo-inositol (Ins), taurine (Tau), the total amount of phosphocholine + glycerophosphocholine (tCho), creatine + phosphocreatine (tCr), and tNAA. Performance of the two sequences was satisfactory except for GABA, for which sLASER yielded higher CRLBs (≥18%) than sSTEAM. Some significant differences in CRLBs were noted, but they were ≤2% except for GABA and Gln. Signal averaging significantly lowered CRLBs for some metabolites but only by a small amount. Measurement repeatability as indicated by median CVs was ≤10% for Gln, Glu, Ins, tCho, tCr, and tNAA in all scans, and that for Asp, GABA, GSH, and Tau was ≥10% under some scanning conditions. The CV for GABA according to sLASER was significantly higher than that according to sSTEAM, whereas the CV for Ins was higher according to sSTEAM. An increase in signal averaging contribute little to lower CVs except for Ins.
CONCLUSIONS: Both sequences quantified brain metabolites with a high degree of precision and repeatability. They are comparable except for GABA, for which sSTEAM would be a better choice. 2021 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  7 Tesla (7 T); Cramér-Rao lower bound; Magnetic resonance spectroscopy (MRS); repeatability; semi-localized by adiabatic selective refocusing (sLASER); short echo time (TE) stimulated echo acquisition mode (sSTEAM)

Year:  2021        PMID: 33392007      PMCID: PMC7719915          DOI: 10.21037/qims-20-517

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  32 in total

1.  Field mapping without reference scan using asymmetric echo-planar techniques.

Authors:  R Gruetter; I Tkác
Journal:  Magn Reson Med       Date:  2000-02       Impact factor: 4.668

2.  Proton T2 relaxation study of water, N-acetylaspartate, and creatine in human brain using Hahn and Carr-Purcell spin echoes at 4T and 7T.

Authors:  Shalom Michaeli; Michael Garwood; Xiao-Hong Zhu; Lance DelaBarre; Peter Andersen; Gregor Adriany; Hellmut Merkle; Kamil Ugurbil; Wei Chen
Journal:  Magn Reson Med       Date:  2002-04       Impact factor: 4.668

3.  Assessing Brain Metabolism With 7-T Proton Magnetic Resonance Spectroscopy in Patients With First-Episode Psychosis.

Authors:  Anna M Wang; Subechhya Pradhan; Jennifer M Coughlin; Aditi Trivedi; Samantha L DuBois; Jeffrey L Crawford; Thomas W Sedlak; Fredrick C Nucifora; Gerald Nestadt; Leslie G Nucifora; David J Schretlen; Akira Sawa; Peter B Barker
Journal:  JAMA Psychiatry       Date:  2019-03-01       Impact factor: 21.596

4.  Dielectric pads and low- B1+ adiabatic pulses: complementary techniques to optimize structural T1 w whole-brain MP2RAGE scans at 7 tesla.

Authors:  Kieran R O'Brien; Arthur W Magill; Jean Delacoste; Jose P Marques; Tobias Kober; Hans-Peter Fautz; Francois Lazeyras; Gunnar Krueger
Journal:  J Magn Reson Imaging       Date:  2013-11-05       Impact factor: 4.813

5.  Determining the in vivo transverse relaxation time of GABA in the human brain at 7T.

Authors:  Jarunee Intrapiromkul; He Zhu; Ying Cheng; Peter B Barker; Richard A E Edden
Journal:  J Magn Reson Imaging       Date:  2012-12-13       Impact factor: 4.813

6.  In vivo detection of GABA and glutamate with MEGA-PRESS: reproducibility and gender effects.

Authors:  Ruth L O'Gorman; Lars Michels; Richard A Edden; James B Murdoch; Ernst Martin
Journal:  J Magn Reson Imaging       Date:  2011-05       Impact factor: 4.813

7.  Short echo time 1H-MRSI of the human brain at 3T with minimal chemical shift displacement errors using adiabatic refocusing pulses.

Authors:  Tom W J Scheenen; Dennis W J Klomp; Jannie P Wijnen; Arend Heerschap
Journal:  Magn Reson Med       Date:  2008-01       Impact factor: 4.668

8.  Estimation of metabolite concentrations from localized in vivo proton NMR spectra.

Authors:  S W Provencher
Journal:  Magn Reson Med       Date:  1993-12       Impact factor: 4.668

9.  Test-retest reproducibility of neurochemical profiles with short-echo, single-voxel MR spectroscopy at 3T and 7T.

Authors:  Melissa Terpstra; Ian Cheong; Tianmeng Lyu; Dinesh K Deelchand; Uzay E Emir; Petr Bednařík; Lynn E Eberly; Gülin Öz
Journal:  Magn Reson Med       Date:  2015-10-26       Impact factor: 4.668

10.  Detection of Glutamate Alterations in the Human Brain Using 1H-MRS: Comparison of STEAM and sLASER at 7 T.

Authors:  Anouk Marsman; Vincent O Boer; Peter R Luijten; Hilleke E Hulshoff Pol; Dennis W J Klomp; René C W Mandl
Journal:  Front Psychiatry       Date:  2017-04-21       Impact factor: 4.157

View more
  4 in total

1.  Reproducibility of 7-T brain spectroscopy using an ultrashort echo time STimulated Echo Acquisition Mode sequence and automated voxel repositioning.

Authors:  Meredith A Reid; Martha R Forloines; Nouha Salibi
Journal:  NMR Biomed       Date:  2021-10-08       Impact factor: 4.044

Review 2.  Fifteen years of urea cycle disorders brain research: Looking back, looking forward.

Authors:  Kuntal Sen; Matthew Whitehead; Carlos Castillo Pinto; Ljubica Caldovic; Andrea Gropman
Journal:  Anal Biochem       Date:  2021-10-09       Impact factor: 3.365

Review 3.  Neuroimaging at 7 Tesla: a pictorial narrative review.

Authors:  Tomohisa Okada; Koji Fujimoto; Yasutaka Fushimi; Thai Akasaka; Dinh H D Thuy; Atsushi Shima; Nobukatsu Sawamoto; Naoya Oishi; Zhilin Zhang; Takeshi Funaki; Yuji Nakamoto; Toshiya Murai; Susumu Miyamoto; Ryosuke Takahashi; Tadashi Isa
Journal:  Quant Imaging Med Surg       Date:  2022-06

Review 4.  Evaluation of Left Ventricular Function Using Four-Dimensional Flow Cardiovascular Magnetic Resonance: A Systematic Review.

Authors:  Jiaxing Jason Qin; Ben Indja; Alireza Gholipour; Mustafa Gök; Stuart M Grieve
Journal:  J Cardiovasc Dev Dis       Date:  2022-09-12